BAROSTIM NEO 1036Recalls & reported problems

Cvrx, Inc. · BAROSTIM NEO Model 1036 Carotid Sinus Lead Kit (US Commercial) · Class 3 · Implant

All models of BAROSTIM NEO (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Barostim NEO carotid sinus nerve stimulator has 76 adverse event reports, including 1 death and 75 injuries. Reports mostly describe nerve damage, Twiddler’s syndrome, swelling, numbness, and neck pain, plus lead issues such as twisted or bent material and high or low impedance. There are no recalls for this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
78
Deaths reported
1
Injuries reported
77
Recalls naming it
0

Summary

The FDA's MAUDE database contains 78 adverse event reports that list this exact device (DI 00859144004463), received between Sep 9, 2022 and Sep 17, 2026 — classified as 1 death, 77 injuries.

The most frequently reported specific device problem is “Material Twisted/Bent” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 5 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 5 reportsMay 2025: 1 reportsJun 2025: 5 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 4 reportsOct 2025: 4 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 4 reportsApr 2026: 3 reportsMay 2026: 6 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem42
  • Material Twisted/Bent11
  • Use of Device Problem9
  • High impedance3
  • Material Split, Cut or Torn3
  • Insufficient Device Problem Information2
  • Low impedance2
  • Unintended Movement2

Patient problems

  • Nerve Damage18
  • Twiddlers Syndrome14
  • No Clinical Signs, Symptoms or Conditions12
  • Numbness9
  • Swelling/ Edema9
  • Neck Pain8
  • Undesired Nerve Stimulation7
  • Facial Paralysis5

Recent adverse event reports

The 25 most recent of 78 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Material Twisted/Bent

    Patient problems: Pain; Twiddlers Syndrome

    MAUDE report 50805739 ↗
  • InjurySep 14, 2026Unintended Movement

    A Barostim system was implanted on (b)(6) 2025 and last titrated on (b)(6) 2025 where therapy amplitude was increased from 6.4mA to 7.0mA. Around (b)(6) 2025, the patient experienced extraneous stimulation in the form of tightness on the right side of their neck which worsened around (b)(6) 2026. The patient was seen for a follow up on (b)(6) 2026 and they were able to replicate the stim by raising their right arm. The patient's therapy was reduced to 5.0mA and the stim resolved. On a later date, the patient continued to experience tightness in their throat. On (b)(6) 2026, the patient's therapy was adjusted to 4.2mA. Labs were ordered as the clinicians believed that the patient's CHF may have worsened. The patient continued to experience issues with stim near their throat and imaging was done. The electrode and the positioning of the device looked good, however the excess lead in the pocket appeared to have migrated underneath the bottom of the IPG which potentially caused a pulling like sensation on the lead. Per the opinion of the physician, the root cause of the migration was possibly due to the patient's bariatric surgery which caused a lot of excess skin and potentially cause

    Patient problems: Neck Pain; Undesired Nerve Stimulation

    MAUDE report 50753367 ↗
  • InjurySep 2, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Implant Pain; Twiddlers Syndrome

    MAUDE report 50571111 ↗
  • InjuryAug 24, 2026Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50420610 ↗
  • InjuryAug 12, 2026Use of Device Problem

    A Barostim system was implanted on (b)(6) 2025. On (b)(6) 2026, impedance was low, and Barostim therapy was turned off. It was noted the system passed compliance as of the last Barostim check on (b)(6) 2026, but the lead impedance had decreased significantly from approximately 850 to 150 Ohms on approximately (b)(6) 2026 and decreased to low around (b)(6) 2026. There were no unusual events or physical stresses on the system reported by the patient, and the patient did not report any symptoms or adverse events. A chest and neck x-ray were ordered and showed the lead completely detached from the IPG header. In the opinion of the physician, the root cause of the lead detachment could not be conclusively determined, but Twiddler Syndrome was possible as the patient reported frequent touching of their IPG pocket. A successful lead revision occurred on 16-Jul-2026. The old lead was extracted, and a new lead was placed on the same carotid artery with a good BP response.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50223996 ↗
  • InjuryAug 11, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Bacterial Infection; Skin Discoloration

