Skip to content

Zip 8i Surgical Skin Closure Device PS2080Recalls & reported problems

Stryker Corporation · Zip 8i Surgical Skin Closure Device · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00865379000002), received between Aug 15, 2023 and Jan 23, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Close4
  • Biocompatibility2
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Rash2
  • Itching Sensation1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Biocompatibility; Adverse Event Without Identified Device or Use Problem

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: ADVERSE EVENT WITHOUT IDENTIFIED DEVICE OR USE PROBLEM. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: Rash; Skin Inflammation/ Irritation

    MAUDE report 24163999 ↗
  • MalfunctionOct 28, 2025Difficult to Open or Close; Difficult to Open or Close

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 6 EVENTS FOR THE FAILURE: DIFFICULT TO OPEN OR CLOSE. - 6 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399793 ↗
  • MalfunctionJul 29, 2025Difficult to Open or Close; Difficult to Open or Close

    THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: DIFFICULT TO OPEN OR CLOSE. - 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22647408 ↗
  • MalfunctionJul 24, 2024Biocompatibility; Biocompatibility

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE PATIENT EXPERIENCED A RASH AND ITCHING SYMPTOMS AFTER THE DEVICE WAS USED ON THE SUTURE SITE FOR TREATMENT. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: Itching Sensation; Rash

    MAUDE report 19823700 ↗
  • MalfunctionJul 24, 2024Difficult to Open or Close; Difficult to Open or Close

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY DIFFICULT TO OPEN OR CLOSE. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823249 ↗
  • MalfunctionApr 16, 2024Difficult to Open or Close; Difficult to Open or Close

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY DIFFICULT TO OPEN OR CLOSE. 2 EVENTS HAD AN ADDITIONAL DEVICE REQUIRED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19116082 ↗
  • MalfunctionAug 15, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE DEVICE LOCK STRAP DETACHED DURING APPLICATION OF THE ZIP TO A PATIENT. ZIP WAS REPLACED WITH A DIFFERENT ZIP DEVICE. THERE WAS NO PATIENT IMPACT, NO DELAY, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17548563 ↗

Other Tape And Bandage, Adhesive devices