AMS 800 Artificial Urinary Sphincter 72400160Recalls & reported problems
Boston Scientific Corporation · Occlusive Cuff · Class 3 · Implant
All models of AMS 800 Artificial Urinary Sphincter (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific AMS 800 artificial urinary sphincter is implanted to treat urinary incontinence. Its 483 adverse event reports mostly describe the device failing to work, fluid leaks, and recurring incontinence that led to replacement surgery. There are no deaths and 440 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 487
- Deaths reported
- 0
- Injuries reported
- 444
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 487 adverse event reports that list this exact device (DI 00878953000718), received between Oct 3, 2018 and Sep 28, 2026 — classified as 444 injuries, 43 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (151 reports).
6 other recalls cover devices cleared under the same FDA submission (P000053); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem151
- Adverse Event Without Identified Device or Use Problem134
- Insufficient Device Problem Information95
- Fluid/Blood Leak94
- Material Puncture/Hole63
- Inadequacy of Device Shape and/or Size30
- Appropriate Device Problem Term/Code Not Available16
- Defective Device12
Patient problems
- Urinary Incontinence192
- Incontinence77
- No Clinical Signs, Symptoms or Conditions66
- Erosion64
- No Known Impact Or Consequence To Patient52
- Unspecified Tissue Injury48
- Unspecified Infection30
- Tissue Damage19
Recent adverse event reports
The 25 most recent of 487 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Mechanical Problem
Patient problems: Muscle Weakness/Atrophy; Urinary Incontinence
MAUDE report 50917616 ↗InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient with an Artificial Urinary Sphincter (AUS) experienced recurring incontinence. A surgical procedure was performed during which the existing device was explanted. There were no further patient complications reported.
Patient problems: Urinary Incontinence
MAUDE report 50855197 ↗InjurySep 11, 2026Mechanical Problem; Material Puncture/Hole
It was reported that the patient with this Artificial Urinary Sphincter (AUS) presented to the hospital with a non-functional device. Two weeks later, a surgical procedure was performed and upon inspection, there was a crack found in the cuff connecting tube. As a result, the physician removed and replaced the entire device. The patient was stable following the procedure, and there was no additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50733147 ↗InjurySep 11, 2026Mechanical Problem; Material Puncture/Hole
It was reported that the patient with this Artificial Urinary Sphincter (AUS) experienced a progressive recurrence of urinary incontinence. Clinical and endoscopic evaluation identified a mechanical issue with the device. A surgical procedure was performed in which the sphincter was explanted and replaced. Examination of the explanted components identified a perforation of the cuff. There were no further patient complications reported.
Patient problems: Urinary Incontinence
MAUDE report 50731940 ↗InjurySep 10, 2026Fluid/Blood Leak; Mechanical Problem; Material Puncture/Hole
Patient problems: Urinary Incontinence
MAUDE report 50724644 ↗InjurySep 2, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient with this Artificial Urinary Sphincter (AUS) experienced recurrent urinary incontinence. The patient had a previously implanted double cuff AUS consisting of a 4.0 cm cuff and a 4.5 cm cuff. A surgical procedure was performed during which the cuffs, balloon, and control pump were explanted. The AUS system was replaced with a 4.0 cm cuff, a 61-70 balloon, and a new control pump. No additional patient complications were reported.
Patient problems: Urinary Incontinence
MAUDE report 50568998 ↗InjuryAug 11, 2026Adverse Event Without Identified Device or Use Problem; Migration
It was reported that this patient with an Artificial Urinary Sphincter (AUS) experienced perineal pain in (b)(6) 2026. On clinical examination, the cuff was adherent to the skin, with a site of suppuration present. It should be noted that the patient had a medical history of transurethral resectin of the prostate (TURP) in 2021, a radical prostatectomy in 2018, radiotherapy in 2020, and laser urethrotomy for a stone located at the anastomosis site associated with a fistula at the cuff site in (b)(6) 2025. Given the clinical picture of erosion and infection of the device, the physician performed a surgical procedure during which the device was explanted and replaced. No additional information was provided.
