OMNI Surgical System FG 07441Recalls & reported problems
Sight Sciences, Inc. · OMNI Surgical System: Ergo series of revisions · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Sight Sciences OMNI Surgical System handpiece is used in glaucoma eye surgery to feed a tiny catheter into the eye's drainage canal. The 24 reports mostly describe the catheter kinking or breaking off inside the eye during the procedure, with the broken piece usually removed using forceps. They include 20 injuries and no deaths. No recalls name this device. Reports dropped from 6 in the prior 12 months to none recently.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 20
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00858027006525), received between May 9, 2023 and Jul 30, 2025 — classified as 20 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation18
- Unintended Ejection3
- Adverse Event Without Identified Device or Use Problem2
- Material Twisted/Bent1
- Retraction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- Intraocular Pressure Increased2
- Hyphema1
- Intraocular Pressure Decreased1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
InjuryJul 30, 2025Material Fragmentation; Material Fragmentation
THE PHYSICIAN WHO PERFORMED AN OMNI PROCEDURE REPORTED THAT THE CATHETER SEVERED WHILE PERFORMING THE 180-DEGREE GONIOTOMY PROCEDURE. THE SEVERED PORTION OF THE CATHETER WAS REMOVED USING FORCEPS. A PARTIAL GONIOTOMY OCCURRED WHILE REMOVING THE SEVERED PORTION OF THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22668258 ↗InjuryJul 10, 2025Material Fragmentation; Material Fragmentation
THE PHYSICIAN WHO PERFORMED AN OMNI SURGERY REPORTED THAT THE CATHETER BROKE OFF DURING THE EXCISION OF THE TRABECULAR MESHWORK. THE CATHETER WAS REMOVED FROM THE EYE AND THE PROCEDURE WAS COMPLETED USING A NEW DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22472412 ↗InjuryMay 22, 2025Material Fragmentation; Material Fragmentation
THE PHYSICIAN WHO PERFORMED AN OMNI SURGERY REPORTED THAT THE CATHETER BROKE-OFF WHILE RETRACTING THE DEVICE DURING THE TRABECULOTOMY PROCEDURE . THE BROKEN PART OF THE CATHETER WAS REMOVED FROM THE EYE. THERE WAS NO FURTHER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22074065 ↗InjuryApr 25, 2025Material Fragmentation; Material Fragmentation
THE PHYSICIAN WHO PERFORMED AN OMNI PROCEDURE REPORTED THAT AFTER SUCCESSFUL COMPLETION OF THE CANALOPLASTY PROCEDURE, THE DEVICE WAS MET WITH RESISTANCE WHEN THE CATHETER WAS RE-ADVANCED INTO THE SCHLEMM' FOR THE TRABECULOTOMY PROCEDURE. THE PHYSICIAN UNPARKED THE CANNULA TIP AND PULLED INTO THE MIDDLE OF THE ANTERIOR CHAMBER TO BEGIN RETRACTING THE CATHETER. IT WAS NOTED THAT THE DEVICE WHEEL WAS HARD TO RETRACT AND AFTER 2-3 CLICK BACKS, THE CATHETER BROKE OFF FROM THE CANNULA. THE PHYSICIAN USED MST MICRO FORCEPS TO REMOVE THE CATHETER AND COMPLETE TRABECULOTOMY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21911326 ↗InjuryJan 22, 2025Material Fragmentation; Material Fragmentation
