OSS™ Orthopedic Salvage System 161071Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
All models of OSS Orthopedic Salvage System (76)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00880304456174), received between Apr 28, 2021 and Apr 28, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
Patient problems
- Failure of Implant1
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 28, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
UPON FURTHER INVESTIGATION, IT WAS DETERMINED THAT THIS ITEM WAS REPORTED IN ERROR. THE LOCKING RING FRACTURE IS WHAT CAUSED THE YOKE TO DISASSOCIATE FROM THE TIBIAL TRAY. THEREFORE, THE LOCKING RING IS THE ONLY REPORTABLE ITEM. THE INITIAL REPORT WAS SUBMITTED IN ERROR AND SHOULD BE VOIDED.
Patient problems: Failure of Implant; Pain
MAUDE report 11739345 ↗
Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices
- ELEOS™ 25001208EOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER3 deaths193 reports
- ELEOS™ THSMWRS01MOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER190 reports
- "UNITED" USTAR II Hip System 1115-1413United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls3 reports
- "UNITED" USTAR II Hip System 1115-1213United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1417United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1217United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/11WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page