Encore System FG0002Recalls & reported problems
Siesta Medical · Implants and delivery tools for hyoid and tongue suspension. · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00862211000206), received between Jan 25, 2023 and May 8, 2026 — classified as 10 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem7
- Insufficient Device Problem Information3
Patient problems
- Skin Infection5
- Post Operative Wound Infection2
- Apnea1
- Implant Pain1
- Pharyngitis1
- Unspecified Infection1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryMay 8, 2026Insufficient Device Problem Information
THE PHYSICIAN TREATED THE PATIENT WITH A HYOID SUSPENSION USING THE ENCORE SYSTEM. THE PATIENT LATER DEVELOPED AN INFECTION IN THE UPPER INCISION SITE OF THE BONE ANCHOR PLACEMENT. THE PATIENT WAS GIVEN A COURSE OF ANTIBIOTICS AND A LAYERED CLOSURE WAS MADE TO ISOLATE THE INFECTION.
Patient problems: Skin Infection
MAUDE report 25124925 ↗InjuryMay 8, 2026Insufficient Device Problem Information
THE PHYSICIAN TREATED THE PATIENT WITH A HYOID SUSPENSION USING THE ENCORE SYSTEM. THE PATIENT LATER DEVELOPED AN INFECTION IN THE UPPER INCISION SITE OF THE BONE ANCHOR PLACEMENT. THE PATIENT WAS GIVEN A COURSE OF ANTIBIOTICS AND A LAYERED CLOSURE WAS MADE TO ISOLATE THE INFECTION.
Patient problems: Skin Infection
MAUDE report 25124767 ↗InjuryAug 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT RECEIVED AN IMPLANT ON (B)(6) 2024. THE AREA BECAME INFECTED WITH STREP. INTERMEDIUS AND WAS TREATED WITH ANTIBIOTICS (PO AND IV). THE DEVICE WAS LATER EXPLANTED ON (B)(6) 2024.
Patient problems: Pharyngitis
MAUDE report 19983586 ↗InjuryApr 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT REPORTED THAT THE INITIAL PROCEDURE WAS NOT EFFECTIVE. A FOLLOW-UP EXAMINATION BY A SECOND PHYSICIAN NOTED THAT THE SUTURES WERE INSUFFICIENTLY TENSIONED. A THIRD PHYSICIAN PERFORMED A SLEEP STUDY AND AIRWAY EXAMINATION. A SECOND PROCEDURE, CONSISTING OF EITHER A TONGUE SUSPENSION OR TIGHTENING OF THE EXISTING SUTURES, DID NOT RESOLVE THE PATIENTS SNORING.
Patient problems: Apnea
MAUDE report 19164296 ↗InjuryFeb 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS TREATED WITH THE ENCORE SYSTEM AND ABOUT 2 WEEKS LATER PRESENTED WITH NECK SWELLING AND AN INFECTION AT THE LOWER INCISION. THE INCISION WAS DRAINED AND PACKED TO TREAT THE INFECTION AND THE PATIENT WAS TREATED WITH AUGMENTIN. THE SYMPTOMS WORSENED, THE PHYSICIAN OPENED AND DRAINED THE INCISION AND CONTINUED THE PATIENT ON AUGMENTIN. ABOUT 1 MONTH LATER, THE DRAINAGE HAS IMPROVED. THE PHYSICIAN DECIDED TO EXPLANT THE DEVICE DUE TO PERSISTENT DRAINAGE. NOTHING OF NOTE WAS OBSERVED DURING THE EXPLANT OR EXPLORATION.
Patient problems: Skin Infection
MAUDE report 18697495 ↗InjuryFeb 9, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
THE PHYSICIAN REPORTED THE PATIENT WAS TREATED WITH THE ENCORE SYSTEM ON 11/16/2023. THE INITIAL TREATMENT WENT WELL, HOWEVER THE INCISION WAS LATER NOTED TO BE "OOZING". A NEGATIVE CULTURE WAS REPORTED.
Patient problems: Skin Infection
MAUDE report 18678453 ↗InjuryFeb 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS IMPLANTED WITH AN ENCORE SYSTEM. HE EXPERIENCED AN INFECTION, FOR WHICH HE RECEIVED ANTIBIOTIC THERAPY AND SURGICAL EXCESION OF GRANULATION TISSUE.
Patient problems: Skin Infection
MAUDE report 18625645 ↗InjuryNov 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PHYSICIAN REPORTED THAT THE PATIENT WAS IMPLANTED WITH THE ENCORE SYSTEM ABOUT A YEAR AGO AND HAD EARLY DRAINAGE FROM THE SUPERIOR INCISION. THIS WAS TREATED WITH ANTIBIOTICS. RECENTLY A PIMPLE-LIKE STRUCTURE WS NOTED AT THE SUPERIOR INCISION. IT WAS NOT CULTURED AS AN INFECTION AND IS NOT PAINFUL. THE PATIENT WAS TREATED WITH ANTIBIOTICS, HOWEVER THIS DID NOT CLEAR THE AREA OF THE INCISION. THE PATIENT IS EXPECTED TO HAVE THE SYSTEM EXPLANTED IN THE NEXT 2-3 WEEKS TO ENSURE THE INCISION HEALS AND THE PATIENT CAN MOVE FORWARD WITH PACEMAKER IMPLANTATION.
Patient problems: Post Operative Wound Infection
MAUDE report 18116224 ↗InjuryJun 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PHYSICIAN REPORTED THE PATIENT WAS SUCCESSFULLY TREATED WITH THE ENCORE SYSTEM ON (B)(6) 2023. ON (B)(6) 2023, THE PHYSICIAN REPORTED AN UNHEALED WOUND, WITH A POST-OPERATIVE INFECTION. A COURSE OF ANTIBIOTICS CLEARED THE INFECTION, HOWEVER A REOCCURRENCE LEAD TO EXPLANT OF THE DEVICE. OF NOTE IS THAT THE PATIENT IS A HEALTHCARE WORKER AND A CULTURE CONFIRMED SERRATIA MARCESCENS, A COMMON ANTIBIOTIC RESISTENT BACTERIA RESPONSIBLE FOR HOSPITAL-ACQUIRED INFECTIONS.
Patient problems: Post Operative Wound Infection
MAUDE report 17214047 ↗InjuryJan 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS IMPLANTED WITH AN ENCORE SYSTEM AND APPROXIMATELY 2 1/2 MONTHS LATER DEVELOPED A SELF-LIMITING STITCH ABSCESS. THE PATIENT RECEIVED 2 DOSES OF STEROIDS AND ANTIBIOTICS. THE DEVICE WAS LATER EXPLANTED, THE WOUND HEALED AND THE PATIENT IS DOING WELL.
Patient problems: Implant Pain; Unspecified Infection
MAUDE report 16242214 ↗
FDA clearance and classification
The FDA regulates product code ORY (prelude tongue suspension system) under 21 CFR 872.5570 as a Class II device. 21 CFR 872.5570 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with a suture. It is indicated for the treatment of obstructive sleep apnea (OSA) and/or snoring.
Common questions about the Encore System FG0002
Has the Encore System FG0002 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00862211000206), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Encore System FG0002?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Jan 25, 2023 and May 8, 2026. In those reports, the most common patient problems are “Skin Infection” (5), “Post Operative Wound Infection” (2) and “Apnea” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Encore System FG0002?
None of the reports are classified as deaths, 10 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Encore System FG0002 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Encore System FG0002, and is it still sold?
The FDA's device database lists Siesta Medical as the company responsible for it (brand Encore System, model FG0002). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page