Skip to content

MicroCross 14 Catheter, 135cm RC-2014Recalls & reported problems

Roxwood Medical, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00858480007138), received between May 31, 2023 and May 31, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Material Rupture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Rupture1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 31, 2023Material Rupture; Material Rupture

    IT WAS REPORTED A RC2014 BURST IN THE PATIENT AND CREATED A HOLE IN THE HEART. PER THE PERFUSIONIST, THE RETROGRADE CANNULA WAS EITHER FROZEN OR BURNED DURING A MAZE PROCEDURE. PERFUSIONIST NOTIFIED PHYSICIAN OF HIGH-PRESSURE WITH LOW FLOW (LESS THAN 30CC/MIN) WHILE DELIVERING RETROGRADE CARDIOPLEGIA INTO CORONARY SINUS WITH THE TIP FROZEN/BURNT CAUSED OCCLUSION OF TIP OF THE CANNULAR WHICH CAUSED THE CATHETER TO RUPTURE. THE PATIENT REQUIRED A BOVINE PERICARDIAL PATCH TO REPAIR DAMAGE TO THE ORGAN CAUSED BY THE CANNULA RUPTURING. THE PATIENT WAS DISCHARGED FROM HOSPITAL.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17031634 ↗