microTargeting™ Guideline 4000 LP+™ MT-LPP-00271Recalls & reported problems
Fhc, Inc. · microTargeting Guideline 4000 LP+ · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00873263002711), received between Jul 3, 2024 and Jul 3, 2024 — classified as 1 injury.
1 other recall cover devices cleared under the same FDA submission (K071364); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Convulsion/Seizure1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
HOSPITAL WAS USING FHC'S MICROELECTRODE RECORDING (MER) SYSTEM ALONG WITH OUR ELECTRODE AND INSERTION TUBES. THE CUSTOMER WAS RECORDING WHEN THE MER SYSTEM HAD A SPACE ISSUE IN THE MEMORY OF THE LAPTOP DUE TO WINDOWS UPDATES, WHICH CAUSED THE USER TO GET AN ERROR MESSAGE REGARDING 'INSUFFICIENT MEMORY', WHILE TRYING TO DELETE SOME OF THE FILES (TO FREE UP MEMORY) AND REBOOTING THE SOFTWARE, THE PATIENT HAD A SEIZURE. THE CASE WAS CANCELED. LATER THAT DAY, THE CUSTOMER PERFORMED FUNCTIONALITY TESTING ON THE SUBJECT PRODUCT WHERE NO ANOMALIES WERE FOUND. THE PRODUCT WAS NOT RETURNED TO FHC FOR EVALUATION AND IS STILL IN USE AT THE HOSPITAL WITHOUT ISSUE.
Patient problems: Convulsion/Seizure
MAUDE report 19670525 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071364. One submission often covers many models, so check the product and lot details in each record.
- Z-1143-2010 · Dec 28, 2009 · TerminatedFHC microTargeting Guideline 4000 ver 1.4 (GL4K) Software, Catalog # MT-GL4K. Intended to assist in functional neurological procedures.
Other ELECTRODE, DEPTH devices
- Spencer Probe Depth Electrode SD04R-AP58X-000Ad-Tech Medical Instrument Corporation · ELECTRODE, DEPTH1 recall0 reports
- NA MICRODEEP DEPTH ELECTRODEDixi Medical · Electrode, Depth1 recall0 reports
- Spencer Probe Depth Electrode SD06R-AP58X-000Ad-Tech Medical Instrument Corporation · ELECTRODE, DEPTH1 recall0 reports
- NA MICRODEEP DEPTH ELECTRODEDixi Medical · Electrode, Depth1 recall0 reports
- NA MICRODEEP DEPTH ELECTRODEDixi Medical · Electrode, Depth1 recall0 reports
- NA MICRODEEP DEPTH ELECTRODEDixi Medical · Electrode, Depth1 recall0 reports
- NA MICRODEEP DEPTH ELECTRODEDixi Medical · Electrode, Depth1 recall0 reports
- NA MICRODEEP DEPTH ELECTRODEDixi Medical · Electrode, Depth1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page