ePlex BCID-FP 12 Kit BoxRecalls & reported problems

Genmark Diagnostics, Inc. · ePlex Blood Culture Identification Fungal Pathogen (FP) 12-test kit box · Class 2

All models of ePlex (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00857167005290), received between Nov 26, 2024 and Jan 13, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result3
  • False Positive Result1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 13, 2026False Negative Result

    THERE WAS AN ALLEGATION OF DISCREPANT RESULTS FOR THREE PATIENT SAMPLES USING THE COBAS® EPLEX BLOOD CULTURE IDENTIFICATION FUNGAL PATHOGEN (BCID-GP) PANEL ON A EPLEX BASE. THE ALLEGED SAMPLES GENERATED NEGATIVE RESULTS FOR CANDIDA AURIS. THE CULTURE RESULTS IDENTIFIED CANDIDA AURIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24056811 ↗
  • MalfunctionJul 24, 2025False Negative Result; False Positive Result; False Negative Result; False Positive Result

    THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FOR THE EPLEX BLOOD CULTURE IDENTIFICATION FUNGAL PATHOGEN (BCID-FP) PANEL USING THE COBAS EPLEX CORE UNIT. SAMPLE 1 INITIAL RESULT WAS C. ALBICANS AND C. FAMATA. ON (B)(6) 2025, THE REPEAT RESULT WAS C. ALBICANS AND C. FAMATA. THE CULTURE RESULTS WERE C. ALBICANS AND C. DUBLINIENSIS SAMPLE 2 INITIAL RESULT WAS C. DUBLINIENSIS AND C. FAMATA. ON (B)(6) 2025, THE REPEAT RESULT WAS C. FAMATA. THE CULTURE RESULT WAS C. DUBLINIENSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22586086 ↗
  • MalfunctionNov 26, 2024False Negative Result; Non Reproducible Results; False Negative Result; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FOR TWO PATIENT SAMPLES USING THE EPLEX BLOOD CULTURE IDENTIFICATION FUNGAL PATHOGEN (BCID-FP) PANEL ON A EPLEX INSTRUMENT. SAMPLE 1 INITIALLY GENERATED NEGATIVE RESULTS, HOWEVER, MALDI IDENTIFICATION OF THE SUBCULTURE GROWTH WAS CANDIDA LUSITANIAE WITH LOG SCORE >2. ON (B)(6) 2024, SAMPLE 2 INITIALLY GENERATED A POSITIVE RESULT FOR THE CANDIDA PARAPSILOSIS TARGET. CULTURE DETECTED CANDIDA PARAPSILOSIS AND CANDIDA LUSITANIAE. THE MALDI LOG SCORE WAS >2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20783765 ↗