Skip to content

Vanguard Knee System CP113618Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of Vanguard Knee System (147)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00880304476844), received between Oct 18, 2023 and Oct 18, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Noise, Audible” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K080204, K171054); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Noise, Audible1

Patient problems

  • Ambulation or Postural Difficulties1
  • Hypersensitivity/Allergic reaction1
  • Joint Laxity1
  • Loss of Range of Motion1
  • Pain1
  • Scar Tissue1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 18, 2023Noise, Audible; Noise, Audible

    IT WAS REPORTED THAT A PATIENT UNDERWENT INITIAL RIGHT TOTAL KNEE ARTHROPLASTY. THE PATIENT DID WELL UNTIL NEARLY A YEAR LATER WHEN THE PATIENT BEGAN TO EXPERIENCE PAIN, INSTABILITY, AUDIBLE CLICKING, AND LIMITED FLEXION DUE TO ARTHROFIBROSIS. METAL ALLERGY TESTING CONFIRMED A REACTIVITY TO NICKEL. THE SURGEON ATTRIBUTED THE PATIENT¿S INSTABILITY TO NICKEL WITHIN THE TITANIUM COMPONENTS. SUBSEQUENTLY, ONE YEAR POST-IMPLANTATION, THE PATIENT UNDERWENT OPEN INCISIONAL BIOPSY FOLLOWED BY RADICAL DEBRIDEMENT AND A REVISION OF ALL TIBIAL AND FEMORAL COMPONENTS. A COMPETITOR SYSTEM WAS IMPLANTED AND THE ORIGINAL PATELLA WAS RETAINED. DUE DILIGENCE IS IN PROGRESS FOR THIS EVENT; TO DATE NO FURTHER INFORMATION HAS BEEN REPORTED.

    Patient problems: Ambulation or Postural Difficulties; Hypersensitivity/Allergic reaction; Joint Laxity; Loss of Range of Motion; Pain; Scar Tissue

    MAUDE report 17956990 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K080204, K171054. One submission often covers many models, so check the product and lot details in each record.

  • Z-0938-2019 · Dec 21, 2018 · Terminated
    Vanguard Knee System Series-A Standard Patella sizes 34mm Item Number: 184766
  • Z-2289-2018 · May 24, 2018 · Terminated
    Vanguard Complete Knee System; E1 Vanguard PS Tibial Bearing; Model Numbers EP-183664, EP-183642 Product Usage: The product is intended for use in knee joint replacement arthropla
  • Z-2177-2018 · May 18, 2018 · Terminated
    Biomet Modular Primary Tibial Tray Implants; Modular Interlok Primary Tray (size 75 mm); Item No. 141215; UDI: (01) 00880304005310 (17) 280306 (10) 385340 Product is used in Total
  • Z-2183-2018 · May 11, 2018 · Terminated
    Vanguard Knee System PS Tibial Bearing, Part Number 183621 Intended for use in knee joint replacement procedures.

Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page