Skip to content

BiowaveHOME BWHD01Recalls & reported problems

Biowave Corporation · BiowaveHOME Neurostimulator · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00859911006072), received between Apr 9, 2024 and Jun 23, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Circuit Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Circuit Failure1
  • Inadequate or Insufficient Training1
  • Insufficient Device Problem Information1
  • Output above Specifications1
  • Use of Incorrect Control/Treatment Settings1

Patient problems

  • Electric Shock2
  • Hypoesthesia1
  • Myalgia1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 23, 2026Insufficient Device Problem Information

    A VETERAN PATIENT WAS USING THE BIOWAVEHOME FOR TREATMENT AT THE BILATERAL LATERAL CERVICAL MUSCLES (C6/C7) AT HER USUAL INTENSITY OF 12-16%. THE PATIENT WAS LYING DOWN AND NOT MOVING DURING TREATMENT. SKIN PREPARATION WAS PERFORMED USING A BABY WIPE PRIOR TO ELECTRODE PLACEMENT. UPON TURNING THE DEVICE ON, THE DISPLAY SHOWED AN ABNORMAL READOUT OF "7 MINUTES." THE DEVICE THEN DELIVERED AN UNINTENDED ELECTRICAL OUTPUT OF SUFFICIENT MAGNITUDE TO PHYSICALLY DISPLACE THE PATIENT FROM A LYING POSITION. THE PATIENT'S HUSBAND WAS PRESENT AND WAS REQUIRED TO PHYSICALLY DISCONNECT THE LEAD WIRE FROM THE UNIT TO TERMINATE THERAPY, AS THE POWER BUTTON WAS NON-FUNCTIONAL. FOLLOWING THE EVENT, THE DEVICE DISPLAY SHOWED A BLUE SCREEN AND THE UNIT COULD NOT BE POWERED OFF VIA THE POWER BUTTON. THE PATIENT REPORTED BEING VERY SORE THE FOLLOWING DAY. NO MEDICAL ATTENTION WAS SOUGHT. NO VISIBLE INJURY OR SKIN MARKS WERE OBSERVED.

    Patient problems: Myalgia

    MAUDE report 25600033 ↗
  • MalfunctionApr 30, 2025Output above Specifications; Output above Specifications

    THE VETS SON CALLED AND INFORMED SHE KEEPS GETTING SHOCKED BY UNIT. SHE GETS SHOCKED WHEN TURNING ON THE UNIT. SHE JUST STARTED USING UNIT (B)(6) 2025. NOW IT KEEPS SHOCKING HER UNTIL SHE INCREASES THE INTENSITY THEN IT STOPS. NO INJURY. SHE IS CLEANING SKIN WITH ONLY WATER SO FAR. SHE HAS GOT PAIN RELIEF BUT CONCERNED ABOUT THE SHOCKING FEELING.

    Patient problems: Electric Shock

    MAUDE report 21938639 ↗
  • MalfunctionOct 11, 2024Circuit Failure; Circuit Failure

    UNIT WAS USED IN SALES DEMO. FROM SALES REP: I PLACED THE ELECTRODES ON HIS WRIST AREA, AND WHEN IT WAS TURNED ON AT 0.0% HE SAID HE FELT A JOLT. WE THEN TRIED THEM AT THE BASE OF HIS NECK AND WHEN IT WAS RAISED TO ABOUT 11%, HE FELT A STRONG SHOCK AND IMMEDIATELY WANTED TO TAKE THEM OFF. I TOOK THE UNIT HOME AND DID SOME TROUBLESHOOTING - TRIED DIFFERENT ELECTRODES ON THE SAME MACHINE, TRIED THE SAME ELECTRODES ON A DIFFERENT MACHINE, ETC. - THE PROBLEM IS DEFINITELY THE MACHINE. I HAD THE ELECTRODES AT THE BASE OF MY NECK, AND WHEN I TURNED ON THE MACHINE AT 0.0% I FELT A SLIGHT JOLT, JUST AS THE VOLUNTEER HAD. I THEN TURNED IT UP TO AROUND 30%, AND THEN FELT A SUDDEN SHOCK THAT WENT ALL THE WAY FROM MY NECK DOWN TO MY FINGERTIPS, SO I THINK THAT'S WHAT HE HAD EXPERIENCED AS WELL.

    Patient problems: Electric Shock

    MAUDE report 20429393 ↗
  • InjuryApr 9, 2024Use of Incorrect Control/Treatment Settings; Inadequate or Insufficient Training; Use of Incorrect Control/Treatment Settings; Inadequate or Insufficient Training

    WHEN PATIENT TURNS ON THE DEVICE , UNIT IS BLINKING AND BEEPING AND SHOCKING AT THE SAME TIME PT WANTED TO PRESS MINUS BUTTON OR POWER BUTTON NONE OF THEM WORKED SO PT JUST PEELED ELECTRODES FROM ARM. INJURY OCCURRED BUT PATIENT DID NOT SEEK ANY MEDICAL ATTENTION. PT DOESN'T HAVE ANY BLISTERS, BUT PT COULDN'T MOVE ARM FOR 3 HOURS. IN THE FOLLOW UP THE P[ATIENT MENTIONED THAT RIGH ARM ISNT FUNCTIONING PROPERLY, NO DAMAGE TO THE SKIN BUT CANT LIFT HER ARM OVER HEAD OR BEHIND THE BACK.

    Patient problems: Hypoesthesia

    MAUDE report 19075352 ↗

Other Stimulator, Nerve, Transcutaneous, For Pain Relief devices

All Stimulator, Nerve, Transcutaneous, For Pain Relief devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page