Active Articulation Hip System EP-200152Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of Active Articulation Hip System (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
17
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00880304485068), received between Nov 27, 2018 and Dec 17, 2025 — classified as 17 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (15 reports).

3 other recalls cover devices cleared under the same FDA submission (K101336, K110555, K161190); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated15
  • Detachment of Device or Device Component3
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1
  • Naturally Worn1
  • Unstable1

Patient problems

  • Joint Dislocation15
  • Failure of Implant3
  • Fall3
  • Insufficient Information2
  • Metal Related Pathology2
  • Ambulation or Postural Difficulties1
  • Hematoma1
  • No Information1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 17, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    NO ADDITIONAL INFORMATION AVAILABLE FOR THE REPORTED EVENT.

    Patient problems: Joint Dislocation

    MAUDE report 23831178 ↗
  • InjuryDec 18, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE 3 YEARS POST IMPLANTATION DUE TO A DISLOCATION AT THE SITE. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Joint Dislocation

    MAUDE report 20967895 ↗
  • InjuryDec 11, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.

    Patient problems: Joint Dislocation

    MAUDE report 20901437 ↗
  • InjurySep 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY A PATIENT THAT ONE OF THE IMPLANTED HIP ITEMS WAS RECALLED FOR A CLEANLINESS ISSUE AND IT WAS SUSPECTED THAT METAL IS LEAKING INTO THE BLOODSTREAM. NO REVISION HAS BEEN REPORTED TO DATE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 20138482 ↗
  • InjurySep 3, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: Joint Dislocation

    MAUDE report 20127445 ↗
  • InjuryJul 5, 2024Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: Failure of Implant; Joint Dislocation

    MAUDE report 19680060 ↗
  • InjuryOct 18, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT HIP REVISION PROCEDURE. SUBSEQUENTLY, THE PATIENT DISLOCATED, UNABLE TO REDUCE AND UNDERWENT A REVISION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Joint Dislocation

    MAUDE report 17961227 ↗
  • InjuryAug 31, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT HAD MULTIPLE DISLOCATIONS POST ACETABULAR REVISIONS TO THE G7 DM. THE PATIENT HAD DISLOCATED AND RETURNED TO THE ER WHERE AN INTRAPROSTHETIC DISLOCATION POST CLOSED REDUCTION WAS FOUND. SUBSEQUENTLY, THE PATIENT WAS REVISED FOUR YEARS POST IMPLANTATION TO A FREEDOM CUP.

    Patient problems: Joint Dislocation

    MAUDE report 17664770 ↗
  • InjuryJan 10, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT EXPERIENCED A HIP DISLOCATION AFTER A FALL AND UNDERWENT A REVISION SURGERY APPROXIMATELY 3 MONTHS POST IMPLANTATION. THE DUAL MOBILITY HEAD AND LINER WERE REMOVED AND REPLACED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 16130734 ↗
  • InjuryAug 4, 2022Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    IT WAS REPORTED PATIENT UNDERWENT A HIP REVISION APPROXIMATELY 5 YEARS POST IMPLANTATION DUE TO A DISLOCATION. THE HEAD AND LINER WERE REPLACED. NO ADDITIONAL INFORMATION.

    Patient problems: Failure of Implant; Hematoma; Joint Dislocation; Metal Related Pathology; Pain; Unequal Limb Length

    MAUDE report 15165314 ↗
  • InjuryFeb 10, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED PATIENT UNDERWENT HIP REVISION SURGERY AFTER EXPERIENCING RECURRENT DISLOCATION OF HER THA. THE PATIENT CURRENTLY HAD AN ACTIVE ARTICULATION CONSTRUCT THAT WAS THE RESULT OF A CONVERSION OF A METAL ON METAL MAGNUM CONSTRUCT TO ACTIVE ARTICULATION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Joint Dislocation

    MAUDE report 13508963 ↗
  • InjuryJul 15, 2021Device Dislodged or Dislocated; Naturally Worn; Device Dislodged or Dislocated; Naturally Worn

    IT WAS REPORTED PRIMARY RIGHT THA WAS PLACED. SUBSEQUENTLY PATIENT EXPERIENCED PAIN, AND ELEVATED METAL IONS, AND UNDERWENT RIGHT THA REVISION OF HEAD TO DUAL MOBILITY CERAMIC INNER HEAD AND POLY E1 OUTER BEARING APPROXIMATELY TEN (10) YEARS POST IMPLANTATION. ADDITIONALLY ELEVEN (11) DAYS AFTER THE FIRST REVISION PATIENT UNDERWENT ORIF/REVISION DUE TO RIGHT THA DISLOCATION AND WAS FOUND DISASSOCIATION OF DUAL MOBILITY COMPONENTS, WITH IMPLANT WEAR, DEBRIS AND SMALL AMOUNT OF METALLOSIS WAS NOTED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Joint Dislocation; Metal Related Pathology

    MAUDE report 12173904 ↗
  • InjuryDec 10, 2020Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT A HIP REVISION APPROXIMATELY 1 MONTH POST IMPLANTATION AND A SECOND REVISION APPROXIMATELY 1 MONTH AFTER THAT DUE TO DISLOCATION AND DISSOCIATION OF THE HEAD AND BEARING. IT WAS REPORTED THAT THE PATIENT EXPERIENCED A FALL. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 10983999 ↗
  • InjuryDec 10, 2020Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT A HIP REVISION APPROXIMATELY 1 MONTH POST IMPLANTATION DUE TO DISLOCATION AND DISSOCIATION OF THE HEAD AND BEARING. IT WAS REPORTED THAT THE PATIENT EXPERIENCED A FALL. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Fall; Joint Dislocation

    MAUDE report 10983984 ↗
  • InjuryAug 5, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THE PATIENT UNDERWENT A HIP ARTHROPLASTY. SUBSEQUENTLY, PATIENT WAS REVISED 1 MONTH LATER DUE TO HIP DISLOCATION. SURGEON DISCUSSED THAT THE SHELL REQUIRES MORE VERSION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Joint Dislocation

    MAUDE report 10368918 ↗
  • InjuryFeb 6, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT UNDERWENT A RIGHT HIP REVISION APPROXIMATELY 10 MONTHS POST IMPLANTATION DUE TO DISLOCATION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Joint Dislocation; Patient Problem/Medical Problem

    MAUDE report 9677576 ↗
  • InjuryNov 27, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    REVISION FOR UNKNOWN REASONS. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Insufficient Information; No Information

    MAUDE report 8107306 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K101336, K110555, K161190. One submission often covers many models, so check the product and lot details in each record.

  • Z-1025-2017 · Dec 19, 2016 · Terminated
    Active Articulation ArComXL Polyethylene Bearings Active Articulation ArComXL Bearings item XL-200152
  • Z-1024-2017 · Dec 19, 2016 · Terminated
    RingLoc ArComXL Highly Crosslinked Polyethylene Liners ArComXL High Wall +3mm Liners Size 23 item XL-108323 ArComXL MaxRom +3mm Liners Size 23 item XL-108223
  • Z-1465-2013 · Apr 26, 2013 · Terminated
    Active Articulation E1. Intended for use with either primary or revision hip arthroplasty.

Other Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented devices

All Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page