AMS 700 LGX MS Pump 72404252Recalls & reported problems
Boston Scientific Corporation · Inflatable Penile Prosthesis, Preconnect Penoscrotal, 1 Pump to 2 Cylinders, InhibiZone treated · Class 3 · Implant
All models of AMS 700 LGX MS Pump (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific AMS 700 inflatable penile implant has 756 reports, with 1 death and 683 injuries. Many reports describe fluid leaks, inflation problems, and holes in the parts, and many describe devices being removed and replaced. Reports fell to 37 in the past 12 months from 75 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 757
- Deaths reported
- 1
- Injuries reported
- 684
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 757 adverse event reports that list this exact device (DI 00878953003603), received between Aug 21, 2017 and Sep 25, 2026 — classified as 1 death, 684 injuries, 72 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (293 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak293
- Mechanical Problem293
- Inflation Problem252
- Material Puncture/Hole184
- Adverse Event Without Identified Device or Use Problem117
- Defective Device43
- Insufficient Device Problem Information42
- Deflation Problem41
Patient problems
- No Clinical Signs, Symptoms or Conditions432
- No Known Impact Or Consequence To Patient127
- No Consequences Or Impact To Patient60
- Erosion29
- Pain29
- Unspecified Infection29
- Insufficient Information27
- Discomfort16
Recent adverse event reports
The 25 most recent of 757 reports, in the reporter's own words as published by the FDA.
InjurySep 25, 2026Fluid/Blood Leak; Mechanical Problem; Material Puncture/Hole; Air/Gas in Device
Patient problems: Perforation
MAUDE report 50899777 ↗InjurySep 23, 2026Collapse; Inflation Problem; Mechanical Problem
It was reported that this Inflatable Penile Prosthesis (IPP) would not inflate properly, as the patient experienced issues with the pump not working well. The pump bulb was reported to stay flat. A surgical procedure was performed in which the existing device was explanted and replaced. There were no patient complications reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870441 ↗InjurySep 21, 2026Mechanical Problem; Nonstandard Device; Material Puncture/Hole
It was reported that the patient was unable to cycle the cylinders and presented for revision surgery. During the procedure, the explanted cylinders were observed to have what appeared to be damage to the silicone layer. A replacement surgery was performed, and new cylinders were implanted. No patient complications were reported, and the patient is expected to fully recover.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50842884 ↗InjurySep 7, 2026Collapse; Fluid/Blood Leak; Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50650936 ↗InjurySep 4, 2026Fluid/Blood Leak; Inflation Problem; Material Puncture/Hole; Air/Gas in Device
It was reported that the Inflatable Penile Prosthesis (IPP) would not inflate due to fluid loss. During revision surgery, a hole was visually identified in the tubing. A revision procedure was performed during which the original device was explanted and replaced. No further patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50616506 ↗InjurySep 4, 2026Inflation Problem; Mechanical Problem
It was reported that the patient's Inflatable Penile Prosthesis (IPP) presented buckling of the cylinders and inflation issues. A surgical procedure was performed, during which the IPP was explanted and replaced. No further information was provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50613126 ↗InjuryAug 18, 2026Inflation Problem; Air/Gas in Device
It was reported that this Inflatable Penile Prosthesis (IPP) was not inflating properly. Upon inspection, air in the pump was noticed. As a result the device was explanted and replaced. No patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50311737 ↗InjuryAug 18, 2026Patient-Device Incompatibility
It was reported that a patient with an Inflatable Penile Prosthesis (IPP) was unable to use the device since implantation due to lack of knowledge on how to operate it. In the course of a medical appointment the physician confirmed that the device was properly working, there were no leaks and the cylinders were inflated. The patient will undergo a device use education session. The device was explanted and replaced. There were no device issues or patient complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50308230 ↗InjuryAug 18, 2026Degraded
It was reported that a patient with an Inflatable Penile Prosthesis (IPP) noticed a curvature when inflated the device, which was attributed to an aneurysm. A surgical procedure was performed to remove and replace the IPP. No further patient complications were reported.
