Skip to content

Cobra PzF 170-03-30012Recalls & reported problems

Celonova Biosciences, Inc. · NanoCoated Coronary Stent System · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00879397000784), received between Sep 6, 2019 and Sep 6, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Deformation1
  • Unstable1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 6, 2019Unstable; Material Deformation; Unstable; Material Deformation

    ON (B)(6) 2019, A PATIENT PRESENTED FOR PLANNED PERCUTANEOUS CORONARY INTERVENTION (PCI) IN AN UNSPECIFIED VESSEL. A 3.0X12MM COBRA PZF¿ NANOCOATED CORONARY STENT SYSTEM WAS PREPPED WITHOUT ISSUE. UPON ENTRY INTO THE GUIDE CATHETER, THE PHYSICIAN NOTED THAT THE STENT WAS DAMAGED / BENT. THE DEVICE WAS NOT USED IN THE PATIENT. ANOTHER UNSPECIFIED DEVICE WAS USED TO TREAT THE TARGET LESION. THE COMPLAINT DEVICE WAS RECEIVED IN THE FOLLOWING CONDITION: THE STENT WAS SHIFTED DISTALLY, OVER THE DISTAL BALLOON MARKER. BLOOD WAS NOTED INSIDE OF THE GUIDE WIRE NOTCH & BALLOON FOLDS; GUIDE WIRE NOTCH DAMAGE WAS OBSERVED. ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE DEVICE WAS PREPPED AND HANDLED WITH BLOODY GLOVES, WAS INTRODUCED OVER GUIDE WIRE, AND INTO HEMOSTATIC VALVE, BUT WAS ULTIMATELY NOT INTRODUCED INTO THE PATIENT SINCE THE STENT DAMAGE (BENT STRUTS AND SHIFTED STENT) WAS NOTED DURING ENTRY INTO VALVE AND GUIDE CATHETER, THEN WAS WITHDRAWN DUE TO THE NOTED DAMAGE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 8970106 ↗

Other Stent, Coronary devices

All Stent, Coronary devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page