RINGLOC BI-POLAR HIP SYSTEM 11-165236Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
All models of RINGLOC BI-POLAR HIP SYSTEM (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00880304209794), received between Nov 30, 2021 and Sep 10, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (K051569, K833175); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Positioning Failure2
- Device Dislodged or Dislocated1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Hematoma1
- Joint Dislocation1
- Muscle/Tendon Damage1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2026Positioning Failure; Material Twisted/Bent
It was reported that the locking ring was slightly bent inside bipolar head and would not allow poly to engage. There was no known impact or consequences to the patient. It was reported that no further information is available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50717722 ↗InjuryNov 11, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT APPROXIMATELY THREE WEEKS POST-IMPLANTATION OF A HEMIARTHROPLASTY, THE PATIENT UNDERWENT A RIGHT HIP REVISION DUE TO MULTIPLE HIP DISLOCATIONS. DURING THE PROCEDURE, THERE WAS HEMATOMA AND BLOOD CLOT EVACUATION AND ENCOUNTERED SHREDDED REMNANTS OF THE SHORT EXTERNAL ROTATORS AND POSTERIOR CAPSULE WHICH HAD BEEN DAMAGED FROM MULTIPLE DISLOCATIONS. ALL COMPONENTS WERE REVISED. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Hematoma; Joint Dislocation; Muscle/Tendon Damage
MAUDE report 20652150 ↗MalfunctionNov 30, 2021Positioning Failure; Positioning Failure
IT WAS REPORTED THAT THE IMPLANT WAS OPENED TO THE STERILE FIELD. THE SCRUB TECH WENT TO PLACE THE PLASTIC AND 28 METAL HEAD INSIDE THE 56MM CUP USING A TWO THUMB METHOD AND IT WOULD NOT ASSEMBLE. SHE PLACED THE 56MM CUP ON HER BACK TABLE AND TRIED USING HER PALM TO ENGAGE. ANOTHER CUP WAS USED TO COMPLETE THE SURGERY. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12895116 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051569, K833175. One submission often covers many models, so check the product and lot details in each record.
- Z-2179-2024 · May 17, 2024 · Open, ClassifiedRingLoc Bi-Polar Hip System Acetabular Cup, 41 mm Outer Diameter - Used in hip hemi-arthroplasty Item Number: 11-165206
- Z-0628-2021 · Nov 25, 2020 · TerminatedRINGLOC BIPOLAR Hip System Bi Polar Locking Ring, Item number 11-165304 - Product Usage: Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phospha
- Z-0097-2021 · Sep 3, 2020 · Terminated(1) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Bi-Polar Cup, E1 Antioxidant Infused, 28 MM, 41 MM: Item No. 110010458; Lot No. 710930; UDI No. (01) 008803045687
- Z-1860-2017 · Feb 17, 2017 · Terminatedvarious polyethylene implants Hips Active Articulation ArComXL Bearings Active Articulation E1 Bearings Arcos Taper Cap BIOLOX delta Option Head E1 RingLoc Bi-Polar Freedom Constr
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