Hydrus® Microstent F00022Recalls & reported problems
Alcon Laboratories, Inc. · Hydrus® Microstent · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Alcon Hydrus Microstent, an implant used to lower eye pressure in glaucoma, has 270 reports, with no deaths and 248 injuries. Many describe the stent sitting in the wrong position or moving, eye pressure rising, and blood in the eye. Some describe inflammation or a stent breaking during insertion. Reports dropped sharply over the last 12 months, to 1 from 10 in the year before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 276
- Deaths reported
- 0
- Injuries reported
- 249
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 276 adverse event reports that list this exact device (DI 00867487000134), received between Oct 29, 2018 and Sep 25, 2026 — classified as 249 injuries, 27 malfunctions.
The most frequently reported specific device problem is “Malposition of Device” (98 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Malposition of Device98
- Adverse Event Without Identified Device or Use Problem61
- Insufficient Device Problem Information58
- Use of Device Problem16
- Obstruction of Flow15
- Migration11
- Separation Failure10
- Difficult to Advance7
Patient problems
- Intraocular Pressure Increased96
- Eye Injury50
- Hyphema46
- Inflammation39
- Insufficient Information35
- Iritis33
- Eye Pain30
- Blurred Vision29
Recent adverse event reports
The 25 most recent of 276 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Difficult to Advance
A non-healthcare professional reported that during a trabecular stent implantation procedure, the stent did not deploy.Additional information was requested but no information is available at the time of this report.
Patient problems: Insufficient Information
MAUDE report 50897322 ↗MalfunctionSep 24, 2026Separation Failure
A health care professional reported that during a trabecular stent implantation procedure, the stent was caught and could not be detached.Additional information has been requested but no information is available at the time of this report.
Patient problems: Insufficient Information
MAUDE report 50881014 ↗MalfunctionSep 21, 2026Separation Failure
The health care professional reported that the arms did not disengage. The timing of the event was unknown.Additional information has been requested but no information is available at the time of this report.
Patient problems: Insufficient Information
MAUDE report 50838308 ↗InjurySep 21, 2026Malposition of Device
A healthcare professional reported via a clinical study that a patient experienced iris inlet adhesions in the left eye following stent implantation. Follow-up information clarified that the event was associated with partial stent-iris touch and intraocular inflammation. Medical treatment was administered, and the inflammation resolved; however, the iris adhesion (synechiae) persisted and could remain unless surgically lysed, although surgical intervention was not considered necessary. The reporter noted no change in intraocular pressure (IOP) and a stable stent, and ongoing monitoring was planned. Iris inlet adhesions, partial stent-iris touch and associated intraocular inflammation were considered related to the stent and not related to the stent implantation procedure or the cataract procedure. No secondary ocular surgical intervention was performed. The inflammation resolved following treatment, while the adhesion/synechiae remained under observation. Onset severity of the inflammation reported as mild.
Patient problems: Adhesion(s); Eye Injury; Inflammation
MAUDE report 50838305 ↗MalfunctionSep 8, 2026Separation Failure
A healthcare professional reported that during a trabecular stent implantation procedure, the stent did not exit the injector.Additional information has been requested but is not available at the time of this report.
Patient problems: Insufficient Information
MAUDE report 50670190 ↗MalfunctionSep 8, 2026Difficult to Advance
A healthcare professional reported that during a trabecular stent implantation procedure, the stent was jammed and was not able to use.Additional information has been requested but no information is available at the time of this report.
Patient problems: Insufficient Information
MAUDE report 50669900 ↗InjuryDec 4, 2025Migration; Migration
A NON-HEALTHCARE PROFESSIONAL HAD REPORTED AFTER TRABECULAR STENT IMPLANTATION SURGERY PATIENT HAD CHOROIDAL FOLDS. BLOOD IN THE CHAMBER HAD PROHIBITED VISUALIZATION, AND ADDITIONAL POSTOPERATIVE CHECKS WERE PERFORMED. INTRAOCULAR PRESSURE HAD SPIKED BUT WAS LATER CONTROLLED WITH DROPS. THE CHOROIDAL FOLDS HAD RESOLVED. THE STENT HAD DIVED. NO HEME HAD BEEN OBSERVED. THE SYMPTOMS WERE RESOLVED.