Patient problems: Adhesion(s); Erosion; Fistula; Pain; Purulent Discharge; Unspecified Infection
MAUDE report 50209693 ↗InjuryAug 5, 2026Mechanical Problem
It was reported that a patient with an Artificial Urinary Sphincter (AUS) experienced recurring incontinence, and the device stopped functioning. A surgical procedure was performed to remove and replace the entire AUS system. No further patient complications were reported.
Patient problems: Urinary Incontinence
MAUDE report 50143176 ↗InjuryJun 19, 2026Inflation Problem; Mechanical Problem
IT WAS REPORTED THAT THIS PATIENT WITH AN ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRING INCONTINENCE AND THE AUS WAS NOT FUNCTIONING PROPERLY BY NOT INFLATING. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH THE DEVICE WAS EXPLANTED AND REPLACED. THE REPORTED CAUSE OF THE CUFF NO LONGER FUNCTIONING PROPERLY WAS SUSPECTED TO BE ATROPHY. UPON EXPLANT THE SYSTEM STILL HAD FLUID AND NO LEAK WAS OBSERVED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Muscle Weakness/Atrophy; Urinary Incontinence
MAUDE report 25563181 ↗InjuryJun 12, 2026Mechanical Problem
IT WAS REPORTED THAT A PATIENT WITH AN ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRING INCONTINENCE AND THE DEVICE STOPPED WORKING. THE SUSPECTED CAUSE OF THE RETURNING OF INCONTINENCE AND THE DEVICE DEFICIENCY WAS REPORTED TO BE THE END OF THE DEVICES SERVICE LIFE. A SURGICAL PROCEDURE WAS PERFORMED, DURING WHICH THE PHYSICIAN REPAIRED A FOSSA STRICTURE AND REMOVED ONLY THE CUFF TO AVOID EROSION, AS POSTOPERATIVE CATHETERIZATION WAS REQUIRED. ALL OTHER COMPONENTS LEFT IN PLACE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Urethral Stenosis/Stricture; Urinary Incontinence
MAUDE report 25486696 ↗InjuryJun 10, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRING URINARY INCONTINENCE. AS A RESULT, THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Urinary Incontinence
MAUDE report 25460990 ↗InjuryMay 20, 2026Mechanical Problem; Material Puncture/Hole
IT WAS REPORTED THAT THE PATIENT'S ARTIFICIAL URINARY SPHINCTER (AUS) WAS NOT FUNCTIONING PROPERLY. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH A CRACK WAS FOUND IN THE CUFF CONNECTING TUBE. THE CUFF WAS REMOVED AND A NEW 3.5 CM CUFF WAS IMPLANTED; NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25232702 ↗InjuryMay 19, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRING URINARY INCONTINENCE. THE DEVICE WAS DESCRIBED AS HAVING BEEN IMPLANTED FOR NUMEROUS YEARS. A SURGICAL PROCEDURE WAS PERFORMED IN WHICH THE EXISTING DEVICE WAS EXPLANTED AND REPLACED. FOLLOWING IMPLANTATION, CYSTOSCOPIC EVALUATION WAS PERFORMED AND THE NEW DEVICE FUNCTIONED AS EXPECTED. THERE WERE NO FURTHER PATIENT COMPLICATIONS REPORTED.
