Pronova-O2 1Recalls & reported problems
Turn Medical, LLC · Pronova-O2 is a kinetic bed to prevent pressure injuries and to assist in the management of patients experiencing pulmonary distress and/or complications due in part to patient immobility. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00860006962203), received between Dec 9, 2025 and Jun 15, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Electrical Overstress” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Electrical Overstress1
- Mechanical Jam1
- Unintended Movement1
Patient problems
- Discomfort1
- Muscle/Tendon Damage1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 15, 2026Electrical Overstress; Mechanical Jam
TURN MEDICAL IS SUBMITTING THIS REPORT OUT OF AN ABUNDANCE OF CAUTION. ON (B)(6) 2026, DURING USE OF THE PRONOVA O2 THERAPY SYSTEM, AN ERROR CODE INDICATING THAT THE PATIENT FRAME ROTATION HAD STALLED WAS DISPLAYED. PER THE ON SCREEN INSTRUCTIONS, THE USER CONTACTED TURN MEDICAL. A TURN MEDICAL EMPLOYEE VISITED THE FACILITY THE SAME DAY TO EVALUATE THE ISSUE. A COMPLETE SYSTEM SHUTDOWN WAS PERFORMED; HOWEVER, UPON RESTART, AN ADDITIONAL ERROR CODE WAS GENERATED. FACILITY STAFF WERE ADVISED TO EXCHANGE THE BED, AND A REPLACEMENT SYSTEM WAS PROVIDED ON (B)(6) 2026. THE PATIENT WAS TRANSFERRED TO THE NEW BED, AND THERAPY WAS RESUMED WITHOUT COMPLICATION. THE PATIENT TOLERATED THE TRANSFER WITHOUT DECOMPENSATION OR CLINICAL CONCERN, AND NO STAFF SAFETY ISSUES WERE REPORTED. FOLLOWING THE TRANSFER, ADDITIONAL IN-SERVICE TRAINING WAS PROVIDED TO DAY SHIFT AND NIGHT SHIFT STAFF OR (B)(6) 2026, WITH FURTHER TRAINING SCHEDULED FOR (B)(6) 2026. DURING EVALUATION OF THIS INCIDENT, TURN MEDICAL DID NOT IDENTIFY ANY DEATH, INJURY, OR DELAY IN THERAPY. THIS REPORT IS BEING SUBMITTED FOR THE IDENTIFIED MALFUNCTION IN THE EVENT THAT IT WERE TO RECUR. THIS INCIDENT WILL BE TRACKED FOR ANY INCREASE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25504493 ↗InjuryDec 9, 2025Adverse Event Without Identified Device or Use Problem; Unintended Movement; Adverse Event Without Identified Device or Use Problem; Unintended Movement
TURN MEDICAL IS SUBMITTING THIS REPORT REGARDING AN ALLEGED INJURY ASSOCIATED WITH THE USE OF THE PRONOVA-O2 THERAPY SYSTEM. ON (B)(6) 2025, DURING A ROUTINE CALL WITH FACILITY MANAGEMENT, IT WAS DISCLOSED THAT THE INCIDENT HAD OCCURRED ON (B)(6) 2025. ACCORDING TO THE REPORTING NURSE MANAGER, THE EMPLOYEE WAS ASSISTING IN POSITIONING A PATIENT INTO SUPINATION WHEN THE HATCH AT THE UPPER END OF THE DEVICE UNEXPECTEDLY SPRANG BACK, STRIKING THE EMPLOYEE'S RIGHT WRIST AND RESULTING IN A SPRAIN. THE EMPLOYEE WAS SUBSEQUENTLY PLACED ON WORKERS' COMPENSATION AND REFERRED TO OCCUPATIONAL MEDICINE. AS OF (B)(6) 2025, THE NURSE MANAGER REPORTED THAT THE EMPLOYEE IS "OK" AND CURRENTLY WORKING UNDER LIGHT DUTY RESTRICTIONS. AN INVESTIGATION INTO THE INCIDENT DID NOT IDENTIFY ANY CONTRIBUTING FACTORS, NOR WAS ANY DEFICIENCY IN THE TURN MEDICAL DEVICE. TURN MEDICAL CONFIRMED THAT THE IFU PROVIDES APPROPRIATE INSTRUCTION FOR OPENING AND CLOSING THE PRONE PANELS.
Patient problems: Discomfort; Muscle/Tendon Damage
MAUDE report 23758758 ↗
Other Bed, Patient Rotation, Powered devices
- Isolibrium 2972Stryker Corporation · BED, PATIENT ROTATION, POWERED1 recall17 reports
- Isolibrium 297300000000Stryker Corporation · BED, PATIENT ROTATION, POWERED25 reports
- Isolibrium 2971Stryker Corporation · BED, PATIENT ROTATION, POWERED11 reports
- Isolibrium 2972Stryker Corporation · BED, PATIENT ROTATION, POWERED1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page