THE PHYSICIAN WHO PERFORMED AN OMNI PROCEDURE REPORTED THAT THE MICROCATHETER KINKED AND BROKE OFF FROM THE DEVICE WHILE ADVANCING IT IN THE SCHLEMM'S CANAL. THE BROKEN PART WAS RETRIEVED USING FORCEPS. THERE WAS NO FURTHER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21208181 ↗InjuryJan 3, 2025Material Fragmentation; Material Fragmentation
THE PHYSICIAN WHO PERFORMED AN OMNI PROCEDURE REPORTED THAT A KINK IN THE MICROCATHETER WAS NOTED WHILE ADVANCING IN TO THE SCHEMM'S CANAL. WHILE RETRACTING IT BACK TO RELEASE THE KINK, THE MICROCATHETER BROKE OFF. THE BROKEN PART WAS RETRIEVED USING A MST FORCEPS AND THE PROCEDURE WAS COMPLETED USING A FRESH DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21068603 ↗InjuryOct 29, 2024Material Fragmentation; Material Fragmentation
THE PHYSICIAN WHO PERFORMED AN OMNI PROCEDURE REPORTED THAT THE MICROCATHETER BROKE OFF DURING THE PROCEDURE WHILE THE DEVICE WAS IN THE SCHLEMM'S CANAL. THE SURGEON USED A PAIR OF FORCEPS TO RETRIEVE THE PORTION OF THE MICROCATHETER THAT BROKE OFF. THERE WAS NO FURTHER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20562521 ↗MalfunctionOct 28, 2024Unintended Ejection; Unintended Ejection
DURING AN OMNI PROCEDURE, THE PHYSICIAN PERFORMED A 180-DEGREE CANALOPLASTY IN THE SUPERIOR HEMISPHERE AND UPON RETRACTION, THE MICROCATHETER AUTOMATICALLY ADVANCED FORWARD IN A SPRING LIKE MANNER ABOUT 2MM IN LENGTH. THE ACTION OCCURRED WHILE THE DEVICE WAS STILL IN THE SCHLEMM'S CANAL. THERE WAS NO INJURY REPORTED. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20547012 ↗InjurySep 13, 2024Material Fragmentation; Material Fragmentation
THE SURGEON WHO PERFORMED AN OMNI PROCEDURE REPORTED THAT DURING THE CANALOPLASTY PROCEDURE IN THE INFERIOR HEMISPHERE, THE CATHETER GOT STUCK WHILE ADVANCING AND BROKE OFF IN THE EYE. THE BROKEN PART THAT WAS LODGED IN THE TRABECULAR MESHWORK WAS RETRIEVED USING MICRO FORCEPS. THE PATIENT HAD NARROW ANGLES, SCARRED TM AND SOME IRIS SYNECHIAE PRIOR TO THE OMNI PROCEDURE. OMNI IS CONTRAINDICATED FOR PATIENTS WITH NARROW ANGLE GLAUCOMA. THERE WAS NO FURTHER INJURY ASSOCIATED WITH THE OMNI PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20211254 ↗MalfunctionAug 22, 2024Unintended Ejection; Unintended Ejection
DURING AN OMNI PROCEDURE IT WAS REPORTED THAT THE MICROCATHETER ADVANCED/SHOT FORWARD ON ITS OWN. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY THAT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20027473 ↗MalfunctionAug 9, 2024Unintended Ejection; Unintended Ejection
THE SURGEON WHO PERFORMED AN OMNI PROCEDURE REPORTED THAT THE MICROCATHETER KEPT SPRINGING BACK IN THE CANAL WHILE TRYING TO RETRACT THE DEVICE AFTER COMPLETING THE CANALOPLASTY IN THE FIRST HEMISPHERE. BOTH INDEX FINGERS HAD TO BE USED TO RETRACT AND HOLD THE MICRO CATHETER IN PLACE SO THE DEVICE COULD BE REMOVED WITHOUT INADVERTENTLY TEARING THE TRABECULAR MESHWORK. A NEW DEVICE WAS USED TO COMPLETE THE CANALOPLASTY AND GONIOTOMY PROCEDURE IN THE SECOND HEMISPHERE. THERE WAS NO PATIENT INJURY THAT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19955502 ↗MalfunctionJun 14, 2024Material Fragmentation; Material Fragmentation