    MAUDE report 50213807 ↗
  • InjuryJun 24, 2026Material Twisted/Bent

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2026. AROUND (B)(6) 2026, THE PATIENT EXPERIENCED MOVEMENT WITH THEIR IPG. AN X-RAY WAS DONE. PER THE PHYSICIAN, THE ROOT CAUSE OF THE EVENT WAS THAT THE PATIENT WAS ROTATING THE DEVICE IN THE POCKET. ON (B)(6) 2026, A SUCCESSFUL POCKET REVISION WAS DONE. THE LEAD WAS STRAIGHTENED AND THE DEVICE WAS SECURED IN THE MEDIAL POCKET. IT WAS NOTED THAT THE LEAD AND IMPEDANCE WERE TESTED AND WERE BOTH WITHIN NORMAL RANGE.

    Patient problems: Twiddlers Syndrome

    MAUDE report 25610246 ↗
  • InjuryMay 27, 2026Appropriate Device Problem Term/Code Not Available

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2024 AND LAST TITRATED ON (B)(6) 2024 WHERE THERAPY WAS INCREASED FROM 7.4MA TO 9.0MA. AROUND (B)(6) 2025, THE PATIENT EXPERIENCED PULLING IN THEIR NECK. THE SURGEON NOTED THAT AFTER PALPATING THE LEAD, IT FELT "AS TIGHT AS A GUITAR STRING". AROUND (B)(6) 2026, THE PATIENT HAD A CT DONE. THE CT OF THE CHEST WAS DONE IN A SUPINE POSITION, AND THE LEAD APPEARED TO HAVE SLACK. ADDITIONAL IMAGING WAS BEING DONE, HOWEVER WAS NOT YET AVAILABLE. A SUGGESTION WAS MADE TO ADJUST THE PATIENT'S THERAPY TO RULE OUT STIMULATION. THE SURGEON WAS WORKING WITH THE CARDIOLOGIST TO SCHEDULE THE PATIENT FOR A LEAD REVISION. IT WAS NOTED THAT THE PATIENT HAD A POOR PROGNOSIS DUE TO CANCER DIAGNOSIS AND CHOSE TO HAVE THE DEVICE EXPLANTED AND THEY NEEDED TO PUT A PORT IN AS THEY WERE UNDERGOING CHEMO. PER THE OPINION OF THE PHYSICIAN, THE ROOT CAUSE OF THE EVENT WAS POSSIBLY DUE TO RADIATION THERAPY ON THE LEAD. THE DEVICE WAS EXPLANTED ON (B)(6) 2026.

    Patient problems: Discomfort

    MAUDE report 25311580 ↗
  • InjuryMay 27, 2026Material Twisted/Bent

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025. ON (B)(6) 2026, THE PATIENT PRESENTED TO THE EMERGENCY ROOM WITH SHORTNESS OF BREATH, SWELLING AND ABDOMINAL PAIN AND AN X-RAY WAS DONE ROUTINELY AND SHOWED THE IPG PROJECTING OVER THEIR RIGHT UPPER CHEST, EXTENDING INTO THE RIGHT NECK, WITH SEVERAL TWISTS AND TURNS ALONG THE PATH OF THE LEAD. AROUND 30-MAR-2026, THE PATIENT EXPERIENCED TIGHTNESS IN THEIR NECK WHEN THEY TURN THEIR HEAD TO THE RIGHT AND ON (B)(6) 2026, THEY HAD A VIRTUAL FOLLOW UP WITH THEIR PHYSICIAN. IT WAS SUSPECTED THAT THE ROOT CAUSE WAS TWIDDLERS SYNDROME DUE TO TWISTING AND TURNING OF THE BATTERY/GENERATOR IN ITS POCKET. AN X RAY WAS DONE, REVEALING A TWIST IN THE LEAD. IT WAS NOTED THAT THE LEAD DID NOT APPEAR FRACTURED AND THEY WERE UNSURE IF THE DEVICE WAS WORKING. THE PATIENT WAS SEEN FOR A FOLLOW UP ON (B)(6) 2026 AND THEIR LEAD IMPEDANCE WAS AT 130. ON (B)(6) 2026, A SUCCESSFUL LEAD REVISION WAS DONE. THE FRACTURED LEAD WAS CUT, AND A NEW LEAD WAS PLACED TO THE LEFT CAROTID. PER THE OPINION OF THE PHYSICIAN, IT WAS SUSPECTED THAT THAT THE SOB MAY HAVE BEEN CAUSED BY BAROSTIM NOT DELIVERING THE THERAPY APPROPRIATELY, AND AS OF (B)(6) 2026, THE PATIENT REPOR