Patient problems: Urinary Incontinence
MAUDE report 25218697 ↗InjuryMay 13, 2026Mechanical Problem; Material Puncture/Hole
IT WAS REPORTED THAT THE PATIENT VISITED THE HOSPITAL BECAUSE HIS ARTIFICIAL URINARY SPHINCTER (AUS) WAS NOT FUNCTIONING PROPERLY. THE PATIENT UNDERWENT SURGERY TWO WEEKS LATER AND UPON INSPECTION, A CRACK WAS IDENTIFIED IN THE CUFF CONNECTING TUBE. THE CUFF WAS REMOVED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25161409 ↗InjuryApr 21, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRING URINARY INCONTINENCE, PAIN AND CUFF INFECTION. AS A RESULT, THE PATIENT ELECTED TO HAVE THE DEVICE REMOVED WITHOUT THE INVOLVEMENT OR KNOWLEDGE OF A COMPANY REPRESENTATIVE. SEVERAL DAYS LATER, A NEW DEVICE WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Pain; Unspecified Infection; Urinary Incontinence
MAUDE report 24941267 ↗InjuryApr 8, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT WITH THIS ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRENT INCONTINENCE. DURING A REVISION SURGERY, THE PHYSICIAN EXPLANTED AND REPLACED THE DEVICE. THERE WERE NO FURTHER PATIENT COMPLICATIONS.
Patient problems: Unspecified Tissue Injury; Urinary Incontinence
MAUDE report 24824288 ↗InjuryMar 31, 2026Mechanical Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRING URINARY INCONTINENCE. DURING A SURGICAL PROCEDURE, IT WAS DETERMINED THAT THE DEVICE WAS NO LONGER PROVIDING ADEQUATE COAPTATION DUE TO ITS AGE. THE AUS WAS REMOVED AND REPLACED. NO ADDITIONAL PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Urinary Incontinence
MAUDE report 24744503 ↗InjuryMar 2, 2026Fluid/Blood Leak; Mechanical Problem
IT WAS REPORTED THAT A PATIENT WITH AN ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRENT INCONTINENCE AND PRESENTED WITH A NONFUNCTIONING DEVICE. DURING REVISION SURGERY, THE PHYSICIAN CONFIRMED THE PRESENCE OF A FLUID LEAK WITHIN THE SYSTEM, AND THE SUSPECTED CAUSE OF THE FLUID LOSS WAS ATTRIBUTED TO THE AGE OF THE DEVICE. THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Urinary Incontinence
MAUDE report 24483925 ↗InjuryFeb 5, 2026Mechanical Problem; Material Puncture/Hole
IT WAS REPORTED THAT THE PATIENT'S ARTIFICIAL URINARY SPHINCTER (AUS) WAS NOT FUNCTIONING PROPERLY. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH A HOLE WAS FOUND IN THE CUFF CONNECTING TUBE. THE CUFF WAS REPLACED AND NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24273993 ↗InjuryJan 21, 2026Mechanical Problem; Material Puncture/Hole
IT WAS REPORTED THAT A PATIENT WITH AN ARTIFICIAL URINARY SPHINCTER (AUS) PRESENTED WITH A NONFUNCTIONING DEVICE. A REVISION SURGERY WAS PERFORMED, DURING WHICH THE PHYSICIAN CONFIRMED A CRACK IN THE CUFF CONNECTION TUBE. THE DEVICE WAS EXPLANTED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24141478 ↗InjuryJan 5, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS ARTIFICIAL URINARY SPHINCTER (AUS) PRESENTED RECURRING INCONTINENCE. THE AUS WAS EXPLANTED AND REPLACED. THERE IS NO ADDITIONAL INFORMATION REPORTED.
Patient problems: Urinary Incontinence
MAUDE report 23959044 ↗InjuryDec 26, 2025Fluid/Blood Leak; Mechanical Problem; Decrease in Pressure; Inadequacy of Device Shape and/or Size; Fluid/Blood Leak; Mechanical Problem; Decrease in Pressure; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT THE PATIENT WITH THIS ARTIFICIAL URINARY SPHINCTER (AUS) PRESENTED RECURRING INCONTINENCE. IT WAS REPORTED A SIZING PROBLEM WITH THE CUFF. THE AUS WAS EXPLANTED AND REPLACED. THERE IS NO ADDITIONAL INFORMATION REPORTED.
