PhaseOne Antimicrobial Solution 15050Recalls & reported problems
Sonoma Pharmaceuticals, Inc. · Class U
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Apr 27, 2026Open, Classified
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
Z-2275-2026 · Root cause: Component design/selection · FDA record ↗
Other Dressing, Wound, Drug devices
- MediHoney® 31815Acell, Inc. · Dressing, wound, drug2 recalls7 reports
- MediHoney® 31515Acell, Inc. · Dressing, wound, drug2 recalls6 reports
- MediHoney® 31805Acell, Inc. · Dressing, wound, drug2 recalls1 death5 reports
- MediHoney® 31535Acell, Inc. · Dressing, wound, drug2 recalls0 reports
- AQUACEL AG FOAM 420681Convatec, Purchasing Department · Dressing, Wound, Drug99 reports
- SteriCare Solutions 6240Nurse Assist, Inc. · Dressing, Wound, Drug1 recall14 reports
- Medline DYND500MLSMedline Industries, Inc. · Dressing, wound, drug1 recall5 reports
- SteriCare Solutions 6270Nurse Assist, Inc. · Dressing, Wound, Drug1 recall5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.