    Patient problems: Discomfort; Dyspnea; Swelling/ Edema; Twiddlers Syndrome

    MAUDE report 25306882 ↗
  • InjuryMay 27, 2026Material Twisted/Bent

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2024. THE PATIENT WAS SEEN FOR A DEVICE INTERROGATION ON (B)(6) 2026, AND IT WAS DISCOVERED THAT AROUND ON (B)(6) 2025 THE SYSTEM WAS OUT OF COMPLIANCE AND THE LEAD IMPEDANCE WAS HIGH. IT WAS NOTED THAT THE PATIENT HAD A CHIROPRACTIC ADJUSTMENT THAT CORRESPONDED WITH THE DATE OF THE IMPEDANCE CHANGE. AN X-RAY WAS ORDERED AND THE PATIENT'S THERAPY WAS DEACTIVATED. PER THE OPINION OF THE PHYSICIAN, THE ROOT CAUSE OF THE EVENT WAS THAT THE CSL MIGRATED INTO SUBCUTANEOUS MUSCLE HOWEVER, PER THE X-RAY'S, IT LOOKED LIKE THE ELECTRODE WAS SEPARATING FROM THE LEAD BODY. THEY WERE SCHEDULED FOR A FOLLOW-UP ON (B)(6) 2026, HOWEVER AS OF ON (B)(6) 2026, THERE WERE NO ADDITIONAL UPDATES PROVIDED. IT WAS NOTED THAT THE SURGEON'S OFFICE WAS GOING THROUGH A TRANSITION AND WAS UNRESPONSIVE ON ALL ISSUES. ON (B)(6) 2026, THE PATIENT'S DEVICE WAS EXPLANTED SUCCESSFULLY AND SENT TO PATHOLOGY. AN ATTEMPT WAS MADE TO RETRIEVE THE EXPLANTED DEVICE BUT WAS UNSUCCESSFUL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25306819 ↗
  • InjuryMay 27, 2026Material Twisted/Bent

    NEBRASKA LOW IMPEDANCE- A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025. AROUND (B)(6) 2026, THE PATIENT EXPERIENCED AN INTERMITTENT GURGLING SENSATION AT THE IPG LOCATION. THE PATIENT DENIED MANIPULATING THE DEVICE OR EXPERIENCING ANY TRAUMA THAT MIGHT HAVE IMPACTED THE DEVICE. THEY REPORTED THAT WHEN SLEEPING AT NIGHT, THE DEVICE "POKED OUT". WHILE REVIEWING THE IMPEDANCE TREND, IT WAS NOTED THAT THERE WAS A STEEP DIP IN IMPEDANCE THAT CORRELATED WITH THE DAY THE PATIENT BEGAN TO EXPERIENCE SYMPTOMS. THE GURGLING SENSATION AND LOW IMPEDANCE OF 330 WAS REPLICABLE WHILE THE PATIENT WAS MOVING AROUND. WHEN THE PATIENT WAS SEATED THE IMPEDANCE WENT BACK TO 803. THE PATIENT'S THERAPY WAS REDUCED TO 5MA AND XRAY WAS ORDERED AND REVEALED TWISTS IN THE LEAD. A LEAD AND POCKET REVISION OCCURRED ON (B)(6) 2026. A NEW LEAD WAS IMPLANTED, AND THE DEVICE WAS SECURED IN THE POCKET. AS OF (B)(6) 2026, THE PATIENT'S SYMPTOMS WERE RESOLVED AND THE IMPEDANCE WAS NORMAL.

