Skip to content

PuraSinus 5mL 651-009Recalls & reported problems

3-D Matrix Inc. · PuraSinus Gel is a space-occupying gel stent intended to be used in nasal surgery (5 mL) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 2, 2019Class IITerminated

    The boxes used by the shipper were not validated for use during transport.

    Z-0744-2022 · Root cause: Packaging process control · Recall details · FDA record ↗

Other Polymer, Ear, Nose And Throat, Synthetic, Absorbable devices

All Polymer, Ear, Nose And Throat, Synthetic, Absorbable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page