EkoSonic® 6779-004Recalls & reported problems

Boston Scientific Corporation · The EkoSonic® Endovascular Device consists of a single use Infusion Catheter and an Ultrasonic Core. The device delivers physician-specified fluids and ultrasound to the intravascular treatment site. It comes in different working lengths and treatment zones. · Class 2

All models of EkoSonic (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00858593006158), received between Mar 29, 2023 and Oct 24, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Leak/Splash1
  • Obstruction of Flow1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 24, 2024Leak/Splash; Obstruction of Flow; Leak/Splash; Obstruction of Flow

    IT WAS REPORTED THAT THERE WAS A DRUG LEAK AND OCCLUSION. AN EKOSONIC ENDOVASCULAR DEVICE, 106X24CM WAS SELECTED FOR USE IN A UNILATERAL THROMBOLYSIS PROCEDURE. AFTER SEVEN HOURS OF EKOS THERAPY, WHILE IN THE ICU, IT WAS OBSERVED THAT THERE WAS A WET SPOT IN THE PATIENT'S BED AND SOME BLOOD BACKING UP IN THE DRUG LUER. THE NURSE ATTEMPTED TO FLUSH THE CATHETER AND OBSERVED THAT FLUID SQUIRTED SIDEWAYS OUT OF THE DRUG INFUSION TUBING, BELOW THE HUB WHERE THE STOPCOCK WAS CONNECTED. THERE WAS A PINHOLE LEAK IN THE HUB OF THE DRUG INFUSION TUBING. THE LINE INTO THE PATIENT WAS COMPLETELY OCCLUDED. TROUBLESHOOTING WAS UNABLE TO RESOLVE THE ISSUE. THE TPA WAS RUN AT A SLOWER RATE TO BYPASS THE ISSUE AND THERAPY WAS COMPLETED IN THE MORNING. THROMBUS WAS RESOLVED. THERE WERE NO REPORTED ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20523485 ↗
  • InjuryMar 29, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE DISTAL MARKER BAND DETACHED FROM THE CATHETER, AND REMAINED IN THE PATIENT. AN EKOSONIC ENDOVASCULAR DEVICE, 106X24CM WAS SELECTED FOR USE IN THE ILIAC. THE CATHETER WAS ADVANCED TO THE TARGET LESION, BUT WAS UNABLE TO PASS DUE TO SEVERE STENOSIS. THE CATHETER WAS ATTEMPTED TO BE WITHDRAWN, AT WHICH POINT THE DISTAL MARKER BAND DETACHED. THE PHYSICIAN ELECTED TO LEAVE THE DEVICE FRAGMENT IN THE PATIENT, AS THERE WAS A LIMITED RISK OF THE MARKER BAND MIGRATING FROM THE SEVERE STENOSIS. THE PROCEDURE WAS ABORTED AT THIS TIME, AND THE PATIENT WAS NOT TREATED. THERE WERE NO FURTHER CONSEQUENCES REPORTED FOR THE PATIENT. THE PATIENT WAS REPORTED TO BE DOING FINE.

    Patient problems: Foreign Body In Patient

    MAUDE report 16639372 ↗

Other Catheter, Continuous Flush devices

All Catheter, Continuous Flush devices →