    Patient problems: Twiddlers Syndrome; Undesired Nerve Stimulation

    MAUDE report 25300149 ↗
  • InjuryMay 19, 2026Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2026. FOLLOWING THE PROCEDURE, THE PATIENT EXPERIENCED NUMBNESS AND A SWOLLEN TONGUE AND ON (B)(6) 2026, THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT. A STROKE WAS RULED OUT AT THE ED, AND THE PATIENT WAS SENT HOME. ON (B)(6) 2026 AT A SCHEDULED TITRATION FOLLOW-UP, THE PATIENT REPORTED THAT THEY EXPERIENCED A SWOLLEN TONGUE, TONGUE DEVIATION, DIFFICULTY SWALLOWING AND SPEAKING, RIGHT LEG SPASMS, INTERMITTENT HEADACHES, PINS AND NEEDLES SENSATION ON THE RIGHT SIDE OF THEIR FACE AND NECK, NUMBNESS ON THE RIGHT SIDE OF THEIR FACE AND NECK, A TEARY RIGHT EYE AND BLURRY VISION IN THEIR RIGHT EYE. PER THE OPINION OF THE PHYSICIAN, THE ROOT CAUSE OF THE TONGUE MOVEMENT ISSUES WAS LIKELY CAUSED BY BAROSTIM DUE TO THE PROXIMITY OF THE HYPOGLOSSAL NERVE, HOWEVER, WAS UNSURE IF THE OTHER ISSUES WERE BEING CAUSED OR CONTRIBUTED TO BY BAROSTIM. THE PHYSICIAN OBSERVED A HARD LUMP UNDER THE INCISION, AT THE SITE OF THE ELECTRODE AND HYPOTHESIZED THAT IT MIGHT BE CAUSING PRESSURE ON OTHER SURROUNDING NERVES TO CREATE THESE OTHER EXTRANEOUS COMPLAINTS. THE PATIENT'S BAROSTIM THERAPY WAS TURNED OFF TO ALLOW MORE TIME FOR HEALING, AND IT WAS NOTED THAT

    Patient problems: Excessive Tear Production; Headache; Nerve Damage; Numbness; Swelling/ Edema

    MAUDE report 25218548 ↗
  • InjuryMay 19, 2026Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2026. ON (B)(6) 2026, FOLLOWING A LONG DAY OF INCREASED ACTIVITY, THE PATIENT EXPERIENCED REDNESS, TENDERNESS, WARMTH AND SWELLING NEAR AND ABOVE THE CHEST INCISION AS WELL AS CHEST DISCOMFORT. ON (B)(6) 2026 THEY PRESENTED TO THE EMERGENCY ROOM WHERE THEY WERE ADMITTED. LABS AND AN EKG WERE DONE AND INTRAVENOUS ANTIBIOTICS WERE ADMINISTERED. IT WAS REPORTED THAT CELLULITIS WAS FOUND ON THE LEAD ATTACHED TO THE RIGHT CAROTID AND ON (B)(6) 2026, THE LEAD WAS EXPLANTED. PER THE OPINION OF THE PHYSICIAN, THE ROOT CAUSE OF THE INFECTION WAS UNKNOWN. THE PATIENT WAS DISCHARGED ON (B)(6) 2026, THE INFECTION RESOLVED.

    Patient problems: Bacterial Infection; Discomfort; Swelling/ Edema

    MAUDE report 25218238 ↗
  • InjuryApr 25, 2026Material Twisted/Bent; Unintended Movement

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025. AROUND ON (B)(6) 2026, THE PATIENT'S LEAD IMPEDANCE DECREASED RAPIDLY. IT WAS NOTED THAT AROUND THAT TIME, THE DEVICE TURNED WHILE THE PATIENT WAS WAKING UP AND THE PATIENT MANUALLY TURNED IT BACK. THE DEVICE WAS TESTED AND CONSTANTLY PASSED COMPLIANCE AND STIM WAS EXPERIENCED ONCE. THERAPY WAS STILL BEING DELIVERED, HOWEVER THE PATIENT REPORTED BEING MORE TIRED IN A TIMEFRAME THAT COINCIDES WITH THE IMPEDANCE DROP. AN X-RAY WAS ORDERED AND THE IMAGES REVEALED THAT THE IPG DROPPED INFERIOR TO THE CHEST WALL, AS WELL AS MULTIPLE COILS IN THE LEAD NEAR THE HEADER OF THE IPG. A LEAD REVISION OCCURRED ON (B)(6) 2026. THE DAMAGED LEAD AND THE IPG WERE EXPLANTED AND A NEW GENERATOR AND LEAD WERE IMPLANTED.

