STar™ Drive MA System ST-DS-MA-00441Recalls & reported problems
Fhc, Inc. · STar Drive MA System, manual operation only · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier.
8 other recalls cover devices cleared under the same FDA submission (K092562); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 10, 2018Class IITerminated
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
Z-2349-2018 · Root cause: Device Design · Recall details · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K092562. One submission often covers many models, so check the product and lot details in each record.
- Z-2357-2018 · May 10, 2018 · TerminatedmicroTargeting" STar" Drive System (Manual) Distributed by Medtronic, Product Number FC8001
- Z-2351-2018 · May 10, 2018 · TerminatedmicroTargeting" STar" Drive System (Manual), Product Number 70-ZD-MA
- Z-2352-2018 · May 10, 2018 · TerminatedmicroTargeting" STar" Drive System (Motorized), Product Number 70-ZD-ME
- Z-2350-2018 · May 10, 2018 · TerminatedmicroTargeting" STar" Drive System (Motorized), Product Number ST-DS-ME
- Z-2354-2018 · May 10, 2018 · TerminatedmicroTargeting" Lead Adapter for STar" Drive, Product Number 70-CN-DB
- Z-2358-2018 · May 10, 2018 · TerminatedmicroTargeting" STar" Drive System (Motorized) Distributed by Medtronic, Product Number FC8002
- Z-0312-2011 · Mar 30, 2010 · TerminatedSTar Drive System, Catalog numbers: ST-DS-MA, ST-DS-ME, 70-ZD-ME Product is the FHC DBS Depth Stop Adapter, a component of the microTargeting Drive and the STar Drive. The Drives
- Z-0314-2011 · Mar 30, 2010 · TerminatedDepth Stop Adapter 66-AC-DS(1.8), E6-01-.5 used with the microTargeting and STar Drives Product is the FHC DBS Depth Stop Adapter, a component of the microTargeting Drive and the
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page