Skip to content

Arcos Modular Revision Hip System 11-301333Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of Arcos Modular Revision Hip System (71)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00880304474260), received between Aug 28, 2024 and Dec 17, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Implant subsidence” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 2, 2021Class IITerminated

    Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention

    Z-2271-2021 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Implant subsidence1
  • Migration1

Patient problems

  • Bone Fracture(s)1
  • Failure of Implant1
  • Joint Dislocation1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 17, 2025Implant subsidence; Implant subsidence

    NO ADDITIONAL INFORMATION AVAILABLE FOR THE REPORTED EVENT.

    Patient problems: Joint Dislocation

    MAUDE report 23831544 ↗
  • InjuryApr 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE TWO MONTHS POST IMPLANTATION DUE TO A FRACTURE AROUND THE STEM IN WHICH THE STEM AND LINER WERE EXCHANGED.THERE IS NO FURTHER INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Bone Fracture(s)

    MAUDE report 21920833 ↗
  • InjuryAug 28, 2024Migration; Migration

    IT WAS REPORTED THAT APPROXIMATELY FOUR MONTHS POST IMPLANTATION OF A RIGHT TOTAL HIP ARTHROPLASTY, THE PATIENT WAS REVISED DUE TO MID-THIGH PAIN WITH STEM SUBSIDENCE. NO ADDITIONAL INFORMATION WAS AVAILABLE.

    Patient problems: Failure of Implant; Pain

    MAUDE report 20086084 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page