EndoStat 0010-0622Recalls & reported problems
American Medical Systems · 0.4mm 12ft · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00878953000107), received between Mar 30, 2018 and Aug 24, 2020 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break5
- Adverse Event Without Identified Device or Use Problem2
- Fracture2
- Thermal Decomposition of Device1
Patient problems
- No Consequences Or Impact To Patient6
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 24, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE THE PHYSICIAN NOTICED THE TIP FLARE FROM THE FIBER WAS POOR. THE PERFORMANCE WAS POOR BUT ENOUGH TO COMPLETE THE CASE WITH NO CLINICAL CONSEQUENCES TO THE PATIENT. THE FIBER WAS RETURNED AND ANALYSIS FOUND THAT THE DISTAL END OF THE GLASS FIBER WAS FRACTURED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10445634 ↗MalfunctionJan 15, 2019Break; Break
ANALYSIS OF THE RETURNED DEVICE FOUND THAT THE FIBER CONNECTOR DETACHED FROM THE FIBER AND AIM BEAM IS VISIBLE AT CONNECTOR OPENING. THERE WAS NO PATIENT OR USER IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8251439 ↗MalfunctionSep 24, 2018Break; Break
IT WAS REPORTED THAT WHEN THE PROCEDURE COMPLETED IT WAS NOTICED THAT THE TIP OF THE FIBER WAS MISSING. MANY ATTEMPTS TO FIND AND REMOVE THE TIP WAS UNSUCCESSFUL. THE LENGTH OF THE TIP WAS 3MMLONG AND 0.4MM IN DIAMETER. THE AREA WAS CAREFULLY INSPECTED AND THE MISSING TIP WAS NOT FOUND. IT IS BEING ASSUMED THAT IT WAS SWALLOWED AND WOULD BE PASSED OUT. IT WAS ALSO REPORTED THAT THERE WERE NO COMPLICATIONS TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7904007 ↗MalfunctionAug 28, 2018Fracture; Fracture
ANALYSIS OF THE RETURN DEVICE REVEALED A FRACTURE ON THE DISTAL TIP OF THE FIBER. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7824475 ↗MalfunctionJun 5, 2018Break; Thermal Decomposition of Device; Break; Thermal Decomposition of Device
IT WAS REPORTED THAT DURING USE OF THE FIBER, THE PHYSICIAN SAW "A FLASH". THE CONSOLED WAS PLACED ON STANDBY BY THE LASER TECHNICIAN AND IT WAS NOTICED THAT THE OUTSIDE OF THE SHEATH HAD MELTED. THE FIBER WAS CLEAVED AND RESTRIPED TO COMPLETE THE CASE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7568941 ↗MalfunctionMay 22, 2018Break; Break
ANALYSIS OF THE RETURN DEVICE REVEALED THAT THE FIBER IS BROKEN AND DETACHED FORWARD OF CONNECTOR. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7531592 ↗MalfunctionApr 13, 2018Break; Break
IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING AN OHNS PROCEDURE (OTOLARYNGOLOGY/HEAD AND NECK SURGERY), THE FIBER BROKE OFF AT THE BASE. PATIENT OUTCOME: "NOT HARMED" - THERE WAS NO INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7431060 ↗MalfunctionMar 30, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS UNDER ANESTHESIA FOR A MICROLARYNGOSCOPY PROCEDURE. UPON STARTING THE PROCEDURE, THE FIBER CONNECTOR END PART (*SUBJECT DEVICE) CONNECTING TO THE CONSOLE, BURNT. A SECOND FIBER WAS UTILIZED TO CONTINUE THE PROCEDURE; HOWEVER, THE SAME ISSUE OCCURRED. THE PHYSICIAN ABORTED THE PROCEDURE AFTER THE SECOND FIBER FAILED. THE PATIENT WAS REPORTED TO BE OKAY. THIS IS THE 1ST OF 2 REPORTS FILED FOR THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7387167 ↗MalfunctionMar 30, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS UNDER ANESTHESIA FOR A MICROLARYNGOSCOPY PROCEDURE. UPON STARTING THE PROCEDURE, THE FIBER CONNECTOR END PART CONNECTING TO THE CONSOLE, BURNT. A SECOND FIBER (*SUBJECT DEVICE) WAS UTILIZED TO CONTINUE THE PROCEDURE; HOWEVER, THE SAME ISSUE OCCURRED. THE PHYSICIAN ABORTED THE PROCEDURE AFTER THE SECOND FIBER FAILED. THE PATIENT WAS REPORTED TO BE OKAY. THIS IS THE 2ND OF 2 REPORTS FILED FOR THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7387163 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page