Skip to content

EluPro Antibiotic-Eluting BioEnvelope CMCV-124-MEDRecalls & reported problems

Elutia Med LLC · The EluPro Antibiotic-Eluting BioEnvelope is a sterile pouch designed to hold a cardiac implantable device or implantable neurostimulator device to create a stable environment when implanted in the body. EluPro consists of decellularized, non-crosslinked, lyophilized extracellular matrix (ECM) and resorbable, ring- shaped poly(D,L Lactide-Co-Glycolide) (PLGA) discs containing the antibiotics rifampin and minocycline. The ECM material is derived from porcine small intestinal submucosa (SIS). The envelope is constructed with four multilaminate sheets, perforated to allow for the natural flow of exudate through the device. The drug-eluting polymer discs are secured between the multilaminate sheets on each side and the envelope is constructed with 5-0 polydioxanone (PDS) suture. EluPro contains 10.5 mg of rifampin and 9.3 mg of minocycline. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00859389005331), received between Aug 28, 2025 and Jun 24, 2026 — classified as 2 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Seroma1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 24, 2026Adverse Event Without Identified Device or Use Problem

    NOTIFICATION WAS RECEIVED FROM BOSTON SCIENTIFIC SALES REPRESENTATIVE THAT A (B)(6) YR. OLD MALE PATIENT WAS IMPLANTED WITH AN ABBOTT DUAL CHAMBER CIED ON (B)(6) 2026 USING AN ELUPRO ANTIBIOTIC-ELUTING BIOENVELOPE (CMCV-124-MED; LOT #M26C1093). IT WAS REPORTED THAT THE PATIENT HAD DEVELOPED A LARGE DRAINING SEROMA (SPECIFIC DATE IS UNKNOWN) THAT REQUIRED THE PHYSICIAN TO PERFORM A POCKET EVACUATION - REMOVAL OF THE ELUPRO BIOENVELOPE AND CLOT(S), ON (B)(6) 2026. SURROUNDING TISSUE WAS DESCRIBED AS RED WITH POSSIBLE GRANULATION, CULTURES WERE TAKEN AT THE TIME OF THE POCKET EVACUATION BUT WERE NEGATIVE FOR ANY BACTERIA. THE ELUPRO BIOENVELOPE WASN'T COMPLETELY RESORBED AS PHYSICIAN REMOVED REMNANTS OF THE BIOENVELOPE AND ANTIBIOTIC DISC. THE PHYSICIAN STATES THAT THE REPORTED SEROMA IS RELATED TO THE ELUPRO BIOENVELOPE AND NOT PROCEDURE RELATED. NO FURTHER DETAILS ARE AVAILABLE REGARDING THE REPORTED EVENT, SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

    Patient problems: Seroma

    MAUDE report 25610885 ↗
  • InjuryAug 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    EMAIL WAS RECEIVED FROM AN ELUTIA SALES REPRESENTATIVE REGARDING A PATIENT ADVERSE EVENT. IT WAS REPORTED THAT, THE PATIENT HAD A POCKET REVISION WITH A CRTD BIV PACEMAKER IMPLANT ON (B)(6) 2025 ALONG WITH THE IMPLANT OF A MEDIUM ELUPRO ANTIBIOTIC-ELUTING BIOENVELOPE (MODEL - CMCV-124-MED, LOT NUMBER - UNKNOWN). THE SALES REPRESENTATIVE NOTED THAT THE PATIENT HAD AN INFECTION ON (B)(6) 2025. THE DEVICE WAS REMOVED ON (B)(6) 2025 AND A NEW DEVICE WAS IMPLANTED ON (B)(6) 2025. ON (B)(6) 2025, THE PHYSICIAN CONFIRMED THAT THE ELUPRO DEVICE WAS NOT THE CAUSE OF THE INFECTION. NO FURTHER INFORMATION REGARDING THIS EVENT IWAS PROVIDED.

    Patient problems: Unspecified Infection

    MAUDE report 22911269 ↗

Other Mesh, Surgical devices

All Mesh, Surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page