Skip to content

Thoraguard FGN-1387Recalls & reported problems

Centese, Inc. · Thoraguard Control Module · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00860000436427), received between Aug 9, 2025 and Aug 9, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Suction Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Suction Failure1

Patient problems

  • Pneumothorax1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 9, 2025Suction Failure; Suction Failure

    ACCORDING TO THE USER FACILITY'S REPORT (MDR MW5171897), ON (B)(6) 2025, A PATIENT UNDERWENT THORACIC SURGERY AND POSTOPERATIVELY HAD A CHEST TUBE CONNECTED TO A THORAGUARD DIGITAL CHEST DRAINAGE SYSTEM. ON POST-OPERATIVE DAY 1, A SCHEDULED MORNING CHEST X-RAY SHOWED A LARGE LEFT-SIDED PNEUMOTHORAX. THE PATIENT WAS ASYMPTOMATIC AND WELL-APPEARING AT THE TIME OF DISCOVERY. THE ATTENDING SURGEON DISCONNECTED THE CHEST TUBE FROM THE THORAGUARD SYSTEM AND CONNECTED IT TO AN ANALOG CHEST DRAIN, WHICH RELEASED AIR FROM THE LEFT SIDE. THROUGHOUT THE EXCHANGE, THE THORAGUARD DEVICE DID NOT DETECT AN AIR LEAK, AND AFTER REMOVAL FROM THE PATIENT'S CHEST TUBE, THE SYSTEM STILL DID NOT REGISTER AN AIR LEAK. THE PATIENT REMAINED ASYMPTOMATIC FOLLOWING THE INTERVENTION.

    Patient problems: Pneumothorax

    MAUDE report 22766654 ↗

Other Pump, Portable, Aspiration (Manual Or Powered) devices

All Pump, Portable, Aspiration (Manual Or Powered) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page