Skip to content

Vanguard Knee System EP-183640Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00880304465749), received between Apr 21, 2022 and Mar 19, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Unstable” (1 report).

9 other recalls cover devices cleared under the same FDA submission (K080528, K113550, K171054); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Unstable1

Patient problems

  • Pain3
  • Loss of Range of Motion2
  • Ankylosis1
  • Arthrofibrosis1
  • Swelling/ Edema1
  • Thromboembolism1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 19, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT MANIPULATION UNDER ANESTHESIA APPROXIMATELY TWO MONTHS POST IMPLANTATION DUE TO PAIN, ARTHROFIBROSIS, AND ANKYLOSIS. SUBSEQUENTLY, THE PATIENT CONTINUES TO REPORT UNRESOLVED EXTREME PAIN TO LEFT KNEE, AND NO ADDITIONAL INTERVENTIONS HAVE BEEN PERFORMED. PATIENT STATED THE BONE CEMENT THAT WAS USED WITH HER IMPLANTS WAS RECALLED DUE TO BEING TOXIC. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE.

    Patient problems: Ankylosis; Arthrofibrosis; Pain

    MAUDE report 24640530 ↗
  • InjuryMar 19, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE MONTH FOLLOWING A LEFT TOTAL KNEE ARTHROPLASTY WAS PERFORMED, A PATIENT WAS ADMITTED TO THE HOSPITAL FOR TREATMENT OF LEFT LEG DVT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE.

    Patient problems: Thromboembolism

    MAUDE report 24640504 ↗
  • InjuryMar 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT KNEE ARTHROPLASTY REVISION TO ADDRESS PAIN, INSTABILITY, LOSS OF RANGE OF MOTION AND LOOSENING OF FEMORAL AND TIBIAL COMPONENTS APPROXIMATELY TWO (2) YEARS POST-OPERATIVELY. ALL COMPONENTS WERE REVISED. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Loss of Range of Motion; Pain

    MAUDE report 21675722 ↗
  • InjuryApr 21, 2022Unstable; Unstable

    IT WAS REPORTED VIA A CLINICAL STUDY THAT A PATIENT UNDERWENT AN INITIAL RIGHT TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, TEN MONTHS POST IMPLANTATION, THE PATIENT WAS REVISED DUE TO PAIN, STIFFNESS, SWELLING, CLICKING, AND INSTABILITY. A CT SCAN PRIOR TO THE REVISION NOTED THAT THE TIBIAL COMPONENT HAD INTERNALLY ROTATED. DURING THE PROCEDURE, THE SURGEON FOUND THAT THERE WAS CEMENT PRESENT ON THE BONE AND TIBIAL IMPLANT, WHICH WAS EASILY REMOVED. THE FEMORAL IMPLANT HAD NO CEMENT ATTACHED. THE PATELLAR COMPONENT WAS LEFT INTACT BUT ALL OTHER COMPONENTS WERE REVISED WITHOUT COMPLICATION.

    Patient problems: Loss of Range of Motion; Pain; Swelling/ Edema

    MAUDE report 14166364 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K080528, K113550, K171054. One submission often covers many models, so check the product and lot details in each record.

  • Z-1042-2023 · Dec 29, 2022 · Open, Classified
    (1) Vanguard Knee System PS Open Box Femoral; Right; 60 mm. Intended for knee joint replacement prostheses. Item Number: 183104 (2) Vanguard Knee System PS Open Box Femoral; Right
  • Z-1043-2023 · Dec 29, 2022 · Open, Classified
    (1) Vanguard Knee System PS Open Box Femoral; Left; 57.5 mm. Intended for Knee joint replacement prostheses. Item Number: 183122 EXPANDED RECALL: (2) Vanguard Knee System PS Ope
  • Z-0938-2019 · Dec 21, 2018 · Terminated
    Vanguard Knee System Series-A Standard Patella sizes 34mm Item Number: 184766
  • Z-0495-2019 · Oct 22, 2018 · Terminated
    Vanguard Knee System -PS Open Box Femoral Left, 62.5 mm Item Number: 183126
  • Z-0496-2019 · Oct 22, 2018 · Terminated
    Vanguard Knee System -PS Open Box Femoral Right, 65 mm Item Number:183108
  • Z-2289-2018 · May 24, 2018 · Terminated
    Vanguard Complete Knee System; E1 Vanguard PS Tibial Bearing; Model Numbers EP-183664, EP-183642 Product Usage: The product is intended for use in knee joint replacement arthropla
  • Z-2177-2018 · May 18, 2018 · Terminated
    Biomet Modular Primary Tibial Tray Implants; Modular Interlok Primary Tray (size 75 mm); Item No. 141215; UDI: (01) 00880304005310 (17) 280306 (10) 385340 Product is used in Total
  • Z-2183-2018 · May 11, 2018 · Terminated
    Vanguard Knee System PS Tibial Bearing, Part Number 183621 Intended for use in knee joint replacement procedures.
  • Z-1011-2018 · Oct 3, 2017 · Terminated
    Vanguard CR Porous Femoral 62.5mm (Left Femur) For use in total knee arthroplasty

Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.