Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00880304489677), received between Mar 24, 2020 and Feb 23, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Device Contaminated During Manufacture or Shipping” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Contaminated During Manufacture or Shipping4
- Packaging Problem4
- Manufacturing, Packaging or Shipping Problem3
- Tear, Rip or Hole in Device Packaging3
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 23, 2026Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT DAMAGED STERILE PACKAGING WITH URETHANE IDENTIFIED DURING INSPECTION OF CIRCULATED STOCK ITEMS. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION AVAILABLE FOR THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24415748 ↗MalfunctionDec 9, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23759568 ↗MalfunctionDec 8, 2025Packaging Problem; Packaging Problem
A PACKAGING ISSUE WAS IDENTIFIED INVOLVING TAPERLOC STEMS. THE INNER VACUUM PACKAGING WAS OBSERVED TO HAVE LOST VACUUM, ALLOWING THE STEM TO CONTACT AND RUB AGAINST THE INNER PLASTIC. SURGEONS REPORTED CONCERN THAT STERILITY MAY BE COMPROMISED UNDER THESE CONDITIONS AND ELECTED NOT TO USE THE AFFECTED UNITS. NO PATIENTS WERE INVOLVED, AND NO PROCEDURES WERE IMPACTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23743646 ↗MalfunctionOct 28, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
NO ADDITIONAL INFORMATION AVAILABLE FOR THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23404882 ↗MalfunctionMay 9, 2025Packaging Problem; Packaging Problem
IT WAS REPORTED DURING INSPECTION THAT THE STERILE PACKAGING OF THE CIRCULATED ITEMS WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21994584 ↗MalfunctionApr 18, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT WHILE INSPECTING THE PRODUCT, THE STERILE PACKAGING WAS FOUND DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO FURTHER INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21864253 ↗MalfunctionMar 24, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT WHEN INSPECTING THE PRODUCT, IT WAS FOUND WITH INNER PACKAGING DAMAGE. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21676539 ↗MalfunctionSep 26, 2024Packaging Problem; Packaging Problem
NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20311186 ↗MalfunctionFeb 8, 2023Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16328802 ↗MalfunctionMar 3, 2022Device Contaminated During Manufacture or Shipping; Packaging Problem; Device Contaminated During Manufacture or Shipping; Packaging Problem
IT WAS REPORTED THAT THE INSIDE STERILE WRAP APPEARED TO CONTAIN SHAVINGS FROM IMPLANT RUBBING ON WRAP PLASTIC. UPON INSPECTION OF THE PACKAGING, THE VACUUM SEAL HAS BEEN BROKEN, THEREFORE COMPROMISING STERILITY. NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13658781 ↗MalfunctionFeb 15, 2022Tear, Rip or Hole in Device Packaging; Device Contaminated During Manufacture or Shipping; Tear, Rip or Hole in Device Packaging; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT THE OUTSIDE PLASTIC AND BOX OF THE IMPLANT WASN'T DAMAGED, NOR WAS THE PLASTIC CONTAINER INSIDE. HOWEVER, THE PLASTIC AROUND IMPLANT WAS DAMAGED WITH HOLES, AND THERE WAS BLACK AND WHITE DEBRIS INSIDE THE PLASTIC IN WITH THE IMPLANT. THE SURGEON DID NOT USE THE PRODUCT AND HAD TO OPEN A DIFFERENT IMPLANT TO COMPLETE THE PROCEDURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13537655 ↗MalfunctionMar 24, 2020Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED, WHILE INVESTIGATING CIRCULATED ITEMS, MANY PACKAGES HAD DEBRIS IN THE STERILE PACKAGING. NO PATIENTS WERE INVOLVED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9872373 ↗
Other Prosthesis, hip, hemi-, femoral, metal devices
- UHR UH1-48-28Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal2 recalls3 reports
- "UNITED" USTAR II Knee System 2515-3111United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3213United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- STAGEONE 431207Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3115United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- ACTIS 101011050Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal73 reports
- UHR UH1-46-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
- UHR UH1-45-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page