Patient problems: Hyphema; Intraocular Pressure Increased; Uveitis
MAUDE report 23715047 ↗InjuryOct 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT ONE YEAR AFTER THE STENT IMPLANTATION PROCEDURE, OPTICAL COHERENCE TOMOGRAPHY SHOWED SIGNIFICANT PROGRESSION IN THE RETINAL NERVE FIBER LAYER. THE PATIENT UNDERWENT OCULAR SECONDARY SURGICAL INTERVENTIONS FOR INTRAOCULAR PRESSURE (IOP) OR GLAUCOMA MANAGEMENT IN THE LEFT EYE. DESPITE THESE INTERVENTIONS, IOP AND VISUAL FIELD REMAINED STABLE. A DIRECT SELECTIVE LASER TRABECULOPLASTY HAD BEEN SCHEDULED FOR THE FOLLOWING MONTH. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 23291514 ↗InjurySep 9, 2025Malposition of Device; Malposition of Device
A HEALTHCARE PROFESSIONAL REPORTED THROUGH A CLINICAL STUDY SITE THAT A PATIENT DEVELOPED AN EIGHT-BALL HYPHEMA WITH ELEVATED INTRAOCULAR PRESSURE ON POSTOPERATIVE DAY ONE FOLLOWING A STENT IMPLANTATION PROCEDURE. TO PREVENT CORNEAL BLOOD STAINING, THE PATIENT WAS TAKEN TO THE OPERATING ROOM THE SAME DAY, AND AN ANTERIOR CHAMBER WASHOUT WAS IMMEDIATELY PERFORMED. AT FOLLOW UP DAY AFTER THE WASHOUT, THE PATIENT HAD PERSISTENT POOR VISION BUT SHOWED A MUCH IMPROVED DEGREE OF HYPHEMA. THE PATIENT¿S VISION SUBSEQUENTLY IMPROVED.
Patient problems: Intraocular Pressure Increased; Visual Impairment
MAUDE report 23004310 ↗MalfunctionJul 10, 2025Migration; Migration
A HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE STENT IMPLANTATION PROCEDURE, THE STENT WAS SUCCESSFULLY PLACED AFTER THREE ATTEMPTS BY REPOSITIONING. THREE DAYS AFTER IMPLANTATION, THE DEVICE MIGRATED, INDICATING DISLODGEMENT OR MOVEMENT. THERE WAS PARTIAL EXTRUSION OF THE DEVICE HOWEVER, NO CONTACT WITH THE CORNEA OR IRIS WAS OBSERVED. NO INTERVENTION WAS REQUIRED AT THE TIME. THE EVENT RESOLVED WITHOUT SEQUELAE.
