Skip to content

PortableConnect 10100Recalls & reported problems

Rom Technologies, Inc · The PortableConnect System is a rehabilitative exercise therapy device to increase range of motion · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00860002828718), received between Jan 23, 2026 and Apr 17, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem2
  • Application Program Problem: Parameter Calculation Error1
  • Computer Software Problem1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Atrial Fibrillation1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2026Computer Software Problem; Application Program Problem: Parameter Calculation Error

    PATIENT WAS ABLE TO START REHAB ON HER OWN WITHOUT A CARDIAC REHAB SPECIALIST CALL. PATIENT WAS ABLE TO PEDAL FOR 19 MINUTES UNSUPERVISED. CARDIAC REHAB SPECIALIST CALLED AND CONNECTED WITH PATIENT, HE STOPPED EXERCISE AS SHE REPORTED 9/10 SHORTNESS OF BREATH. SHE WAS ALLOWED TO SIT AND REST FOR A BIT. SYMPTOMS DID NOT IMPROVE. PATIENT'S PHYSICIAN WAS CALLED AT 4:18PM CST AND WAS LET KNOW OF THE SITUATION, HE RECOMMENDED THAT EMS BE SENT TO THE PATIENTS HOME. THE PATIENT DID NOT RECEIVE TREATMENT FOR ANY LIFE-THREATENING CONDITION AND ANY IMPAIRMENT WAS PREVENTED BASED ON PATIENT'S RELEASE FROM THE HOSPITAL WITHOUT ADMISSION. THIS WAS THE RESULT OF A SOFTWARE BUG THAT ALLOWED THE FIELD "NO_SESSION_ALLOWED_WITHOUT_CALL" BEING SET TO NULL. THE ISSUE WAS FIXED THE SAME DAY AS THE INCIDENT WITH A SOFTWARE UPDATE, AND ADDITIONAL PROTECTIONS ARE BEING ADDED AT THE DATABASE, API, TESTING, AND MONITORING LEVELS.

    Patient problems: Atrial Fibrillation

    MAUDE report 24913522 ↗
  • MalfunctionMar 26, 2026Material Integrity Problem

    POWER CORD WRAPPED AROUND THE PEDALS AND IS NOW FRAYED, EXPOSING THE CONDUCTORS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24705009 ↗
  • MalfunctionMar 26, 2026Material Integrity Problem

    THE POWER CORD GOT TANGLED IN THE PEDALS AND WAS BROKEN, EXPOSING CONDUCTORS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24703899 ↗
  • MalfunctionJan 23, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THE POWER CORD WRAPPED AROUND THE PEDAL AND BROKE, EXPOSING CONDUCTORS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24168025 ↗

Other System, Isokinetic Testing And Evaluation devices

All System, Isokinetic Testing And Evaluation devices →

Sources and updates