Skip to content

Cobra PzF 170-03-40018Recalls & reported problems

Celonova Biosciences, Inc. · NanoCoated Coronary Stent System · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00879397000920), received between Feb 22, 2018 and Dec 16, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Advance2
  • Device Dislodged or Dislocated1
  • Material Deformation1
  • Unstable1

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 16, 2019Unstable; Failure to Advance; Material Deformation; Unstable; Failure to Advance; Material Deformation

    ON (B)(6) 2019, A (B)(6)-YEAR OLD FEMALE PATIENT WITH ST-ELEVATED MYOCARDIAL INFARCTION (STEMI) PRESENTED FOR PLANNED PERCUTANEOUS CORONARY INTERVENTION (PCI) OF AN UNKNOWN VESSEL. AFTER PRE-DILATATION, A 4.0X18MM COBRA PZF¿ CORONARY STENT SYSTEM WAS ADVANCED TOWARD THE TARGET LESION BUT UNABLE TO CROSS. A 3.5 X 16MM SYNERGY STENT WAS THEN PLACED AT THE TARGET LESION. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. THE DEVICE WAS RECEIVED 22-NOV-2019 WITH EXTENSIVE DAMAGE TO THE STENT, WITH THE STENT WAS STRETCHED AND SHIFTED DISTALLY PAST THE DISTAL TIP OF THE DELIVERY SYSTEM, BUT NOT COMPLETELY DISLODGED. ADDITIONAL INFORMATION WAS REQUESTED FROM THE SITE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9473188 ↗
  • MalfunctionFeb 22, 2018Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated

    A (B)(6) YEAR OLD MALE PATIENT PRESENTED WITH LESIONS IN THE PROXIMAL AND DISTAL RIGHT CORONARY ARTERY (RCA). AN ATTEMPT WAS MADE TO CROSS THE PROXIMAL LESION WITH A 4.0X18MM COBRA PZF STENT DELIVERY SYSTEM, BUT IT WAS UNABLE TO CROSS DUE TO VESSEL CALCIFICATION. DURING WITHDRAWAL OF THE STENT SYSTEM THROUGH THE TOUHY VALVE, THE STENT CAME OFF OF THE BALLOON. THE DISLODGED STENT WAS RETRIEVED FROM INSIDE OF THE TOUHY. THE PATIENT WAS SUCCESSFULLY TREATED WITH A NEW 4.0X18MM COBRA. THERE WERE NO ADVERSE PATIENT EFFECTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7289082 ↗

Other Stent, Coronary devices

All Stent, Coronary devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page