OSS™ Orthopedic Salvage System 161073Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of OSS Orthopedic Salvage System (76)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00880304456181), received between Apr 28, 2021 and Dec 30, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Device Dislodged or Dislocated1

Patient problems

  • Failure of Implant3
  • Pain2
  • Ambulation or Postural Difficulties1
  • Fall1
  • Hyperextension1
  • Joint Dislocation1
  • Muscle/Tendon Damage1
  • Subluxation1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A LEFT TOTAL KNEE REVISION. THREE MONTHS POSTOP, THE PATIENT REPORTED FALLING AND RUPTURED THE MEDIAL RETINACULAR AND UNDERWENT AN IRRIGATION AND DEBRIDEMENT TO ENSURE THERE WAS NO INFECTION AND THE CAPSULE WAS NOT DAMAGED. NO IMPLANTS WERE EXCHANGED. SIX MONTHS AFTER THE FALL, THE PATIENT REPORTS SHE WOKE UP AND HYPEREXTENDED HER KNEE RESULTING IN PAIN AND THE INABILITY TO BEAR WEIGHT. X-RAY IMAGES SHOWED METAL FRAGMENTS FROM THE LOCKING RING AND POLY FRAGMENTS IN THE JOINT. THE PATIENT WAS REVISED ONE MONTH LATER IN WHICH THE METAL AND POLY HINGE MECHANISMS WERE REPLACED.

    Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Fall; Hyperextension; Muscle/Tendon Damage; Pain; Subluxation

    MAUDE report 21035705 ↗
  • InjuryNov 7, 2024Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant; Joint Dislocation

    MAUDE report 20629816 ↗
  • InjuryApr 28, 2021Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY FIVE MONTHS¿ POST-IMPLANTATION DUE TO LOCKING NUT FRACTURED IN TWO PIECES AND ALLOWED PISTON TO DISENGAGE FROM THE TIBIAL COMPONENT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Pain

    MAUDE report 11739283 ↗

Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page