Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00857545008011), received between Mar 8, 2024 and Mar 8, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • Vascular Dissection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 8, 2024Use of Device Problem; Use of Device Problem

    RETROACTIVE REPORTING TO THE FDA OF AN INCIDENT THAT OCCURED IN THE UK MARKET ON 4/20/2023. Q'APEL MEDICAL BECAME AWARE OF THE INCIDENT ON 4/20/2023. Q'APEL MEDICAL FILED AN INITIAL MIR REPORT ON 4/26/23. Q'APEL MEDICAL FILED A FINAL MIR REPORT ON 5/24/23. CONCLUSION: THE DEVICE DID NOT MALFUNCTION. THE OPERATOR CONFIRMED HE OVER-INFLATED ACCORDING TO THE VESSEL SIZE AND DID NOT FOLLOW RECOMMENDATIONS OR THE MANUFACTURER'S INSTRUCTIONS. WALRUS USED WAS THE 90CM VERSION BG8087-090 LOT NO. FG220811J-03. PHYSICIAN NAME: DR (B)(6). HOSPITAL NAME: (B)(6) HOSPITAL. 75 YO MALE WITH A RIGHT M1 OCCLUSION SUFFERING AN ACUTE ISCHEMIC STROKE. PATIENT WAS TREATED UNDER GENERAL ANESTHESIA. OTHER DEVICES UTILIZED IN THE PROCEDURE WERE A SOFIA PLUS ASPIRATION CATHETER, PHENNOM 21 MICRO CATHETER AND A SOLITAIRE STENT. RETRIEVER THE WALRUS BGC WAS PLACED IN THE DISTAL CERVICAL SEGMENT OF THE R ICA PHYSICIAN INITIALLY INFLATED THE WALRUS WITH APPROX 0.3CC'S OF CONTRAST/SALINE MIX. THEY FELT THAT THE BALLOON WAS NOT STABLE SO CONTINUED TO INFLATE WITH MAXIMUM RECOMMENDED VOLUME OF 0.6CC'S. PHYSICIAN COMPLETED ONE PASS WITH THE SOLITAIRE STENT RETRIEVER AND THE SOFIA PLUS ASPIRATION CATHETER. ONCE THAT

    Patient problems: Vascular Dissection

    MAUDE report 18871670 ↗

Other Catheter, Percutaneous devices