THE PHYSICIAN WHO PERFORMED OMNI SURGERY REPORTED THAT THE CATHETER KINKED UPON FEELING RESISTANCE AT ABOUT 90-100 DEGREES DURING THE PROCEDURE. THE CATHETER WAS STUCK WHEN TRYING TO PULL BACK TO REENGAGE. THE CATHETER WAS STUCK AND COULD NOT BE PULLED BACK SO IT WAS PULLED OUT BY DOING A GONIOTOMY. THE DEVICE WAS PULLED OUT OF THE EYE WHILE TRYING TO MOVE THE WHEEL FORWARD OR BACKWARDS AND THE CATHETER BROKE OFF (OUTSIDE THE EYE). THERE WAS NO INTERVENTION REQUIRED AS THE CATHETER BROKE OUTSIDE THE EYE. THE PROCEDURE WAS COMPLETED USING A FRESH DEVICE. AN EVALUATION OF THE RETURNED DEVICE AND A REVIEW OF THE LOT HISTORY AND RELEVANT COMPONENTS OF THE RETURNED DEVICE DETERMINED THAT THERE IS NO EVIDENCE THAT THE DEVICE DESIGN OR MANUFACTURING PROCESS WAS THE CAUSE OF THE INCIDENT. A VIDEO OF THE ACTUAL PROCEDURE WAS NOT AVAILABLE AND THEREFORE A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED; RETURNED DEVICE FOUND TO MEET SPECIFICATION, AND USER HISTORY OF THIS PHYSICIAN AS WELL AS REVIEW OF PAST SIMILAR COMPLAINTS REPORTED BY OTHER PHYSICIANS, WHERE VIDEOS OF THE PROCEDURE WAS PROVIDED THE PROBABLE ROOT CAUSE FOR THIS INCIDENT IS LIKELY DUE TO A SURGICAL TECHNIQUE OR USE ERROR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19542548 ↗InjuryApr 29, 2024Material Fragmentation; Material Fragmentation
THE PHYSICIAN USING AN OMNI DEVICE REPORTED THAT THE CATHETER KINKED ON ADVANCEMENT AND BROKE-OFF IN THE SCHLEMM'S CANAL. THE SEVERED PORTION OF THE MICROCATHETER WAS REMOVED USING FORCEPS. THERE WERE NO FURTHER INJURIES REPORTED. THE PROCEDURE WAS COMPLETED USING A FRESH DEVICE. AN EVALUATION OF THE RETURNED DEVICE AND A REVIEW OF THE LOT HISTORY AND RELEVANT COMPONENTS OF THE RETURNED DEVICE DETERMINED THAT THERE IS NO EVIDENCE THAT THE DEVICE DESIGN OR MANUFACTURING PROCESS WAS THE CAUSE OF THE INCIDENT. A VIDEO OF THE ACTUAL PROCEDURE WAS NOT AVAILABLE AND THEREFORE A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED; HOWEVER, BASED ON: 1) RETURNED DEVICE FOUND TO MEET SPECIFICATION, AND 2) USER HISTORY OF THIS PHYSICIAN AS WELL AS REVIEW OF PAST SIMILAR COMPLAINTS REPORTED BY OTHER PHYSICIANS, WHERE VIDEOS OF THE PROCEDURE WAS PROVIDED THE PROBABLE ROOT CAUSE FOR THIS INCIDENT IS LIKELY DUE TO A SURGICAL TECHNIQUE OR USE ERROR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19209735 ↗InjuryApr 10, 2024Retraction Problem; Material Twisted/Bent; Retraction Problem; Material Twisted/Bent
THE PHYSICIAN USING AN OMNI PROCEDURE REPORTED THAT THE CATHETER KINKED DURING THE GONIOTOMY PROCEDURE AND DIDN'T ALLOW FOR A SMOOTH RETRACTION LIKE IT USUALLY DOES. THERE WAS SIGNIFICANT HEME AT THE ENTRY POINT OF THE CANNULA THAT IS SUSPECTED TO BE RELATED TO THE RETRACTION NOT BEING SMOOTH. DAY 1 POST-OP IOP WAS 18 WITH 0.5MM HYPHEMA. AT THE 1 WEEK FOLLOW UP, IOP WAS 14 AND THE HYPHEMA HAD CLEARED UP. AT THE 4 WEEK FOLLOW UP, IOP HAD SPIKED TO 38MMHG. THE PATIENT WAS PRESCRIBED PREDNISONE POST OP IN ADDITION TO THE LATANOPROST THAT THE PATIENT WAS ON FOR THE PAST YEAR INCLUDING THROUGH THE SURGERY AND POST OP. AT THE 4 WEEK FOLLOW UP, THE PATIENT WAS ADVISED TO STOP THE STEROIDS AND ADD PILOCARPINE. PATIENT'S LAST REPORTED IOP MEASUREMENT WAS 18MMHG.
