Skip to content

Wahoo FG 00074Recalls & reported problems

Q'Apel Medical LLC · SelectFlex 072 Neurovascular Access System, 105cm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00857545008004), received between May 2, 2025 and May 2, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Obstruction of Flow” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 2, 2025Obstruction of Flow; Obstruction of Flow

    DURING A RETROSPECTIVE LEGACY COMPLAINT FILE REVIEW FOR POTENTIAL MDR REPORTABLE EVENTS, (B)(4) (EVENT DATE-01/13/2021), WAS OBSERVED TO HAVE MET THE REPORTABILITY CRITERIA AFTER THE INVESTIGATION OF THE RETURNED PRODUCT WAS COMPLETED ((B)(6) 2021). NOTE: ON (B)(6) 2023, Q'APEL MEDICAL INC. DECIDED TO CEASE THE PRODUCTION OF WAHOO SELECTFLEX 072 NEUROVASCULAR ACCESS SYSTEM. TIMELINE OF EVENTS: -(B)(4) INITIAL MDR EVALUATION: AT THE TIME OF THE COMPLAINT, Q'APEL MEDICAL DETERMINED THE EVENT WAS NOT REPORTABLE UNDER 21 CFR 803.50 BASED ON AVAILABLE INFORMATION. SPECIFICALLY, THE EVALUATION CONCLUDED THE DEVICE DID NOT: 1. CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY; OR 2. MALFUNCTION IN A WAY THAT, IF IT WERE TO RECUR, WOULD LIKELY RESULT IN A DEATH OR SERIOUS INJURY. -CASE OVERVIEW (PROVIDED BY SALES REPRESENTATIVE): 1. WAHOO CATHETER WAS TRACKED TO THE TARGET ANATOMY AND USED TO DEPLOY A SURPASS FLOW DIVERTER. 2. UPON ATTEMPT TO PASS A WIRE POST-DEPLOYMENT, RESISTANCE WAS NOTED. 3. RESISTANCE DURING WITHDRAWAL PROMPTED SIMULTANEOUS REMOVAL OF THE WAHOO AND 8F SHEATH. 4. INITIAL INSPECTION SUGGESTED THE CATHETER WAS INTACT. 5. THE PATIENT HAD HIGHLY TORTUOUS VASCULATURE. NO HAR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21952073 ↗

Other Catheter, Percutaneous devices

All Catheter, Percutaneous devices →