Skip to content

FlexBand 31051Recalls & reported problems

Artelon · Dynamic Matrix · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00868047000335), received between Dec 16, 2024 and Dec 16, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 16, 2024Break; Break

    IT WAS REPORTED THAT THE SURGEON RE-OPERATED ON A FEMALE PATIENT 18 MONTHS AFTER A RIGHT PERONEUS BREVIS REPAIR. UPON REOPERATION THE SURGEON NOTICED THE DEVICE WAS BROKEN IN HALF IN THE MIDDLE. THE SURGEON SAID SHE WAS DOING GREAT OVER A YEAR AND A HALF ANOTHER ANKLE SPRAIN LIKELY TEARING THE BAND AT THAT TIME.

    Patient problems: Insufficient Information

    MAUDE report 20946899 ↗

Other Mesh, Surgical, Polymeric devices

All Mesh, Surgical, Polymeric devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page