Patient problems: Hyphema; Intraocular Pressure Increased
MAUDE report 19084363 ↗InjuryFeb 27, 2024Material Fragmentation; Material Fragmentation
THE PHYSICIAN USING AN OMNI DEVICE REPORTED THAT THE CATHETER GOT STUCK AND DETACHED WHILE RETRACTING TO PERFORM THE SECOND HALF OF THE 360 DEGREE CANALOPLASTY PROCEDURE. THE DETACHED MICROCATHETER WAS REMOVED USING FORCEPS. THERE WAS NO FURTHER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18789340 ↗InjuryFeb 22, 2024Material Fragmentation; Material Fragmentation
PHYSICIAN REPORTED CATHETER BREAKING OFF IN THE SCHLEMM'S CANAL DURING TRABECULOTOMY, REQUIRING REMOVAL WITH MICRO FORCEPS. THERE WAS NO FURTHER REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18763322 ↗InjuryJan 10, 2024Material Fragmentation; Material Fragmentation
PHYSICIAN SUCCESSFULLY PERFORMED 180 DEGREE CANALPLASTY UTILIZING AN OMNI DEVICE. UPON SECOND ADVANCEMENT, THE PHYSICIAN REPORTER THE MICROCATHETER BUCKLED AND SNAPPED. THE BROKEN PART OF THE MICROCATHETER WAS REMOVED FROM THE CANAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18490139 ↗InjuryNov 27, 2023Material Fragmentation; Material Fragmentation
THE PHYSICIAN USING AN OMNI DEVICE, REPORTED THAT THE CATHETER SEVERED IN THE EYE DURING PROCEDURE. THE BROKEN PART OF THE MICROCATHETER WAS REMOVED FROM THE CANAL USING A MICRO FORCEPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18213694 ↗InjuryNov 27, 2023Material Fragmentation; Material Fragmentation
THE PHYSICIAN USING AN OMNI DEVICE, REPORTED THAT THE CATHETER SNAPPED IN THE EYE UNDER NO PRESSURE WHILE PERFORMING THE PROCEDURE .THE SEVERED PART OF THE MICROCATHETER WAS RETRIEVED USING A MICROFORCEPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18213693 ↗InjuryJul 26, 2023Material Fragmentation; Material Fragmentation
THE PHYSICIAN USING AN OMNI DEVICE, REPORTED THAT THE CATHETER BROKE WHILE PERFORMING THE GONIOTOMY PROCEDURE. THE BROKEN OFF PART OF THE DEVICE WAS REMOVED USING A FORCEPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17401642 ↗InjuryJun 30, 2023Material Fragmentation; Material Fragmentation
THE PHYSICIAN USING AN OMNI DEVICE, REPORTED THAT THE CATHETER KINKED AND BROKE-OFF WHILE TURNING AND ADVANCING FOR 180 DEGREE GONIOTOMY. THE BROKEN OFF PART OF THE CATHETER WAS RETRIEVED USING A RETINA GRASPING FORCEPS AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17243356 ↗InjuryMay 25, 2023Material Fragmentation; Material Fragmentation
THE PHYSICIAN USING AN OMNI DEVICE, REPORTED THAT THE CATHETER BROKE-OFF IN THE SCHLEMM'S CANAL WHILE TRYING TO PULL THE DEVICE THROUGH THE TRABECULAR MESHWORK FOR THE GONIOTOMY PROCEDURE. THE BROKEN OFF PART OF THE CATHETER WAS RETRIEVED USING A MST FORCEPS FROM THE SCHLEMM'S CANAL AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17004870 ↗InjuryMay 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WITH SEVERE GLAUCOMA UNDERWENT OMNI PROCEDURE, POST-OP DAY 1 IOP WAS 9MMHG AND WEEK 1 THE IOP WAS 6MMHG. STEROID WAS STOPPED AFTER THE FIRST WEEK. UPON PHYSICIAN FOLLOW-UP AT 2 WEEKS POST-OP, PATIENT PRESENTED WITH HIGH IOP THAT NECESSITATED MEDICAL INTERVENTION AND THE PATIENT'S IOP IS BEING MONITORED ON A WEEKLY BASSIS.
