Versipass Suture Passer Needle 4727Recalls & reported problems
Lincotek Medical LLC · Suture passer needle for the Versipass Suture Passer · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00860794002679), received between Apr 23, 2026 and Jun 8, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation1
- Material Integrity Problem1
Patient problems
- Laceration(s)1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 8, 2026Material Fragmentation
THE COMPLAINT WAS REPORTED FROM STRYKER INVOLVING THE VERSIPASS NEEDLE MANUFACTURED BY LINCOTEK MEDICAL LLC. THE EVENT TOOK PLACE IN THE UNITED STATES. DURING A SURGICAL PROCEDURE ON (B)(6) 2026, TWO VERSIPASS NEEDLES FRACTURED DURING USE. THE FIRST NEEDLE FRACTURED APPROXIMATELY HALFWAY DOWN THE SHAFT DURING DEPLOYMENT. A SECOND NEEDLE WAS SUBSEQUENTLY OPENED AND USED; HOWEVER, THIS NEEDLE FRACTURED AT THE DISTAL TIP IN TWO LOCATIONS RESULTING IN BROKEN NEEDLE FRAGMENTS BECOMING DISLODGED INTO THE PATIENT'S MENISCUS. X-RAYS WERE UTILIZED TO IDENTIFY AND LOCATE THE RETAINED NEEDLE FRAGMENTS. APPROXIMATELY ONE HOUR OF ADDITIONAL SURGICAL TIME WAS REQUIRED TO LOCATE AND RETRIEVE THE BROKEN NEEDLE PIECES FROM THE PATIENT. DUE TO THE PRESENCE OF RETAINED FRAGMENTS AND THE EFFORTS REQUIRED FOR RETRIEVAL THE PLANNED MENISCUS REPAIR PROCEDURE COULD NOT BE COMPLETED AS INTENDED. INSTEAD, A PARTIAL LATERAL MENISCECTOMY WAS PERFORMED IN ORDER TO REMOVE THE NEEDLE FRAGMENTS ALONG WITH THE AFFECTED MENISCAL TISSUE. THE EVENT RESULTED IN PATIENT INVOLVEMENT, MEDICAL INTERVENTION, AND A SURGICAL DELAY. PHOTOGRAPHS OF THE BROKEN NEEDLES AND X-RAYS WERE PROVIDED WITH THE COMPLAINT.
Patient problems: Laceration(s)
MAUDE report 25427387 ↗MalfunctionApr 23, 2026Material Integrity Problem
LINCOTEK MEDICAL WAS NOTIFIED BY STRYKER ON 04/08/2026 OF A NEEDLE BREAKAGE EVENT THAT OCCURRED AT THE SURGICAL SITE. DURING A MENISCUS ROOT REPAIR PROCEDURE USING THE VERIPASS SUTURE PASSER AND NEEDLE, TWO NEEDLE TIPS FRACTURED WITHIN THE PATIENT. ONE FRAGMENT WAS SUCCESSFULLY RETRIEVED, WHILE THE OTHER REMAINED EMBEDDED IN THE MENISCAL TISSUE AT THE SITE OF BREAKAGE. THE ISSUE WAS IDENTIFIED DURING THE PROCEDURE, WHICH WAS COMPLETED SUCCESSFULLY WITHOUT ANY REPORTED DELAY. HOWEVER, THE ADVERSE OUTCOME WAS THAT ONE NEEDLE TIP REMAINED IN THE PATIENT'S MENISCUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24972039 ↗
FDA clearance and classification
The FDA regulates product code GAB (needle, suturing, disposable) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Versipass Suture Passer Needle 4727
Has the Versipass Suture Passer Needle 4727 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00860794002679), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Versipass Suture Passer Needle 4727?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Apr 23, 2026 and Jun 8, 2026. In those reports, the most common specific device problems are “Material Fragmentation” (1) and “Material Integrity Problem” (1); the most common patient problem is “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Versipass Suture Passer Needle 4727?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Versipass Suture Passer Needle 4727 FDA approved?
Devices of this type (product code GAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Lincotek Medical LLC.
Who makes the Versipass Suture Passer Needle 4727, and is it still sold?
The FDA's device database lists Lincotek Medical LLC as the company responsible for it (brand Versipass Suture Passer Needle, model 4727). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Needle, Suturing, Disposable devices
- Quattro® Suture Passer Needle CM-9011Cayenne Medical, Inc. · Needle, Suturing, Disposable27 reports
- Injector CAT01857Stryker Corporation · NEEDLE, SUTURING, DISPOSABLE8 reports
- Richard-Allan® Needle 1/2 Circle Taper 216704Aspen Surgical Products, Inc. · NEEDLE, SUTURING, DISPOSABLE5 reports
- Richard-Allan® Needle 1/2 Circle Taper 216735Aspen Surgical Products, Inc. · NEEDLE, SUTURING, DISPOSABLE5 reports
- Richard-Allan® Needle 1/2 Circle Taper 216706Aspen Surgical Products, Inc. · NEEDLE, SUTURING, DISPOSABLE4 reports
- Richard-Allan® Needle 1/2 Circle Taper 216701Aspen Surgical Products, Inc. · NEEDLE, SUTURING, DISPOSABLE2 reports
- Richard-Allan® Needle 1/2 Circle Taper 216705Aspen Surgical Products, Inc. · NEEDLE, SUTURING, DISPOSABLE2 reports
- Arthrex® AR-7283Arthrex, Inc. · NEEDLE, SUTURING, DISPOSABLE1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page