Skip to content

WayPoint™ Surgical Kit 66-WP-SKS-00304Recalls & reported problems

Fhc, Inc. · WayPoint Disposable Surgical Kit, sterile · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00873263003046), received between Mar 31, 2025 and Mar 31, 2025 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Post Operative Wound Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON 3/1 DR. (B)(6) INFORMED FHC REP. ABOUT A POSSIBLE FIDUCIAL INFECTION IN THE PATIENT (B)(6) (SN (B)(6)) DURING THE DBS PROCEDURE ON 2/27. HE SAID IT WAS THE ANTERIOR-RIGHT FIDUCIAL. HE WASN'T SURE IF IT WAS REALLY AN INFECTION OR FAT NECROSIS. THEY DIDN'T CULTURE IT, BUT PROCEEDED AS IF IT WAS AN INFECTION BY REMOVING THE FIDUCIAL AND CLEANING THE COMPROMISED AREA AND ISOLATING IT SO THE CASE DIDN'T CANCEL. DR. (B)(6) CONFIRMED THE PLATFORM WAS STABLE AND THE CASE WAS COMPLETED SUCCESSFULLY. PATIENT WILL COME BACK THE FOLLOWING WEEK TO GET THE IPG GENERATOR IMPLANTED. ON 3/18 FHC REACHED OUT TO DR. (B)(6) TO FOLLOW UP ON ANY UPDATES FROM THIS PATIENT AFTER HIS VISIT TO GET THE GENERATOR. DR. (B)(6) REPLIED AND SAID "SO FAR, NO ISSUES. HE CAME BACK FOR HIS SECOND STAGE PROCEDURE AND EVERYTHING LOOKED LIKE IT WAS HEALING APPROPRIATELY. WE WILL SEE HIM NEXT IN NEUROSURGERY IN A COUPLE OF MONTHS". THERE WAS NO INDICATION OF LARGER HARM TO THE PATIENT. THERE WAS NO INDICATION OF ANY FAILURE OR DEFECT OF THE ANCHOR SCREW (FIDUCIAL). PRODUCT WAS DISCARDED.

    Patient problems: Post Operative Wound Infection

    MAUDE report 21734311 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page