Patient problems: Intraocular Pressure Increased
MAUDE report 16911763 ↗InjuryMay 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PATIENT WHO UNDERWENT A COMBINATION OF OMNI AND CATARACT EXTRACTION PROCEDURE, WAS PRESENTED WITH AN IOP OF 10MMHG DURING WEEK 1 POST-OP. ON DAY 11 POSTOP, THE IOP FURTHER DECREASED TO 3MMHG. THE PHYSICIAN INCREASED THE DOSAGE OF PREDNISONE FROM 4X TO 6X A DAY. AT A SUBSEQUENT FOLLOW UP, IT WAS DETERMINED THAT THE PATIENT HAD A SMALL CLEFT WHICH COULD HAVE INDUCED HYPOTONY IN THE PATIENT'S EYE.
Patient problems: Intraocular Pressure Decreased
MAUDE report 16899886 ↗
FDA clearance and classification
Cleared through 510(k) K202678 on Mar 1, 2021 (decision: Substantially Equivalent). Submitted by Sight Sciences, Inc. as “OMNI Surgical System”. The FDA review took 167 days. FDA record ↗
The FDA regulates product code MRH (pump, infusion, ophthalmic) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the OMNI Surgical System FG 07441
Has the OMNI Surgical System FG 07441 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00858027006525), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the OMNI Surgical System FG 07441?
The FDA's MAUDE database lists 24 adverse event reports for this device, received between May 9, 2023 and Jul 30, 2025. In those reports, the most common specific device problems are “Material Fragmentation” (18), “Unintended Ejection” (3) and “Material Twisted/Bent” (1); the most common patient problems are “Intraocular Pressure Increased” (2), “Hyphema” (1) and “Intraocular Pressure Decreased” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the OMNI Surgical System FG 07441?
None of the reports are classified as deaths, 20 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the OMNI Surgical System FG 07441 FDA approved?
It was cleared through 510(k) K202678 on Mar 1, 2021. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the OMNI Surgical System FG 07441, and is it still sold?
The FDA's device database lists Sight Sciences, Inc. as the company responsible for it (brand OMNI Surgical System, model FG 07441). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Pump, Infusion, Ophthalmic devices
- VISCO(TM)360 Viscosurgical System FG 03916Sight Sciences, Inc. · Pump, Infusion, Ophthalmic23 reports
- VIA360 Surgical SystemNew World Medical, Incorporated · Pump, infusion, ophthalmic3 reports
- OMNI Edge Surgical System 1-112Sight Sciences, Inc. · Pump, Infusion, Ophthalmic2 reports
- Streamline Surgical System 10-0066New World Medical, Incorporated · Pump, infusion, ophthalmic1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page