COMPREHENSIVE SHOULDER SYSTEM 113635Recalls & reported problems
Biomet Orthopedics, LLC · Class 2 · Implant
All models of Comprehensive Shoulder System (68)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00880304462663), received between Dec 8, 2020 and Dec 8, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Packaging Problem” (1 report).
1 FDA recall record name this device's identifier.
6 other recalls cover devices cleared under the same FDA submission (K060692); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Jan 7, 2022Class IITerminated
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.
Z-0677-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 8, 2020Packaging Problem; Packaging Problem
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10967860 ↗
FDA clearance and classification
Cleared through 510(k) K060692 on May 30, 2006 (decision: Substantially Equivalent). Submitted by Biomet Manufacturing Corp as “Comprehensive Primary Shoulder Stems”. The FDA review took 76 days. FDA record ↗
The FDA regulates product code KWT (prosthesis, shoulder, non-constrained, metal/polymer cemented) under 21 CFR 888.3650 as a Class II device. 21 CFR 888.3650 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K060692. One submission often covers many models, so check the product and lot details in each record.
- Z-1543-2020 · Feb 17, 2020 · TerminatedComprehensive Shoulder System Mini Humeral Stem 9mm,, Model No. 113629 - Product Usage: The intended use of the comprehensive humeral stem mini is to act as the humeral component o
- Z-0591-2019 · Oct 10, 2018 · TerminatedComprehensive Mini Stem, Item No. 113631:
- Z-0979-2018 · Dec 8, 2017 · TerminatedHumeral Stem 83mm(Size 17) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
- Z-0980-2018 · Dec 8, 2017 · TerminatedHumeral Stem 55mm(Size 13) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
- Z-0299-2018 · Jun 5, 2017 · TerminatedCOMP PRIMARY STEM 8MM MINI, model # 113628
- Z-1755-2015 · Apr 16, 2015 · TerminatedComprehensive Shoulder System, 4 mm, 122 mm long, Primary Shoulder Stem, Ti-6Al-4V, Standard Length Porous Plasma. Shoulder joint replacement components include humeral stems, hum
Common questions about the COMPREHENSIVE SHOULDER SYSTEM 113635
Has the COMPREHENSIVE SHOULDER SYSTEM 113635 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0677-2022 (Class II), began on Jan 7, 2022: The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the COMPREHENSIVE SHOULDER SYSTEM 113635?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 8, 2020. That report lists device problem “Packaging Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the COMPREHENSIVE SHOULDER SYSTEM 113635?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the COMPREHENSIVE SHOULDER SYSTEM 113635 FDA approved?
It was cleared through 510(k) K060692 on May 30, 2006. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the COMPREHENSIVE SHOULDER SYSTEM 113635 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the COMPREHENSIVE SHOULDER SYSTEM 113635, and is it still sold?
The FDA's device database lists Biomet Orthopedics, LLC as the company responsible for it (brand COMPREHENSIVE SHOULDER SYSTEM, model 113635). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices
- Equinoxe 320-42-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented3 recalls105 reports
- Equinoxe 320-38-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented3 recalls105 reports
- Equinoxe 320-42-03Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls85 reports
- Equinoxe 320-38-03Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls42 reports
- Equinoxe 320-42-13Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls41 reports
- Equinoxe 320-10-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented129 reports
- Equinoxe 320-36-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls27 reports
- Equinoxe 320-42-10Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls14 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page