Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00875197003024), received between Sep 15, 2025 and Sep 16, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
- Output Problem2
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Non Reproducible Results; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Non Reproducible Results
THE INITIAL REPORTER ALLEGED A FALSE POSITIVE RESULT USING THE KIT CAP-G/CTM HIV-1 V2.0 US-IVD ON THE COBAS AMPLIPREP INSTRUMENT. THE PATIENT, A PREGNANT WOMAN, HAD PREVIOUSLY TESTED NEGATIVE FOR HIV ON AN HIV 1/2 ANTIBODY/ANTIGEN SCREEN CONDUCTED SIX MONTHS EARLIER IN (B)(6) 2020. ON (B)(6) 2021, THE PATIENT WAS TESTED USING AN HIV 4TH GENERATION ANTIBODY/ANTIGEN RAPID SCREEN, WHICH GENERATED A PRESUMPTIVE POSITIVE RESULT. A REFLEX TEST WAS PERFORMED ON THE SAME SAMPLE USING THE HIV 1/2 DIFFERENTIAL ASSAY (BIORAD GENIUS SUPPLEMENTAL ASSAY), WHICH REPORTED A NON-REACTIVE RESULT. ON (B)(6) 2021, THE CUSTOMER CONDUCTED A POLYMERASE CHAIN REACTION (PCR) TEST USING THE KIT CAP-G/CTM HIV-1 V2.0 US-IVD, WHICH GENERATED A POSITIVE RESULT. THE POSITIVE PCR RESULT LED TO THE INITIATION OF A THREE-DRUG TREATMENT REGIMEN FOR THE PATIENT¿S NEWBORN BABY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23064623 ↗MalfunctionSep 15, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem
THE INITIAL REPORTER ALLEGED DISCREPANT RESULTS WITH THE COBAS AMPLIPREP/COBAS TAQMAN HIV-1 TEST, V2.0 FOR TWO PATIENT SAMPLES ON A COBAS AMPLIPREP ANALYZER. SAMPLE 1: THE INITIAL RESULT WAS 425 CP/ML. THE REPEAT RESULT WAS <2.00E+01CP/ML. SAMPLE 2: THE INITIAL RESULT WAS 159 CP/ML. THE REPEAT RESULT WAS TARGET NOT DETECTED. IT IS UNKNOWN IF THE SAMPLES WERE FROM SEPARATE OR THE SAME DRAW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23057780 ↗
Other Test, hiv detection devices
- Determine™ HIV–1/2 Ag/Ab Combo 7D2648Alere Scarborough, Inc. · Test, Hiv Detection409 reports
- Alere Determine™ HIV–1/2 Ag/Ab Combo 7D2648Alere Scarborough, Inc. · Test, Hiv Detection238 reports
- Elecsys HIV Duo 08836973190Roche Diagnostics GmbH · Test, HIV detection105 reports
- ARCHITECT 2P36-01Abbott GmbH · Test, hiv detection1 recall0 reports
- Alere Determine™ HIV–1/2 Ag/Ab Combo 7D2649Alere Scarborough, Inc. · Test, Hiv Detection27 reports
- cobas® HIV-1 09040803190Roche Molecular Systems, Inc. · Test, HIV detection27 reports
- VITROS 6842781Ortho Clinical Diagnostics · Test, hiv detection16 reports
- RealTime 6L18-90Abbott Molecular Inc. · Test, hiv detection12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page