Patient problems: Insufficient Information
MAUDE report 22468198 ↗InjuryMar 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT DURING THE POST OPERATIVE WEEK ONE VISIT AFTER THE STENT IMPLANTATION PROCEDURE, THE PATIENT EXPERIENCED BLURRY VISION AND FLOATERS SINCE POST OPERATIVE DAY 5. THE PATIENT HAD DIFFUSE GRANULOMATOUS KERATIC PRECIPITATES (KPS), ANTERIOR CHAMBER AND VITREOUS CELL. THE PHYSICIAN DID TAP AND INJECT JUST TO BE SAFE, BUT CONJUNCTIVA WAS WHITE AND QUIET, NO PAIN, DIDN¿T REALLY SEEM LIKE ENDOPHTHALMITIS. THE PHYSICIAN THOUGHT IT COULD BE EARLY SYMPATHETIC OPHTHALMIA OR REACTION TO STENT (GRANULOMATOUS INFLAMMATORY REACTION). ADDITIONAL INFORMATION WAS REQUESTED AND RECEIVED STATING POST-OPERATIVELY, GRANULOMATOUS UVEITIS WAS FIRST NOTED ON POST-OPERATIVE DAY 8, ACCOMPANIED BY WORSENED VISION AND ELEVATED INTRAOCULAR PRESSURE. GIVEN THE PRESENTATION OF ANTERIOR CHAMBER AND VITREOUS CELL WITHIN ONE WEEK OF SURGERY, THE CONDITION HAD INITIALLY BEEN TREATED AS A POSSIBLE CASE OF POST-OPERATIVE ENDOPHTHALMITIS. THE PATIENT UNDERWENT A VITREOUS TAP AND RECEIVED AN INTRAVITREAL INJECTION OF CEPHALOSPORIN ANTIBIOTICS AND GLYCOPEPTIDE ANTIBIOTICS ON THE SAME DAY. THE PATIENT HAD BEEN LAST SEEN ON POST OPERATIVE DAY 14 (LAST SEEN DATE), AT WHICH POINT BACTERIAL CULTURES AND
Patient problems: Blurred Vision; Intraocular Pressure Increased; Uveitis; Vitreous Floaters
MAUDE report 21596918 ↗InjuryJan 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTHCARE PROFESSIONAL REPORTED VIA STUDY THAT PATIENT EXPERIENCED A PERIPHERAL ANTERIOR SYNECHIAE WITH A MILD SEVERITY AND RE-POSITIONING.
Patient problems: Adhesion(s)
MAUDE report 21078323 ↗MalfunctionDec 6, 2024Fracture; Fracture
A FACILITY REPRESENTATIVE REPORTED THAT DURING AN STENT IMPLANTATION PROCEDURE, STENT WAS BROKEN WHILE IT WAS BEING INSERTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE PROCEDURE WAS COMPLETED WITH ANOTHER STENT.
Patient problems: Insufficient Information
MAUDE report 20863439 ↗InjuryNov 29, 2024Malposition of Device; Malposition of Device
ADDITIONAL INFORMATION STATED THAT WORSENING OF VISUAL FIELD (MEAN DEVIATION GREATER THAN OR EQUAL TO 2.5 DECIBELS (DB) OR MORE CONFIRMED BY 2 REPEAT MEASUREMENTS).
Patient problems: Insufficient Information; Loss of Vision
MAUDE report 20810726 ↗MalfunctionNov 12, 2024Malposition of Device; Malposition of Device
A PHYSICIAN REPORTED THAT FOLLOWING AN STENT IMPLANT PROCEDURE, PATIENT HAD INCREASED INTRAOCULAR PRESSURE (IOP) OF 13 OS. ON GONIOSCOPY, THE STENT WAS CLEARLY PLUGGED WITH TENTED-UP IRIS TISSUE COMPLETELY OCCLUDING THE LUMEN OF THE PORTION OF THE STENT EXPOSED IN THE ANTERIOR CHAMBER (AC). ADDITIONAL INFORMATION HAS BEEN REQUESTED. THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS ASSOCIATED WITH THE LEFT EYE.
Patient problems: Intraocular Pressure Increased
MAUDE report 20664832 ↗MalfunctionNov 12, 2024Malposition of Device; Malposition of Device
A PHYSICIAN REPORTED THAT FOLLOWING AN STENT IMPLANT PROCEDURE, PATIENT HAD INCREASED INTRAOCULAR PRESSURE (IOP) OF 19 OD. ON GONIOSCOPY, THE STENT WAS CLEARLY PLUGGED WITH TENTED-UP IRIS TISSUE COMPLETELY OCCLUDING THE LUMEN OF THE PORTION OF THE STENT EXPOSED IN THE ANTERIOR CHAMBER (AC). ADDITIONAL INFORMATION HAS BEEN REQUESTED. THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS PATIENT. THIS REPORT IS ASSOCIATED WITH THE RIGHT EYE.
