The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00879397000746), received between Aug 20, 2020 and Aug 20, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Material Rupture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Rupture1

Patient problems

  • Angina1
  • Perforation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 20, 2020Material Rupture; Material Rupture

    A (B)(6)-YEAR-OLD MALE PATIENT WITH MEDICAL HISTORY OF MULTI-VESSEL CORONARY ARTERY DISEASE, AND PRIOR ST-ELEVATED MYOCARDIAL INFARCTION (STEMI) TREATED VIA PERCUTANEOUS CORONARY INTERVENTION (PCI) OF NON-TARGET VESSEL, AND PANCREATIC CANCER RECENTLY DIAGNOSED WITH METASTATIS, PRESENTED (B)(6) 2020 FOR PLANNED PCI OF A SEVERELY CALCIFIED LESION IN THE VERY TORTUOUS PROXIMAL LEFT CIRCUMFLEX (LCX) CORONARY ARTERY. THE PATIENT RECEIVED 7000 UNITS HEPARIN TO ACHIEVE AN ACT OF GREATER THAN 300 SECONDS. AFTER ENGAGING A 4.0 GUIDING CATHETER VIA LEFT COMMON FEMORAL ACCESS, AN ATTEMPT WAS MADE TO CROSS THE LESION WITH A WHISPER GUIDE WIRE, BUT IT WAS UNABLE TO CROSS. A FIELDER XT WIRE WAS ABLE TO CROSS. PRE-DILATATION WAS PERFORMED WITH A 2.5 X 12 BALLOON AT 14 ATMOSPHERES (ATMS) FOR 2 INFLATIONS. OF NOTE, THE PROXIMAL CIRCUMFLEX DID HAVE SOME ANEURYSMAL ECTASIA. THERE WAS EVIDENCE OF A SIGNIFICANT STENOSIS DESPITE THE BALLOON INFLATION, SUGGESTING A RING OF CALCIUM. A CUTTING BALLOON WAS UNABLE TO CROSS; A GUIDELINER SUPPORT CATHETER WAS ADVANCED THEN ADDITIONAL DILATATION WAS PERFORMED USING A 2.5X12MM NON-COMPLIANT BALLOON, INFLATED TO 20 ATMS, WITH SATISFACTORY RESULT. A 2.75X18MM COBR

    Patient problems: Angina; Perforation