    Patient problems: Fatigue; Twiddlers Syndrome; Undesired Nerve Stimulation

    MAUDE report 24991398 ↗
  • InjuryApr 21, 2026Loss of or Failure to Bond

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2022. DURING A NORMAL IPG REPLACEMENT ON (B)(6) 2026, HIGH IMPEDANCE WAS NOTICED. THE CSL WAS REMOVED FROM THE IPG, AND THE CSL PIN HAD PULLED OUT FROM THE RING. IT WAS UNKNOWN IF THE PHYSICIAN HAD FULLY LOOSENED BOTH SET SCREWS AS THEY WERE MOVING FAST, BUT THE PHYSICIAN STATED THEY THOUGHT THERE WAS A DEFECT IN THE CONNECTION. THE CSL WAS CAPPED, AND NO NEW IPG WAS IMPLANTED. THE PROCEDURE WAS NOT DELAYED BY GREATER THAN 30 MINUTES DUE TO THE ISSUE, AND THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS. A CSL EXPLANT AND BAROSTIM REIMPLANT OCCURRED ON (B)(6) 2026. A NEW IPG WAS PLACED, AND A NEW CSL PLACED ON THE OPPOSITE CAROTID. THE INITIAL CSL WAS LEFT IMPLANTED AND CAPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24944530 ↗
  • InjuryApr 15, 2026High impedance; Unraveled Material; Detachment of Device or Device Component

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2026. ON (B)(6) 2026, IT WAS NOTED THAT THE LEAD IMPEDANCE WAS HIGH AND OUT OF COMPLIANCE. THE PHYSICIAN ORDERED X RAYS AND A FOLLOW UP WAS SCHEDULED FOR (B)(6) 2026. THE PATIENT'S FOLLOW UP WAS CANCELLED AS THEY WERE ADMITTED FOR ABDOMINAL PAIN UNRELATED TO BAROSTIM. IT WAS NOTED THAT THE IPG SET SCREWS LOOSENED AND THE CSL ADVANCED FURTHER. ON (B)(6) 2026, A LEAD REVISION WAS DONE, AND THE IMPEDANCE WAS IN COMPLIANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24888316 ↗
  • InjuryMar 20, 2026Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025. ON (B)(6) 2026, DURING A PROCEDURE FOR A DIFFERENT IMPLANTABLE DEVICE, THE SURGEONS EXPERIENCED DIFFICULTY DUE TO THE BAROSTIM LEAD THAT WAS ALREADY IMPLANTED. IT WAS NOTED THAT THE VENOUS ACCESS COULD NOT PASS THE STRAIN RELIEF TAB. A FEMORAL ACCESS METHOD WAS USED AND THERE WERE NO FURTHER ISSUES. THE PROCEDURE WAS DELAYED AS A RESULT OF THE DIFFICULTY WITH THE BAROSTIM LEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24654561 ↗
  • InjuryMar 18, 2026Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON 18-DEC-2025. SINCE IMPLANT, THE PATIENT EXPERIENCED RIGHT FACIAL DROOPING, NUMBNESS AND SWELLING IN THE RIGHT FACE AND NECK, AND PAIN IN THEIR HEAD, NECK, AND THROAT. THE PHYSICIAN STATED A NECK NERVE WAS MANIPULATED DURING THE IMPLANT PROCEDURE, DROOPING MAY RESOLVE IN 6-12 MONTHS, AND NO INTERVENTION WAS PLANNED FOR THE DROOPING. AS OF (b)(6) 2026, THE NERVE DAMAGE WAS SLIGHTLY IMPROVED, AND A FOLLOW-UP APPOINTMENT WAS SCHEDULED FOR (b)(6) 2026.