Patient problems: Intraocular Pressure Increased
MAUDE report 20663702 ↗InjuryNov 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING AN STENT IMPLANT PROCEDURE, THE PATIENT HAD HYPOTONY AND DECREASE VISUAL ACUITY (VA). PATIENT WAS STABLE, OBSERVING MAC AND INTRAOCULAR PRESSURE (IOP). THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS 2 OF 2.
Patient problems: Blurred Vision; Intraocular Pressure Decreased
MAUDE report 20598625 ↗InjuryOct 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTHCARE PROFESSIONAL REPORTED WITH A DESCRIPTION OF THE MICROSTENT IMPLANT SURGERY COMBINED WITH CATARACT SURGERY WAS PERFORMED AND DETACHMENT OF DESCEMETS MEMBRANE OCCURRED. AS PER THE SURGEON, IT IS UNCLEAR WHEN THIS SYMPTOM OCCURRED, BUT THAT THE TIP OF THE MICROSTENT CANNULA MAY HAVE PROBABLY HIT THIS AREA DURING ENGAGEMENT. THERE IS NO SPECIFIC TREATMENT BEING TAKEN FOR THE PATIENT AT THIS TIME. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Eye Injury
MAUDE report 20570965 ↗MalfunctionOct 29, 2024Migration; Migration
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FOLLOWING A GLAUCOMA SHUNT IMPLANT PROCEDURE, THE STENT SEEMED TO HAVE BEEN PLACED WELL. HOWEVER, WHEN IT WAS ADVANCED, IT APPEARED TO "BOUNCE BACK" SLIGHTLY. THE DECISION WAS MADE TO LEAVE IT AS THE STENT SEEMED ADEQUATELY ADVANCED. ON POSTOPERATIVE DAY ONE, THE STENT WAS VISIBLE AT THE SLIT LAMP WITHOUT GONIOSCOPY AND WAS PROTRUDING FROM THE BEGINNING OF THE FIRST WINDOW. THERE WAS NO NOTABLE CONTACT WITH THE IRIS OR CORNEA, AND THE PATIENT OTHERWISE APPEARED STABLE, WITH AN INTRAOCULAR PRESSURE (IOP) OF 7 MMHG. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 20562052 ↗InjuryOct 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTH CARE PROFESSIONAL PUBLISHED ORIGINAL ARTICLE WITH A PURPOSE EARLY TO LATE EXPLANTATION OF HYDRUS MICROSTENT MIGS DEVICE: A CASE SERIES. THE HYDRUS MICROS TENT WAS APPROVED BY THE FDA IN (B)(6) 2018 FOR USE WITH CATARACT SURGERY TO REDUCE IOP IN PATIENTS WITH MILD TO MODERATE PRIMARY OPEN ANGLE GLAUCOMA (POAG). PIVOTAL CLINICAL TRIALS DEMONSTRATED ITS OVERALL SAFETY AND EFFICACY IN LOWERING IOP. HOWEVER, MISPOSITIONING OF THE IMPLANT CAN RESULT IN UVEITIS GLAUCOMA- HYPHEMIA (UGH) SYNDROME NECESSITATING DEVICE EXPLANTATION. HERE WE REPORT FOUR SUCH CASES AND THEIR ASSOCIATED CHALLENGES. WE ALSO HIGHLIGHT THE IMPORTANCE OF EARLY RECOGNITION OF POST-OPERATIVE COMPLICATIONS FOR EASE OF IMPLANT REMOVAL. A 71-YEAR-OLD MAN WITH POAG ON DUAL ANTI-PLATELET THERAPY (ASPIRIN AND CLOPIDOGREL) PRESENTED WITH PERSISTENT HYPHEMA AND UNCONTROLLED IOP AFTER THE HYDRUS IMPLANT. DURING HIS CATARACT AND HYDRUS IMPLANT SURGERY, SIGNIFICANT BLEEDING WAS NOTED FROM THE INITIAL GONIOTOMY SITE FOR HYDRUS INSERTION, AND ADDITIONAL VISCOELASTIC WAS ABLE TO TAMPONADE THE BLEED INTRAOPERATIVELY. IN THE POST-OPERATIVE PERIOD, THE PATIENT DEVELOPED A LARGE HYPHEMA AND NASAL CLOT WITH AN IOP OF 64 MMHG AN
