Skip to content

COBAS® AmpliPrep / COBAS® TaqMan® HBV Test, v2.0Recalls & reported problems

Roche Molecular Systems, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00875197002706), received between Aug 30, 2022 and Aug 30, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Non Reproducible Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 30, 2022Non Reproducible Results; Non Reproducible Results

    A CUSTOMER IN CHINA ALLEGED DISCREPANT RESULTS FOR ONE PATIENT WHILE USING THE COBAS® AMPLIPREP AND COBAS® TAQMAN® HBV TEST, V2 (CAP/CTM HBV V2). THE ALLEGED SAMPLE INITIALLY A GENERATED POSITIVE RESULT WITH A TITER OF 15,800 IU/ML WITH A COMPETITOR PLATFORM. A REPEAT OF THE SAMPLE ON THAT SYSTEM GAVE CONSISTENT RESULTS. A NEW BLOOD DRAW WAS TAKEN AND TESTED ON THE CAP/CTM HBV V2 ASSAY AND GAVE A RESULT OF TARGET NOT DETECTED (TND). THEN, THE ORIGINAL AND THE NEW COLLECTED SAMPLES WERE RETESTED ON CAP/CTM HBV V2 ASSAY AND BOTH GAVE RESULTS OF TND. THE NEW COLLECTED SAMPLE WAS ALSO RETESTED ON THE COMPETITOR PLATFORM AND PRODUCED A POSITIVE RESULT OF 2,230 IU/ML. THE POSITIVE AND NEGATIVE RESULTS WERE RELEASED. NO HARM IS ALLEGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15321268 ↗

Other Hepatitis Viral B DNA Detection devices

All Hepatitis Viral B DNA Detection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page