    Patient problems: Nerve Damage; Numbness; Pain; Swelling/ Edema

    MAUDE report 24626241 ↗
  • InjuryMar 16, 2026Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2026. AROUND (B)(6) 2026, THE PATIENT EXPERIENCED A HEMATOMA AT THE NECK INCISION SITE AND PRESENTED TO THE HOSPITAL WHERE THEY WERE ADMITTED. THE HEMATOMA WAS SUCCESSFULLY SURGICALLY EVACUATED. PER THE OPINION OF THE PHYSICIAN, THE HEMATOMA WAS POSSIBLY CAUSED DURING CLOSURE SUTURING OF THE NECK INCISION, IN ADDITION TO THE PATIENT BEING ON BLOOD THINNERS. THE PATIENT WAS DISCHARGED ON (B)(6) 2026.

    Patient problems: Hematoma

    MAUDE report 24612574 ↗
  • InjuryMar 13, 2026Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025. FOLLOWING THE PROCEDURE, THE PATIENT EXPERIENCED LIP DROOPING. THERE WAS NO PLAN FOR FURTHER INTERVENTION, AND THE PATIENT WOULD NOT RETURN TO SEE THEIR SURGEON AND THE CARDIOLOGIST. THE PATIENT REPORTED THAT THEY HAD TOO MUCH GOING ON TO DEAL WITH THE EVENT AT THIS TIME AND WAS OK WITH ALLOWING MORE TIME TO CONTINUE TO SEE IF THE LIP DROOP RESOLVED.

    Patient problems: Facial Paralysis; Nerve Damage

    MAUDE report 24591756 ↗
  • InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025. FOLLOWING IMPLANT, THE PATIENT EXPERIENCED DIFFICULTY SWALLOWING AND SWELLING AT THE NECK INCISION. THE CERTIFIED PHYSICIAN ASSISTANT RULED OUT AN INFECTION AS THE PATIENT'S WHITE BLOOD CELL COUNT WAS NOT ELEVATED. A CT WAS PERFORMED AND REVEALED FLUID AND GAS COLLECTION ALONG THE LEFT NECK STIMULATOR LEAD AND ANTERIOR TO LEFT CAROTID SPACE. PER THE OPINION OF THE PHYSICIAN, BASED ON THE IMAGING RESULTS THE ROOT CAUSE OF THE EVENT WAS RELATED TO ABSCESS. THE PATIENT WAS PUT ON ANTIBIOTICS, AND AN ASPIRATION OF THE ABSCESS WAS PERFORMED. THE PATIENT WAS SEEN FOR A FOLLOW UP ON (B)(6) 2025 AND IT WAS CONFIRMED THAT THEY WERE DOING WELL. THE PATIENT REPORTED THAT THEY WERE HOSPITALIZED FOR A HIGH FEVER AND CONFUSION ON (B)(6) 2025 UNRELATED TO BAROSTIM. THE PATIENT WAS SEEN FOR A FOLLOW UP ON (B)(6) 2026 AND THEY REPORTED THAT THE DIFFICULTY SWALLOWING AND DIFFICULTY OF PRESSING THEIR LIPS TOGETHER RESOLVED AND THE SWELLING WAS REDUCED. AS OF (B)(6) 2026, IT WAS CONFIRMED THAT THE PATIENT WAS DOING WELL.

    Patient problems: Swelling/ Edema

    MAUDE report 24326559 ↗
  • InjuryFeb 12, 2026Use of Device Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025 AND THE PATIENT LEAD IMPEDANCE WAS 1300. ON (B)(6) 2025, THE PATIENT'S LEAD IMPEDANCE WAS 3900 AND AS OF (B)(6) 2025, REMAINED STABLE. IT WAS SUSPECTED THAT THERE WAS AN ISSUE WITH THE CONNECTION BETWEEN THE LEAD PIN AND SET PIN IN THE HEADER OF THE IPG AND THE ROOT CAUSE OF THE EVENT WAS BEING FURTHER INVESTIGATED. THE PATIENT WAS SEEN FOR A FOLLOW UP ON 10-DEC-2025 AND IT WAS NOTED THAT THE IMPEDANCE WAS STILL HIGH AND THE DEVICE WAS NOT IN COMPLIANCE. AN X-RAY REVEALED AND CONFIRMED THE CONNECTION ISSUE BETWEEN THE LEAD AND THE HEADER AS THE LEAD PIN WAS NOT VISIBLE PAST THE DISTAL BLOCK IN THE HEADER. A LEAD REVISION OCCURRED ON (B)(6) 2026 AND THE IMPEDANCE ISSUE RESOLVED. THE PATIENT WAS SEEN FOR A FOLLOW UP ON 27-JAN-2026 AND IT WAS CONFIRMED THAT THE LEAD IMPEDANCE TREND RETURNED TO NORMAL SINCE THE REVISION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24324524 ↗
  • InjuryDec 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025. SINCE IMPLANT, THE PATIENT EXPERIENCED WORSENING VOCAL ISSUES IN THE FORM OF PARTIAL VOCAL PARALYSIS AND INCREASED DYSPNEA ON EXERTION (DOE) AND SHORTNESS OF BREATH (SOB). IN THE OPINION OF THE PHYSICIAN, THE ROOT CAUSE OF THE PARTIAL VOCAL PARALYSIS WAS UNKNOWN BUT POSSIBLY RELATED TO NERVE DAMAGE OR VOCAL CORD TRAUMA DURING BAROSTIM IMPLANT AS THE ONSET WAS POST-BAROSTIM IMPLANT. IN THE OPINION OF THE PHYSICIAN, THE DOE AND SOB WERE NOT DUE TO BAROSTIM AS THE PATIENT HAD RECENTLY EXPERIENCED COVID. A BAROSTIM FOLLOW-UP WAS SCHEDULED FOR (B)(6) 2025. THE PATIENT WAS REFERRED TO PHYSICAL THERAPY FOR THEIR VOICE AND VOCAL CORDS, AND PHYSICAL THERAPY WAS SCHEDULED TO BEGIN IN MID-(B)(6). AS OF (B)(6) 2025, NO CHANGE TO THE VOCAL ISSUES HAD OCCURRED.

    Patient problems: Paralysis; Voice alteration

    MAUDE report 23790210 ↗
  • InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025 AND LAST TITRATED FROM 3.0 TO 4.0MA ON (B)(6) 2025. SINCE IMPLANT, THE PATIENT EXPERIENCED AN OVERACTIVE SALIVARY GLAND AND LIP NUMBNESS ON THE RIGHT SIDE OF THEIR MOUTH CAUSING A LISP. THE SURGEON EXPECTED THE SYMPTOMS TO RESOLVE; NO INTERVENTION WAS PLANNED. BAROSTIM THERAPY WAS INCREASED TO 6.0MA ON (B)(6) 2025.

    Patient problems: Numbness; Salivary Hypersecretion; Voice alteration

    MAUDE report 23649481 ↗
  • InjuryNov 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A BAROSTIM SYSTEM WAS IMPLANTED ON (B)(6) 2025 WITH THE IPG AND CSL PLACED ON THE RIGHT SIDE OF THE PATIENT'S BODY. SINCE IMPLANT, THE PATIENT EXPERIENCED LEFT-SIDED FACIAL DROOPING WHEN SMILING. AS OF (B)(6) 2025, IT WAS NOTED THE DROOPING MAY HAVE SLIGHTLY IMPROVED. A FOLLOW-UP WAS PLANNED FOR (B)(6) 2025; HOWEVER, THE APPOINTMENT WAS CANCELLED. IN THE OPINION OF THE PHYSICIAN, THE SYMPTOMS COULD HAVE BEEN CAUSED BY THE INTUBATION AND SHOULD RESOLVE OVER 4-6 MONTHS. AS OF (B)(6) 2025, THERE HAD NOT BEEN MUCH CHANGE IN THE PATIENT'S STATUS. A FOLLOW-UP WAS PLANNED FOR (B)(6) 2025.

    Patient problems: Facial Paralysis

    MAUDE report 23638800 ↗

Other Stimulator, Carotid Sinus Nerve devices