Compress® Device 178406Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
All models of Compress Device (71)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00880304259744), received between Feb 8, 2021 and Dec 18, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
6 other recalls cover devices cleared under the same FDA submission (K031804, K043547); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Oct 9, 2023Open, Classified
Anchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
Z-0373-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture3
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Failure of Implant1
- No Code Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 18, 2023Fracture; Fracture
IT WAS REPORTED THAT PATIENT UNDERWENT A KNEE ARTHROPLASTY. DURING THE PROCEDURE, THE ANCHOR PLUG FRACTURED DURING ASSEMBLY. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18345196 ↗MalfunctionAug 2, 2023Fracture; Fracture
IT WAS REPORTED THAT THE ANCHOR PLUG BROKE WHEN THREADING ON THE NUT. THERE WAS NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17439895 ↗InjuryFeb 8, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; No Code Available
MAUDE report 11294194 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K031804, K043547. One submission often covers many models, so check the product and lot details in each record.
- Z-1526-2009 · May 19, 2009 · TerminatedBiomet OSS FINN Proximal Femoral Trial 7 cm resection/right, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472086. A non-powered, orthopedic manual surgical instru
- Z-1527-2009 · May 19, 2009 · TerminatedBiomet OSS FINN Proximal Femoral Trial 7 cm resection/left, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472087. A non-powered, orthopedic manual surgical instrum
- Z-1529-2009 · May 19, 2009 · TerminatedBiomet OSS Letson Proximal Femoral Trial 7 cm resection/left, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472089. A non-powered, orthopedic manual surgical instr
- Z-1528-2009 · May 19, 2009 · TerminatedBiomet OSS Letson Proximal Femoral Trial 7 cm resection/right, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472088. A non-powered, orthopedic manual surgical inst
- Z-1530-2009 · May 19, 2009 · TerminatedBiomet OSS Elliptical Proximal Femoral Trial 7 cm, right provisional, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472124. A non-powered, orthopedic manual surgic
- Z-1531-2009 · May 19, 2009 · TerminatedBiomet OSS Elliptical Proximal Femoral Trial 7 cm left provisional, stainless steel, Biomet Orthopedics, Inc., Warsaw, IN; REF 32-472125. A non-powered, orthopedic manual surgical
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page