Patient problems: Adhesion(s); Intraocular Pressure Increased; UGH (Uveitis Glaucoma Hyphema) Syndrome
MAUDE report 20415701 ↗InjuryOct 9, 2024Malposition of Device; Malposition of Device
A HEALTH CARE PROFESSIONAL PUBLISHED ORIGINAL ARTICLE WITH A PURPOSE EARLY TO LATE EXPLANTATION OF HYDRUS MICROSTENT MIGS DEVICE: A CASE SERIES. THE HYDRUS MICROS TENT WAS APPROVED BY THE FDA IN AUGUST 2018 FOR USE WITH CATARACT SURGERY TO REDUCE IOP IN PATIENTS WITH MILD TO MODERATE PRIMARY OPEN ANGLE GLAUCOMA (POAG). PIVOTAL CLINICAL TRIALS DEMONSTRATED ITS OVERALL SAFETY AND EFFICACY IN LOWERING IOP. HOWEVER, MISPOSITIONING OF THE IMPLANT CAN RESULT IN UVEITIS GLAUCOMA- HYPHEMIA (UGH) SYNDROME NECESSITATING DEVICE EXPLANTATION. HERE WE REPORT FOUR SUCH CASES AND THEIR ASSOCIATED CHALLENGES. WE ALSO HIGHLIGHT THE IMPORTANCE OF EARLY RECOGNITION OF POST-OPERATIVE COMPLICATIONS FOR EASE OF IMPLANT REMOVAL. A 76-YEAR-OLD MAN WITH POAG PRESENTED WITH CHRONIC GRANULOMATOUS IRITIS AFTER UNCOMPLICATED CATARACT AND HYDRUS IMPLANT SURGERY. THE PROXIMAL END OF THE HYDRUS WAS MALPOSITIONED, BEING NOTED TO PROTRUDE FURTHER INTO THE ANTERIOR CHAMBER THAN DESIRED RATHER THAN LYING CLOSE TO THE TM. STILL, IT WAS JUDGED TO BE IN ACCEPTABLE POSITION GIVEN THAT THERE WAS NO CONTACT WITH THE IRIS AND MORE THAN 2/3 OF THE TRANSITION ZONE WAS COVERED. AT THE POM 1, VISUAL ACUITY WAS 20/25 + 1 WITH AN
Patient problems: Uveitis
MAUDE report 20406756 ↗InjuryOct 9, 2024Malposition of Device; Malposition of Device
A HEALTH CARE PROFESSIONAL PUBLISHED ORIGINAL ARTICLE WITH A PURPOSE EARLY TO LATE EXPLANTATION OF HYDRUS MICROSTENT MIGS DEVICE: A CASE SERIES. THE HYDRUS MICROS TENT WAS APPROVED BY THE FDA IN AUGUST 2018 FOR USE WITH CATARACT SURGERY TO REDUCE IOP IN PATIENTS WITH MILD TO MODERATE PRIMARY OPEN ANGLE GLAUCOMA (POAG). PIVOTAL CLINICAL TRIALS DEMONSTRATED ITS OVERALL SAFETY AND EFFICACY IN LOWERING IOP. HOWEVER, MISPOSITIONING OF THE IMPLANT CAN RESULT IN UVEITIS GLAUCOMA- HYPHEMIA (UGH) SYNDROME NECESSITATING DEVICE EXPLANTATION. HERE WE REPORT FOUR SUCH CASES AND THEIR ASSOCIATED CHALLENGES. WE ALSO HIGHLIGHT THE IMPORTANCE OF EARLY RECOGNITION OF POST-OPERATIVE COMPLICATIONS FOR EASE OF IMPLANT REMOVAL. THIS FILE WAS CREATED FOR A 75-YEAR-OLD FEMALE WITH A HISTORY OF BILATERAL CHRONIC ANGLE CLOSURE GLAUCOMA AND AGE-RELATED MACULAR DEGENERATION WAS REFERRED FOR CHRONIC GRANULOMATOUS UVEITIS WITH ELEVATED IOP AFTER CATARACT SURGERY WITH HYDRUS MICROSTENT IMPLANTATION IN THE LEFT EYE 8 MONTHS PRIOR. THE PATIENT COMPLAINED OF NO IMPROVEMENT IN VISION AFTER THE SURGERY AND HAD A HISTORY OF RECURRENT INFLAMMATION ON TAPERING TOPICAL STEROIDS. SHOWED CME WITH A CENTRAL MACULAR THICKNES
