Skip to content

Cobra PzF 170-03-25012Recalls & reported problems

Celonova Biosciences, Inc. · NanoCoated Coronary Stent System · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00879397000661), received between Jan 4, 2019 and Jan 4, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Failure to Advance1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 4, 2019Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated

    ON (B)(6) 2018, (B)(6)-YEAR-OLD PATIENT WITH CORONARY ARTERY DISEASE WITH VESSEL CALCIFICATION AND TORTUOSITY PRESENTED WITH A NON-ST-ELEVATED MYOCARDIAL INFARCTION (NSTEMI). A HEAVILY CALCIFIED LESION IN THE TORTUOUS OBTUSE MARGINAL (OM) WAS IDENTIFIED. AFTER PRE-DILATING THE VESSEL, A 2.75X12MM COBRA STENT WAS DEPLOYED IN THE PROXIMAL TO MID OM WITHOUT ISSUE. AN ATTEMPT WAS THEN MADE TO CROSS THIS STENT VIA RIGHT RADIAL ACCESS FOR TREATMENT OF THE DISTAL OM AS FOLLOWS: A 2.5X12MM COBRA PZF NANOCOATED STENT SYSTEM WAS ADVANCED OVER A WHISPER ES 190CM GUIDE WIRE, AND THROUGH A 5F AL1 LBT GUIDE CATHETER, BUT, DESPITE PUSHING THE DEVICE AGAINST RESISTANCE, IT WAS UNABLE TO CROSS THE 1ST STENT DUE TO TORTUOSITY AND SEVERE CALCIFICATION. WHILE RETRACTING THE STENT SYSTEM, THE STENT STRIPPED (DISLODGED) OFF OF THE BALLOON. ATTEMPTS WERE MADE TO SNARE THE STENT USING A 2ND GUIDE WIRE AND A BALLOON, BUT IT WAS UNABLE TO BE RETRIEVED AND IT REMAINS IN THE ANATOMY. AS THE TARGET LESION WAS PRE-DILATED, THE PHYSICIAN FELT THAT THIS TREATMENT WAS SUFFICIENT AND LEFT THE DISTAL OM LESION AS IS. THERE WERE NO ADVERSE PATIENT SEQUELAE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 8217769 ↗

Other Stent, Coronary devices

All Stent, Coronary devices →