Skip to content

cobas® HIV-1/HIV-2 Qualitative 09040528190Recalls & reported problems

Roche Molecular Systems, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00875197006384), received between Sep 18, 2025 and Aug 12, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result2
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • False Negative Result1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2026False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50219809 ↗
  • MalfunctionJun 3, 2026False Positive Result

    THERE WAS AN ALLEGATION OF DISCREPANT COBAS® HIV-1/HIV-2 QUALITATIVE (192T) RESULTS FROM THE COBAS 5800 ANALYZER FOR A SINGLE PATIENT. ON (B)(6) 2026, THE INITIAL SAMPLE WAS GENERATED HIV-1 REACTIVE (38.9), HIV-2 NON-REACTIVE RESULTS. REPEAT ON (B)(6) 2026 GENERATED HIV-1 REACTIVE (38.21), HIV-2 NON-REACTIVE RESULTS. THE ELECSYS HIV COMBI GENERATED A POSITIVE RESULT (3,63 - 3,68 ¿ 3,73 S/CO). THE BIO-RAD GEENIUS HIV 1/2 SUPPLEMENTAL ASSAY GENERATED A NEGATIVE RESULT. SEROLOGY WAS NEGATIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25373029 ↗
  • MalfunctionSep 18, 2025False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE INITIAL REPORTER ALLEGED FIVE DISCREPANT RESULTS WHEN TESTING WITH THE KIT C58/68/88 HIV-1/HIV-2 QUAL 192T IVD ASSAY ON THE COBAS 8800 SYSTEM. THREE OF THE SAMPLES WERE PLASMA PREPARATION TUBES (PPT), AND TWO WERE PLASMA SAMPLES. ONE OF THE FIVE DISCREPANT RESULTS INVOLVED A KNOWN HIGH-TITER HIV PATIENT SAMPLE TESTED AS PART OF A VALIDATION PROCESS, WHICH YIELDED A NEGATIVE RESULT. THE SAME SAMPLE WAS ALSO TESTED ON A HOLOGIC PLATFORM; HOWEVER, THE RESULTS OF THIS TEST ARE NOT KNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23085584 ↗

Other Test, HIV detection devices

All Test, HIV detection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page