Skip to content

COBAS® AmpliPrep/ COBAS® TaqMan® HIV- 1 Test, version 2.0Recalls & reported problems

Roche Molecular Systems, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00875197003024), received between Sep 15, 2025 and Sep 16, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Output Problem2
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Non Reproducible Results; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Non Reproducible Results

    THE INITIAL REPORTER ALLEGED A FALSE POSITIVE RESULT USING THE KIT CAP-G/CTM HIV-1 V2.0 US-IVD ON THE COBAS AMPLIPREP INSTRUMENT. THE PATIENT, A PREGNANT WOMAN, HAD PREVIOUSLY TESTED NEGATIVE FOR HIV ON AN HIV 1/2 ANTIBODY/ANTIGEN SCREEN CONDUCTED SIX MONTHS EARLIER IN (B)(6) 2020. ON (B)(6) 2021, THE PATIENT WAS TESTED USING AN HIV 4TH GENERATION ANTIBODY/ANTIGEN RAPID SCREEN, WHICH GENERATED A PRESUMPTIVE POSITIVE RESULT. A REFLEX TEST WAS PERFORMED ON THE SAME SAMPLE USING THE HIV 1/2 DIFFERENTIAL ASSAY (BIORAD GENIUS SUPPLEMENTAL ASSAY), WHICH REPORTED A NON-REACTIVE RESULT. ON (B)(6) 2021, THE CUSTOMER CONDUCTED A POLYMERASE CHAIN REACTION (PCR) TEST USING THE KIT CAP-G/CTM HIV-1 V2.0 US-IVD, WHICH GENERATED A POSITIVE RESULT. THE POSITIVE PCR RESULT LED TO THE INITIATION OF A THREE-DRUG TREATMENT REGIMEN FOR THE PATIENT¿S NEWBORN BABY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23064623 ↗
  • MalfunctionSep 15, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE INITIAL REPORTER ALLEGED DISCREPANT RESULTS WITH THE COBAS AMPLIPREP/COBAS TAQMAN HIV-1 TEST, V2.0 FOR TWO PATIENT SAMPLES ON A COBAS AMPLIPREP ANALYZER. SAMPLE 1: THE INITIAL RESULT WAS 425 CP/ML. THE REPEAT RESULT WAS <2.00E+01CP/ML. SAMPLE 2: THE INITIAL RESULT WAS 159 CP/ML. THE REPEAT RESULT WAS TARGET NOT DETECTED. IT IS UNKNOWN IF THE SAMPLES WERE FROM SEPARATE OR THE SAME DRAW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23057780 ↗

Other Test, hiv detection devices

All Test, hiv detection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page