Cobra PzF 170-03-40030Recalls & reported problems
Celonova Biosciences, Inc. · NanoCoated Coronary Stent System · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00879397000944), received between Feb 6, 2018 and Oct 10, 2019 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated4
- Failure to Advance2
- Difficult to Advance1
- Difficult to Remove1
- Unstable1
Patient problems
- No Consequences Or Impact To Patient4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 10, 2019Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated; Difficult to Remove; Failure to Advance; Device Dislodged or Dislocated
A 60-YEAR-OLD MALE PATIENT PRESENTED ON (B)(6) 2019 FOR PLANNED PERCUTANEOUS CORONARY INTERVENTION (PCI). ANGIOGRAPHY SHOWED A MODERATELY CALCIFIED LESION IN THE SEVERELY TORTUOUS MID RIGHT CORONARY ARTERY (MRCA) WITH A PREVIOUSLY IMPLANTED STENT IN THE PROXIMAL SEGMENT. THE PHYSICIAN GAINED ARTERIAL ACCESS VIA THE RIGHT RADIAL ARTERY. AFTER PRE-DILATING THE LESION WITH A 3.5X15MM SEMI-COMPLIANT BALLOON, AN ATTEMPT TO CROSS WAS MADE WITH A 4.0X30MM COBRA PZF¿ NANOCOATED CORONARY STENT SYSTEM, BUT THIS DEVICE WAS UNABLE TO CROSS THE LESION DUE TO PATIENT ANATOMY. DURING WITHDRAWAL, RESISTANCE WAS FELT AGAINST THE ANATOMY, AND THE STENT DISLODGED IN THE PROXIMAL PART OF THE RCA. THE DISLODGED STENT WAS SNARED AND RETRIEVED FROM THE VESSEL. A NEW, SMALLER STENT (3.5X30MM COBRA PZF¿) WAS DEPLOYED IN THE MRCA. THERE WERE NO ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED. SUBSEQUENT TO THE INITIAL FILED MEDWATCH REPORT, USER FACILITY MEDWATCH REPORT WAS RECEIVED, AND STATES THE FOLLOWING: "DESCRIBE THE EVENT OR PROBLEM: PHYSICIAN INSERTED A STENT INTO A PATIENT'S CORONARY ARTERY. AS THE PHYSICIAN WAS TAKING THE DEPLOYMENT SYSTEM OUT, IT WAS THEN NOTED
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9174766 ↗MalfunctionAug 20, 2019Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated
A (B)(6)-YEAR-OLD MALE PATIENT PRESENTED WITH ST-ELEVATED MYOCARDIAL INFARCTION (NSTEMI); TREATMENT OF AN UNSPECIFIED VESSEL VIA PERCUTANEOUS CORONARY INTERVENTION (PCI) WAS SCHEDULED. ON (B)(6) 2019, AFTER UNPACKAGING A 4.0X30MM COBRA PZF¿ NANOCOATED CORONARY STENT SYSTEM, THE STENT APPEARED TO BE SHIFTED PROXIMALLY ON THE BALLOON, BUT WAS NOT DISLODGED. THE DEVICE WAS NOT USED IN THE PATIENT; IT WAS RE-PACKAGED BY THE SITE FOR RETURN. CELONOVA BIOSCIENCES RETURNED GOODS LAB RECEIVED THE DEVICE WITHOUT ITS STENT. ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE SITE THOUGHT THE STENT HAD BEEN REPACKAGED WITH ITS DELIVERY SYSTEM AND, WHILE THE CURRENT STENT LOCATION IS NOT KNOWN, THE DEVICE WAS NEVER USED IN THE PATIENT. NO ADDITIONAL INFORMATION AS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8910644 ↗MalfunctionOct 3, 2018Difficult to Advance; Device Dislodged or Dislocated; Difficult to Advance; Device Dislodged or Dislocated
