The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00880304462489), received between Nov 11, 2019 and Nov 4, 2025 — classified as 13 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 17, 2020Class IITerminated

    Potential presence of elevated endotoxin levels that exceed the specification limit

    Z-2144-2020 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Device Dislodged or Dislocated3
  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Insufficient Device Problem Information1
  • Migration1
  • Naturally Worn1
  • Unstable1

Patient problems

  • Failure of Implant4
  • Pain4
  • Joint Dislocation3
  • Inflammation2
  • Insufficient Information2
  • Loss of Range of Motion2
  • Bacterial Infection1
  • Hypersensitivity/Allergic reaction1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT UNDERWENT AN INITIAL SHOULDER PROCEDURE AND WAS SUBSEQUENTLY REVISED AFTER A ROTATOR CUFF FAILURE DUE TO A FALL. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Insufficient Information; Muscle/Tendon Damage

    MAUDE report 23468391 ↗
  • InjuryOct 17, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    NO ADDITIONAL INFORMATION IS AVAILABLE REGARDING THE INCIDENT.

    Patient problems: Joint Dislocation

    MAUDE report 20474182 ↗
  • InjuryOct 5, 2023Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT IS BEING INDICATED FOR AN UPCOMING REVISION PROCEDURE DUE TO SUBLUXATION/DISLOCATION. DURING THE PATIENT'S APPOINTMENT LIGAMENT LAXITY AND VOLUNTARY INSTABILITY WERE NOTED. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Bacterial Infection; Inadequate Osseointegration; Joint Dislocation; Joint Laxity; Medical device site infection; Pain

    MAUDE report 17876078 ↗
  • InjuryOct 3, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT HAD AN INITIAL LEFT SHOULDER ARTHROPLASTY APPROXIMATELY SIX (6) AND A HALF YEARS AGO. SUBSEQUENTLY, THE PATIENT WAS REVISED APPROXIMATELY FIVE (5) YEARS AGO FOR AN UNKNOWN REASON. THE PATIENT WAS REVISED FROM AND ANATOMIC SHOULDER TO A REVERSE SHOULDER.

    Patient problems: Insufficient Information

    MAUDE report 17861417 ↗
  • InjurySep 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint Impingement; Loss of Range of Motion; Pain

    MAUDE report 17818718 ↗
  • InjuryJun 27, 2023Naturally Worn; Naturally Worn

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17210739 ↗
  • InjuryFeb 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A STUDY PATIENT UNDERWENT AN ANATOMIC SHOULDER ARTHROPLASTY. IMMEDIATELY POSTOPERATIVE, THE PATIENT WAS NOTED TO HAVE RADIAL NERVE PALSY AS DIAGNOSED BY DROP HAND. THE PATIENT UNDERWENT EMG AND RADIOLOGIC DIAGNOSTICS. THE PATIENT STARTED USING A DROP HAND SPLINT/ BRACE AND PHYSIOTHERAPY WAS STARTED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.

    Patient problems: Peripheral Nervous Injury

    MAUDE report 16326239 ↗
  • InjuryDec 5, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY DUE TO A TEAR OF THE ROTATOR CUFF. FURTHER, THE PATIENT HAD LOOSENING OF THE GLENOID COMPONENT AND DISLOCATION OF THE POLYETHYLENE FROM THE POST IMPLANT.

    Patient problems: Failure of Implant; Joint Dislocation

    MAUDE report 15914637 ↗
  • InjuryDec 27, 2021Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY APPROXIMATELY FOUR (4) YEARS AND ONE (1) MONTH POST-IMPLANTATION DUE TO ROTATOR CUFF FAILURE AND LOOSE GLENOID COMPONENT. SURGEON STATED THAT AT THE TIME OF THE PRIMARY IMPLANTATION, THE PATIENT HAD UNDIAGNOSED INFLAMMATORY ARTHROPATHY. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Inflammation

    MAUDE report 13087667 ↗
  • InjuryDec 10, 2021Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant

    MAUDE report 12975001 ↗
  • InjurySep 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Hypersensitivity/Allergic reaction; Inflammation; Pain

    MAUDE report 12474517 ↗
  • InjuryMay 17, 2021Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Loss of Range of Motion; Pain

    MAUDE report 11835713 ↗
  • InjuryNov 11, 2019Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT THE PATIENT RECEIVED AN INITIAL LEFT SHOULDER ARTHROPLASTY APPROXIMATELY A YEAR AGO. SUBSEQUENTLY A REVISION OCCURRED ABOUT 2 MONTHS AGO DUE TO A LOOSENING BASEPLATE AND BONE EROSION.

    Patient problems: Failure of Implant

    MAUDE report 9307716 ↗

Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices →