cobas&#x00AE WNV (96 tests) 07001061190Recalls & reported problems

Roche Diagnostics GmbH

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00875197004960), received between Sep 19, 2025 and Sep 19, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result2
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Output Problem2
  • False Negative Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025False Positive Result; False Positive Result

    THE INITIAL REPORTER ALLEGED A DISCREPANT RESULT WITH THE KIT COBAS 6800/8800 WNV 96T IVD ASSAY DURING PARTICIPATION IN THE INSTAND RINGVERSUCH 391 WEST NILE VIRUS (WNV) PANEL. THE CUSTOMER REPORTED THAT ONE SAMPLE, IDENTIFIED AS (B)(6), GENERATED A REACTIVE RESULT WITH A CYCLE THRESHOLD (CT) VALUE OF 31.82, WHILE THE EXPECTED RESULT PROVIDED BY THE PANEL WAS NEGATIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23100572 ↗
  • MalfunctionSep 19, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE INITIAL REPORTER ALLEGED DISCREPANT RESULTS WHEN USING THE KIT COBAS 6800/8800 WNV 96T IVD ASSAY ON THE COBAS 6800 INSTRUMENT. THE CUSTOMER REPORTED THAT A POOL OF 8 SAMPLES WAS REACTIVE FOR WEST NILE VIRUS (WNV). UPON RESOLUTION TESTING, THE POOL WAS REACTIVE WITH A CYCLE THRESHOLD (CT) VALUE OF 38. AN IN-HOUSE POLYMERASE CHAIN REACTION (PCR) TEST PERFORMED ON THE SAME SAMPLE WAS NON-REACTIVE. THE DONATION ASSOCIATED WITH THE SAMPLE WAS NOT USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23100162 ↗
  • MalfunctionSep 19, 2025False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE INITIAL REPORTER ALLEGED THEY OBSERVED INVALID RESULTS USING THE KIT COBAS 6800/8800 WNV 96T IVD ASSAY ON THE COBAS 6800 SYSTEM. THE CUSTOMER USED WEST NILE VIRUS (WNV) CONTROL MATERIAL FROM HELVETICA HEALTH CARE (HHC), CONCENTRATED AT 10,000 COPIES/ML, WHICH THEY DILUTED TO THE DESIRED CONCENTRATION USING FRESH FROZEN PLASMA (FFP). THE CUSTOMER REPORTED NON-REACTIVE RESULTS FOR SOME DILUTIONS OF THE WNV QUALITY CONTROL MATERIAL, WHICH WERE EXPECTED TO BE REACTIVE AS THEY WERE ABOVE THE ASSAY'S LIMIT OF DETECTION (LOD) OF 12.9 COPIES/ML (LINEAGE 1). THE CUSTOMER DID NOT ACCEPT AN INITIAL EXPLANATION THAT THESE RESULTS WERE NEAR THE LOD OF THE ASSAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23097845 ↗