Patient problems: Deformity/ Disfigurement
MAUDE report 50306525 ↗InjuryAug 4, 2026Fluid/Blood Leak; Mechanical Problem; Material Puncture/Hole
It was reported that an Inflatable Penile Prosthesis (IPP) stopped functioning. A surgical procedure was performed, during which a small break was identified in the tubing between the pump and one cylinder. This was identified as the site of the leak, as no fluid remained within the device. As a result, the entire IPP system was removed and replaced. Additionally, the physician inadvertently created a hole in one cylinder during device removal. No patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50128734 ↗InjuryJul 7, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) EXPERIENCED PAIN UPON ACTIVATION OF THE DEVICE. DURING REVISION SURGERY, THE PHYSICIAN CONFIRMED THE PRESENCE OF INFECTION. THE DEVICE WAS EXPLANTED AND REPLACED WITH A MALLEABLE PENILE PROSTHESIS. THERE WERE NO FURTHER PATIENT COMPLICATIONS.
Patient problems: Pain; Unspecified Infection
MAUDE report 25742690 ↗InjuryJun 25, 2026Degraded
IT WAS REPORTED THAT THE CYLINDER OF THIS INFLATABLE PENILE PROSTHESIS (IPP) HERNIATED. AS A RESULT, THE DEVICE WAS EXPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25631411 ↗InjuryJun 17, 2026Fluid/Blood Leak; Inflation Problem; Material Puncture/Hole; Connection Problem
IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS (IPP) EXHIBITED INFLATION ISSUES. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH IT WAS IDENTIFIED A FLUID LOSS DUE TO A TUBING DISCONNECTION AND BOTH CYLINDERS RUPTURED REPORTED FROM WEAR AND TEAR. THE IPP WAS EXPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25535438 ↗InjuryJun 1, 2026Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT THE INFLATABLE PENIS PROSTHESIS (IPP) PRESENTED WITH FLUID LOSS. A SURGICAL PROCEDURE WAS PERFORMED, AND THE DEVICE WAS REMOVED AND REPLACED WITH NEW HARDWARE. IT WAS DETERMINED THAT THE RIGHT CYLINDER HAD AN OUT-LAYER BREAK CAUSING THE DEVICE MALFUNCTION. NO COMPLICATIONS HAVE BEEN REPORTED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25350631 ↗InjuryMay 27, 2026Fluid/Blood Leak; Inflation Problem; Material Puncture/Hole; Air/Gas in Device
IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS NOT INFLATING. UPON INSPECTION, A HOLE IN THE TUBING ON THE RIGHT SIDE OF THE CYLINDERS, AIR IN THE SYSTEM, AND ASSOCIATED FLUID LOSS WERE NOTED. AS A RESULT, THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25305989 ↗InjuryMay 22, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) COMPLAINED OF PAIN. A REVISION SURGERY WAS PERFORMED, DURING WHICH THE PHYSICIAN STATED THAT UPON OPENING THE SCROTUM, PURULENT MATERIAL WAS IDENTIFIED WITH AN INFECTION LOCATED IN THE SCROTAL AREA EXTENDING TO THE CYLINDERS. THE PHYSICIAN REMOVED ALL COMPONENTS OF THE PREVIOUS IPP AND PERFORMED A WASHOUT WITH ANTIBIOTICS. THERE WERE NO FURTHER PATIENT COMPLICATIONS.
Patient problems: Pain; Purulent Discharge; Unspecified Infection
MAUDE report 25256598 ↗InjuryApr 30, 2026Collapse; Fluid/Blood Leak; Inflation Problem; Material Puncture/Hole; Air/Gas in Device
IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS (IPP) EXHIBITED INFLATION ISSUES. DURING A SUBSEQUENT SURGICAL PROCEDURE, FLUID LOSS WAS IDENTIFIED AND ATTRIBUTED TO A HOLE IN THE TUBING, WHICH WAS NOTED TO BE RELATED TO THE AGE AND LONGEVITY OF THE DEVICE. ADDITIONALLY, AIR WAS OBSERVED WITHIN THE DEVICE, AND THE PUMP WAS FOUND TO BE COLLAPSED. THE IPP WAS FULLY REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25048853 ↗InjuryApr 28, 2026Inflation Problem; Mechanical Problem; Nonstandard Device
IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) EXHIBITED INFLATION/DEFLATION ISSUES, AND THE PUMP WAS NOT WORKING. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE THE IPP. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25021255 ↗InjuryApr 17, 2026Fluid/Blood Leak; Mechanical Problem
IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) IMPLANT DEVICE WAS NON-FUNCTIONING. A SURGICAL PROCEDURE WAS PERFORMED DURING WHICH THE DEVICE WAS EXPLANTED AND REPLACED. UPON EXPLANT, THE PHYSICIAN IDENTIFIED FLUID LOSS. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24915994 ↗InjuryApr 8, 2026Fluid/Blood Leak; Inflation Problem
IT WAS REPORTED THAT AN INFLATABLE PENILE PROSTHESIS (IPP) EXHIBITED INFLATION ISSUES, AND THE RESERVOIR WAS FOUND TO BE EMPTY, INDICATING FLUID LOSS. HOWEVER, THE SOURCE AND LOCATION OF THE LEAK ARE UNKNOWN. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE THE IPP. NO COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24824514 ↗InjuryApr 2, 2026Fluid/Blood Leak; Inflation Problem
IT WAS REPORTED THAT THE PATIENT WITH AN INFLATABLE PENILE PROSTHESIS (IPP) WAS UNABLE TO INFLATE THE DEVICE DUE TO FLUID LOSS. A SURGICAL PROCEDURE WAS PERFORMED TO REMOVE AND REPLACE THE IPP. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24765913 ↗InjuryFeb 27, 2026Fluid/Blood Leak; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS INFLATABLE PENILE PROSTHESIS (IPP) WAS EXPERIENCING ISSUES WITH THE IMPLANT NOT FUNCTIONING PROPERLY. FLUID LOSS WAS IDENTIFIED. A SURGICAL PROCEDURE WAS PERFORMED IN WHICH THE EXISTING DEVICE WAS EXPLANTED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24467165 ↗InjuryFeb 17, 2026Deflation Problem; Fluid/Blood Leak; Inflation Problem; Material Puncture/Hole
IT WAS REPORTED THAT THE PATIENT'S INFLATABLE PENILE PROSTHESIS (IPP) REQUIRED A REVISION SURGERY ONE MONTH AFTER IMPLANTATION DUE TO THE PUMP DID NOT RESPOND TO EITHER INFLATION OR DEFLATION ATTEMPTS. DURING THE PROCEDURE, THE SURGEON IDENTIFIED A CUT ON ONE OF THE CYLINDERS, WHICH WAS FILLED WITH BLOOD. THE ENTIRE IPP WAS SUBSEQUENTLY REMOVED, AND NO NEW DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24371153 ↗InjuryFeb 2, 2026Fluid/Blood Leak; Mechanical Problem
IT WAS REPORTED THAT THIS INFLATABLE PENILE PROSTHESIS (IPP) PRESENTED A MECHANICAL PROBLEM DUE TO A FLUID LOSS. THE IPP WAS EXPLANTED AN REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24233070 ↗InjuryJan 5, 2026Fluid/Blood Leak; Air/Gas in Device
IT WAS REPORTED THAT THE PATIENT WITH A HISTORY OF ERECTILE DYSFUNCTION AND AN INFLATABLE PENILE PROSTHESIS (IPP) IMPLANT DEVICE NOTED THAT THE DEVICE WAS NOT WORKING AS THERE WAS FLUID LOSS. THE FLUID LOSS WAS IDENTIFIED BY THERE BEING AIR IN THE PUMP. THE PHYSICIAN REMOVED AND REPLACED THE ENTIRE DEVICE THROUGH REPLACEMENT SURGERY, AND THERE WERE NO PATIENT SIGNS OR SYMPTOMS REPORTED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23968850 ↗
Other Prosthesis, Penis, Inflatable devices
- AMS 700 MS Pump 72404310Boston Scientific Corporation · Prosthesis, Penis, Inflatable992 reports
- AMS 700 CX MS Pump 72404233-12Boston Scientific Corporation · Prosthesis, Penis, Inflatable656 reports
- AMS 700 CX MS Pump 72404232Boston Scientific Corporation · Prosthesis, Penis, Inflatable639 reports
- AMS 700 LGX MS Pump 72404251Boston Scientific Corporation · Prosthesis, Penis, Inflatable611 reports
- Conceal 720185-01Boston Scientific Corporation · Prosthesis, Penis, Inflatable586 reports
- AMS 700 LGX MS Pump 72404252-10Boston Scientific Corporation · Prosthesis, Penis, Inflatable561 reports
- AMS 700 CX MS Pump 72404232-10Boston Scientific Corporation · Prosthesis, Penis, Inflatable533 reports
- AMS 700 CX MS Pump 72404231Boston Scientific Corporation · Prosthesis, Penis, Inflatable423 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page