EndoStat 0010-6122Recalls & reported problems
American Medical Systems · 0.2mm 12ft SL · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00878953000053), received between Nov 8, 2017 and Jul 31, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break1
- Detachment Of Device Component1
- Detachment of Device or Device Component1
- Temperature Problem1
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 31, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT WHILE THE SURGEON WAS LASERING THE PATIENT'S EAR; THE END OF THE FIBER BURNT OFF AND BECAME DETACHED. THE SURGEON CONFIRMED THAT THE DETACHED FRAGMENT HAD NOT REMAINED IN THE PATIENT¿S EAR BY LOOKING THROUGH THE MICROSCOPE AND A 30-DEGREE SCOPE. THERE WAS NO PATIENT INJURY DUE TO THIS EVENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7734935 ↗MalfunctionNov 8, 2017Break; Detachment Of Device Component; Temperature Problem; Break; Detachment Of Device Component; Temperature Problem
IT WAS REPORTED THAT DURING STAPEDOTOMY PROCEDURE AIM BEAM WAS TESTED AND NO AIM BEAM PRESENT AT THE END OF THE FIBER WAS OBSERVED. THE FIBER WAS EXCHANGED AND SECOND FIBER WAS USED TO COMPLETE THE PROCEDURE. IT WAS FURTHER REPORTED THAT AFTER THE PROCEDURE THE HOSPITAL TECHNICIAN TRIED THE FIBER ONCE MORE AND LASER SHOT STRAIGHT THROUGH THE FIBER. PATIENT OUTCOME: NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7009647 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.