Patient problems: Urinary Incontinence
MAUDE report 23897057 ↗InjuryDec 23, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRENT URINARY INCONTINENCE. CYSTOSCOPY REVEALED THAT THE CUFF REMAINED WIDELY OPEN AND WAS UNABLE TO PROVIDE ADEQUATE CONTINENCE AFTER REACTIVATION. A SURGICAL PROCEDURE WAS PERFORMED IN WHICH THE SYSTEM WAS REMOVED AND REPLACED. THERE WERE NO FURTHER PATIENT COMPLICATIONS REPORTED.
Patient problems: Urinary Incontinence
MAUDE report 23877058 ↗InjuryDec 22, 2025Mechanical Problem; Nonstandard Device; Material Puncture/Hole; Mechanical Problem; Nonstandard Device; Material Puncture/Hole
IT WAS REPORTED THAT A PATIENT WITH AN ARTIFICIAL URINARY SPHINCTER (AUS) PRESENTED WITH A NONFUNCTIONING DEVICE. A REVISION SURGERY WAS PERFORMED, DURING WHICH THE PHYSICIAN CONFIRMED THAT A CRACK WAS FOUND IN THE CUFF CONNECTING TUBE. THE DEVICE WAS EXPLANTED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23873687 ↗InjuryDec 16, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT THE PATIENT WITH AN ARTIFICIAL URINARY SPHINCTER (AUS) EXPERIENCED RECURRING INCONTINENCE DUE TO URETHRAL ATROPHY. ADDITIONALLY, IT WAS NOTED THAT THE CUFF BECAME TOO LARGE FOLLOWING THE ATROPHY. A REVISION SURGERY HAS BEEN REQUESTED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Unspecified Tissue Injury; Urinary Incontinence
MAUDE report 23822352 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P000053. One submission often covers many models, so check the product and lot details in each record.
- Z-0531-2025 · Oct 28, 2024 · Open, ClassifiedAMS Artificial Urinary Sphincter Pressure Regulating Balloon 61-70 cm H2O
- Z-0530-2025 · Oct 28, 2024 · Open, ClassifiedAMS Artificial Urinary Sphincter Control Pump
- Z-2442-2019 · Jul 19, 2019 · TerminatedAMS 800 Artificial Urinary Sphincter AMS 800 Belt Cuff, 4.5 cm Part Number: 72400161
- Z-1280-2013 · Mar 1, 2013 · TerminatedAMS 800 Urinary Control System. The AMS 800 Urinary Control System is used to treat urinary incontinence due to reduced outlet resistance (intrinsic sphincter deficiency) followi
- Z-1279-2013 · Mar 1, 2013 · TerminatedAMS 800 Urinary Control System. The AMS 800 Urinary Control System is used to treat urinary incontinence due to reduced outlet resistance (intrinsic sphincter deficiency) follow
- Z-2669-2011 · May 9, 2011 · TerminatedAMS 800 Urinary Control System Control Pump, REF 72400098, (720172-01 Japan only ), Sterilization Method Steam, American Medical systems Inc., 10700 Bren Road West, Minnetonka, M
Other Device, Incontinence, Mechanical/Hydraulic devices
- AMS 800 Artificial Urinary Sphincter 72404130Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic1 death1,803 reports
- AMS 800 Artificial Urinary Sphincter 72404131Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic1,394 reports
- AMS 800 Artificial Urinary Sphincter 72400024Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic1 recall1 death999 reports
- AMS 800 Artificial Urinary Sphincter 720157-01Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic591 reports
- AMS 800 Artificial Urinary Sphincter 72404132Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic576 reports
- AMS 800 Artificial Urinary Sphincter 72404127Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic443 reports
- AMS 800 Artificial Urinary Sphincter 72400161Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic358 reports
- AMS 800 Artificial Urinary Sphincter 72404133Boston Scientific Corporation · Device, Incontinence, Mechanical/Hydraulic220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page