ON (B)(6) 2018, THE PATIENT PRESENTED FOR TREATMENT OF A MILDLY CALCIFIED LESION IN THE NON-TORTUOUS MID RIGHT CORONARY ARTERY (RCA). AFTER PRE-DILATING THE LESION, A 4.0X30MM COBRA PZF NANOCOATED STENT SYSTEM WAS ADVANCED TO THE LESION WITHOUT DIFFICULTY, HOWEVER, THE PHYSICIAN REPORTED RESISTANCE IN THE MID SEGMENT OF THE STENT WHEN THE GUIDE WIRE PASSED THROUGH IT. IN ADDITION, THE PHYSICIAN HAD ALSO DECIDED TO TREAT A DISTAL LESION; THUS, THE PHYSICIAN DECIDED NOT TO USE THIS STENT SYSTEM AND IT WAS SUBSEQUENTLY WITHDRAWN WITHOUT DIFFICULTY. THE PHYSICIAN NOTICED WHAT APPEARED TO BE A POSSIBLE PROTRUDING DISTAL STRUT. A 3.5X24MM COBRA STENT WAS USED TO TREAT DISTAL SEGMENT AND A 4.0X24MM COBRA STENT WAS DEPLOYED TO MID RCA SEGMENT. THE PATIENT WAS NEVER AT RISK AND THERE WAS A GREAT OUTCOME. DURING START OF ANALYSIS OF THE RECEIVED DEVICE, IT WAS NOTED ON 04-SEP-2018 BY CELONOVA BIOSCIENCES THAT THE STENT WAS SHIFTED DISTALLY OVER THE BALLOON, PAST THE DISTAL TIP (PARTIAL DISLODGEMENT). ADDITIONAL INFORMATION RECEIVED FROM THE PHYSICIAN CONFIRMED THAT THE PARTIAL DISLODGEMENT VISUALIZED BY CELONOVA IS WHAT THE PHYSICIAN WAS REFERRING TO WHEN REPORTING THE PROTRUDING DISTAL STRU
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7930041 ↗InjuryFeb 6, 2018Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated
ON (B)(6) 2017, DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE TO IMPLANT A 4.0X30MM COBRA PZF STENT IN LEFT CIRCUMFLEX (LCX), PRE-DILATATION WITH A 2.5X15MM APEX BALLOON WAS PERFORMED. THEN, DURING ADVANCEMENT OF THE COBRA PZF STENT DELIVERY SYSTEM TO THE LCX LESION, IT WAS UNABLE TO CROSS, REPORTEDLY DUE TO VESSEL TORTUOSITY AND THE STENT DISLODGED IN THE PATIENT. THE WHOLE DISLODGED STENT WAS SUCCESSFULLY RETRIEVED USING A GOOSENECK MICROSNARE MICROCATHETER. A 4.0X28MM DRUG-ELUTING STENT (DES) WAS THEN SUCCESSFULLY IMPLANTED. THERE WERE NO ADVERSE PATIENT SEQUELAE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7245812 ↗
Other Stent, Coronary devices
- WALLSTENT® Endoprosthesis H965403130Boston Scientific Corporation · STENT, CORONARY35 reports
- WALLSTENT® Endoprosthesis H965402130Boston Scientific Corporation · STENT, CORONARY23 reports
- Tryton Side Branch Stent T5-2530-191-USTryton Medical, Inc. · Stent, Coronary1 death22 reports
- WALLSTENT® Endoprosthesis H965403330Boston Scientific Corporation · STENT, CORONARY21 reports
- WALLSTENT® Endoprosthesis H965403320Boston Scientific Corporation · STENT, CORONARY10 reports
- WALLSTENT® Endoprosthesis H965403120Boston Scientific Corporation · STENT, CORONARY9 reports
- Tryton Side Branch Stent T5-3035-151-USTryton Medical, Inc. · Stent, Coronary2 deaths7 reports
- MULTI-LINK VISION 1007850-15Abbott Vascular Inc. · STENT, CORONARY7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page