Skip to content

PowerLook AMP SLD GE US V7.2Recalls & reported problems

Icad, Inc. · The SecondLook Computer-Aided Detection (CAD) system for mammography is intended to identify and mark regions of interest on screening and diagnostic mammograms from full-field digital mammography (FFDM) systems to bring them to the attention of the radiologist after an initial reading has been completed. Thus the system prompts the radiologist to areas on mammograms for second review only. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00858012005014), received between Jan 11, 2018 and Jan 11, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Problem with Software Installation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Problem with Software Installation1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 11, 2018Problem with Software Installation; Problem with Software Installation

    A CUSTOMER SITE INDICATED THAT NO CAD MARKINGS WERE REPORTED FROM THE CAD AFTER PROCESSING PATIENT IMAGES. THE OEM MAMMOGRAPHY FIELD SERVICE ENGINEER (FSE) DID NOT APPLY THE CAD LICENSE AFTER INSTALLING THE CAD SOFTWARE. THIS RESULTED IN THE SITE NOT HAVING CAD MARKS ON CASES FOR SOME TIME. NO PATIENT INJURY OR ADVERSE EVENT WAS ASSOCIATED WITH THIS INCIDENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7183668 ↗

Other Analyzer,Medical Image devices

All Analyzer,Medical Image devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page