EkoSonic® 6779-005Recalls & reported problems
Boston Scientific Corporation · The EkoSonic® Endovascular Device consists of a single use Infusion Catheter and an Ultrasonic Core. The device delivers physician-specified fluids and ultrasound to the intravascular treatment site. It comes in different working lengths and treatment zones. · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00858593006165), received between Feb 11, 2020 and Sep 5, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash2
- Adverse Event Without Identified Device or Use Problem1
- Audible Prompt/Feedback Problem1
- Device Alarm System1
- Material Deformation1
- Temperature Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
- Pneumothorax1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 5, 2024Leak/Splash; Material Deformation; Leak/Splash; Material Deformation
IT WAS REPORTED THAT THE DRUG LUER WAS CRACKED AND LEAKING. THIS 106X30CM EKOSONIC ENDOVASCULAR DEVICE WAS SELECTED FOR USE FOR A DEEP VEIN THROMBOSIS POPLITEAL PROCEDURE. THE CATHETER WAS PLACED, AND THE PATIENT WAS BROUGHT TO THE INTENSIVE CARE UNIT FOR TREATMENT. AT THAT TIME, LYTIC THERAPY COULD NOT BE STARTED DUE TO A CRACKED AND LEAKING DRUG LUER. AN ATTEMPT WAS MADE TO REPAIR THE DRUG LUER AND RUN THE CATHETER AS A STANDARD INFUSION CATHETER; HOWEVER, DRUG CONTINUED TO LEAK FROM THE LUER. NO PATIENT COMPLICATIONS WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED BUT WAS NOT AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20146967 ↗MalfunctionMar 26, 2024Leak/Splash; Adverse Event Without Identified Device or Use Problem; Leak/Splash; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THIS DEVICE WAS LEAKING AND THE PATIENT DEVELOPED PNEUMOTHORAX. THIS 106X30CM EKOSONIC ENDOVASCULAR DEVICE WAS SELECTED FOR USE IN A PULMONARY EMBOLISM PROCEDURE. THE CATHETER WAS PLACED, AND THE PATIENT WAS TAKEN TO THE INTENSIVE CARE UNIT FOR TREATMENT OVERNIGHT. LEAKING FROM THE DRUG PORT WAS LATER OBSERVED. PER PHYSICIANS ORDERS, THE COOLANT AND DRUG WERE INFUSED TOGETHER THROUGH THE COOLANT PORT. ADDITIONALLY, IT WAS NOTED THAT THE PATIENT DEVELOPED PNEUMOTHORAX. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE.
Patient problems: Pneumothorax
MAUDE report 18979077 ↗MalfunctionNov 6, 2023Device Alarm System; Temperature Problem; Device Alarm System; Temperature Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 17OCT2023. AN EKOSONIC ENDOVASCULAR DEVICE, 106X30CM WAS SELECTED FOR USE IN A BILATERAL EKOS PROCEDURE. AFTER APPROXIMATELY 90 MINUTES OF EKOS THERAPY, THERE WERE MULTIPLE ERROR MESSAGES OBSERVED ON THE CONTROL UNIT. TROUBLESHOOTING WAS UNABLE TO RESOLVE THE ISSUE. THE EKOS DEVICE WAS RAN AS A STANDARD INFUSION CATHETER WITHOUT THE USE OF ULTRASOUND TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED PATIENT COMPLICATIONS. HOWEVER, DEVICE ANALYSIS REVEALED THE ULTRASONIC CORE (USC) WAS FRACTURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18075173 ↗MalfunctionFeb 11, 2020Audible Prompt/Feedback Problem; Audible Prompt/Feedback Problem
ON (B)(6) 2020, 17 HOURS INTO THERAPY, ICU NURSE CALLED WITH AN ALL MSD GROUPS DISABLED ALARM. THE CATHETER WAS USED TO TREAT A PERIPHERAL ARTERIAL OCCLUSION (PAO). SHE MENTIONED THAT THE BLACK CABLE MAY HAVE BEEN KINKED WHEN THE CONTROL UNIT (CU) ALARMED, BUT THAT IT WAS THEN STRAIGHTENED OUT. TROUBLESHOOTING PERFORMED DID NOT CLEAR THE ALARM. THE NURSE WAS INSTRUCTED TO NOTIFY THE DOCTOR THAT THE CATHETER WOULD NOW RUN AS A BASIC INFUSION CATHETER. THE PATIENT WAS "DOING GREAT". ON 30 JANUARY 2020, INVESTIGATION OF THE RETURNED DEVICE REVEALED THAT THE MICROSONIC DEVICE (MSD) SHORTED DUE TO FLUID INGRESS INTO THE HUB AFTER 17 HOURS OF THERAPY. THIS WAS NOT REPORTED BY THE CUSTOMER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9696617 ↗
Other Catheter, Continuous Flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.