Patient problems: Intraocular Pressure Increased; Macular Edema; Uveitis
MAUDE report 20406727 ↗InjuryOct 4, 2024Defective Device; Defective Device
A PHYSICIAN REPORTED THAT AFTER GLAUCOMA SHUNT IMPLANTATION PROCEDURE, THERE WAS DETACHMENT OF DESCEMET MEMBRANE. HE REPORTED THAT IT WAS THE FIRST TIME THAT THE SURGEON WAS USING THE MICROSTENT. HE INCISED TRABECULAR MESHWORK WITH THE TIP OF A CANNULA AND INSERTED IT INTO SCHLEMM CANAL. HOWEVER, HE FELT THAT THE CUT WAS DIFFICULT, BECAUSE THE TIP OF THE CANNULA WAS BIGGER THAN HE EXPECTED OR HE FELT THE TIP WAS NOT SHARP BUT DULL; IT WAS TOO DULL TO MAKE AN INCISION. HE CONSIDERS THAT THOSE FACTORS WOULD AFFECT THE DETACHMENT OF DESCEMET MEMBRANE. CURRENTLY THERE HAS BEEN NO TREATMENT PLAN FOR THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO FURTHER INFORMATION IS AVAILABLE. THERE ARE TWO MEDICAL DEVICE REPORTS ASSOCIATED WITH THIS REPORT. THIS IS 2 OF 2.
Patient problems: Eye Injury
MAUDE report 20369179 ↗InjuryOct 4, 2024Defective Device; Defective Device
ACCORDING TO THE ADDITIONAL INFORMATION RECEIVED, THE PATIENT HAS BEEN RECOVERING. NO CORNEAL EDEMA OR OTHER CORNEAL EFFECTS HAVE OCCURRED AT THIS TIME. VISUAL ACUITY HAS BEEN ACHIEVED. THE TYPE OF PROCEDURE PERFORMED WAS CATARACT SURGERY + GLAUCOMA SHUNT INSERTION.
Patient problems: Eye Injury
MAUDE report 20369047 ↗InjurySep 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT FOLLOWING AN STUNT IMPLANT PROCEDURE, THE PATIENT HAD INTRAOCULAR PRESSURE (IOP) 60 POST IMPLANT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Intraocular Pressure Increased
MAUDE report 20322643 ↗
Other Intraocular Pressure Lowering Implant devices
- iStent GTS100LGlaukos Corporation · Intraocular Pressure Lowering Implant107 reports
- iStent inject W G2-WGlaukos Corporation · Intraocular Pressure Lowering Implant97 reports
- CyPass 8065754003Alcon Laboratories, Inc. · Intraocular Pressure Lowering Implant1 recall1 report
- Cypass 8065754007Alcon Laboratories, Inc. · Intraocular pressure lowering implant1 recall0 reports
- iStent inject G2-M-ISGlaukos Corporation · Intraocular Pressure Lowering Implant33 reports
- iStent GTS100RGlaukos Corporation · Intraocular Pressure Lowering Implant17 reports
- iStent inject W Trabecular Micro-Bypass System with Infinite Injector Technology iS2Glaukos Corporation · Intraocular Pressure Lowering Implant12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page