Compress® Device 178404Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00880304002890), received between Jan 7, 2020 and Sep 3, 2025 — classified as 5 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

6 other recalls cover devices cleared under the same FDA submission (K031804, K043547); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 9, 2023Open, Classified

    Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery

    Z-0372-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Fracture2
  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Migration1

Patient problems

  • Pain3
  • Aseptic loosening1
  • Failure of Implant1
  • Joint Dislocation1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 3, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Aseptic loosening; Pain

    MAUDE report 22954641 ↗
  • InjuryDec 11, 2024Detachment of Device or Device Component; Loosening of Implant Not Related to Bone-Ingrowth; Detachment of Device or Device Component; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant; Pain

    MAUDE report 20908217 ↗
  • InjuryDec 20, 2023Fracture; Migration; Fracture; Migration

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN OSS PROCEDURE. SUBSEQUENTLY, THE PATIENT STARTED EXPERIENCING PAIN WHILE SQUATTING. IT WAS DETERMINED THAT THE ANCHOR PLUG CONTACTED THE SPINDLE AND BROKE. THE SURGEON ALSO FOUND THE ANCHOR PLUG HAD MIGRATED ABOUT 1 CM. A SUCCESSFUL REVISION WAS COMPLETED. NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Pain

    MAUDE report 18368318 ↗
  • InjuryFeb 26, 2021Fracture; Fracture

    IT WAS REPORTED THAT PATIENT UNDERWENT A LEFT KNEE REVISION APPROXIMATELY SEVEN MONTHS POST-IMPLANTATION DUE TO IMPLANT FRACTURE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11385238 ↗
  • InjuryJan 7, 2020Device Dislodged or Dislocated; Loosening of Implant Not Related to Bone-Ingrowth; Device Dislodged or Dislocated; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THE PATIENT¿S RIGHT KNEE WAS REVISED APPROXIMATELY ONE MONTH POST IMPLANTATION DUE TO LOOSENING, DISASSOCIATION, AND DISLOCATION OF THE FEMORAL COMPONENTS. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

    Patient problems: Joint Dislocation

    MAUDE report 9561770 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K031804, K043547. One submission often covers many models, so check the product and lot details in each record.

  • Z-1528-2009 · May 19, 2009 · Terminated
    Biomet OSS Letson Proximal Femoral Trial 7 cm resection/right, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472088. A non-powered, orthopedic manual surgical inst
  • Z-1530-2009 · May 19, 2009 · Terminated
    Biomet OSS Elliptical Proximal Femoral Trial 7 cm, right provisional, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472124. A non-powered, orthopedic manual surgic
  • Z-1531-2009 · May 19, 2009 · Terminated
    Biomet OSS Elliptical Proximal Femoral Trial 7 cm left provisional, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472125. A non-powered, orthopedic manual surgical
  • Z-1527-2009 · May 19, 2009 · Terminated
    Biomet OSS FINN Proximal Femoral Trial 7 cm resection/left, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472087. A non-powered, orthopedic manual surgical instrum
  • Z-1529-2009 · May 19, 2009 · Terminated
    Biomet OSS Letson Proximal Femoral Trial 7 cm resection/left, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472089. A non-powered, orthopedic manual surgical instr
  • Z-1526-2009 · May 19, 2009 · Terminated
    Biomet OSS FINN Proximal Femoral Trial 7 cm resection/right, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472086. A non-powered, orthopedic manual surgical instru

Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices

All Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.