Taperloc Complete Primary Femoral 51-103160Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of Taperloc Complete Primary Femoral (40)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Biomet Taperloc hip implant is a metal femoral stem used in partial hip replacement. The 39 reports mostly describe damaged, torn or unsealed sterile blister packaging, along with contamination or items shipped unsterile, all found while inspecting stock rather than during surgery. One report mentions a bacterial infection; there are no deaths and one injury. There are no recalls naming this device. Reports have dropped sharply, from 13 in the prior 12 months to 2 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 00880304489691), received between Mar 23, 2020 and Nov 15, 2025 — classified as 1 injury, 37 malfunctions.

The most frequently reported specific device problem is “Tear, Rip or Hole in Device Packaging” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Tear, Rip or Hole in Device Packaging11
  • Manufacturing, Packaging or Shipping Problem10
  • Packaging Problem7
  • Device Contaminated During Manufacture or Shipping5
  • Delivered as Unsterile Product4
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Insert1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions32
  • No Patient Involvement4
  • Bacterial Infection1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23564206 ↗
  • MalfunctionSep 30, 2025Packaging Problem; Packaging Problem

    IT WAS REPORTED WHILE INSPECTING THE CIRCULATED ITEMS IN STOCK THAT THE STERILE PACKAGING WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23181338 ↗
  • MalfunctionSep 8, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT WHILE INSPECTING CIRCULATED ITEMS, IT WAS FOUND THAT THE STERILE PACKAGING WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT, AND IT WAS REPORTED NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22991080 ↗
  • MalfunctionAug 13, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT DAMAGED STERILE PACKAGING WAS IDENTIFIED DURING INSPECTION OF CIRCULATED STOCK ITEMS. NO ADDITIONAL INFORMATION AVAILABLE FOR THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22797716 ↗
  • MalfunctionJul 17, 2025Delivered as Unsterile Product; Tear, Rip or Hole in Device Packaging; Delivered as Unsterile Product; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THE INNER POUCH PACKAGING OF THE IMPLANT WAS FOUND DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22531053 ↗
  • MalfunctionJul 14, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT WHILE INSPECTING CIRCULATED ITEMS AT THE DISTRIBUTORSHIP, IT WAS FOUND THAT THE STERILE PACKAGING WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT, AND IT WAS REPORTED NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22492733 ↗
  • MalfunctionJun 20, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT WHILE INSPECTING STOCK, THE STERILE PACKAGE WAS FOUND DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22298617 ↗
  • MalfunctionJun 3, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT DURING THE SURGERY, THE STERILE PACKAGING OF THE STEM WAS FOUND DAMAGED. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22132569 ↗
  • MalfunctionMay 1, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT THE WHILE INSPECTING THE PRODUCT, THE PACKAGING WAS FOUND DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21941326 ↗
  • MalfunctionApr 29, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21930979 ↗
  • MalfunctionApr 28, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT WHILE INSPECTING CIRCULATED ITEMS, IT WAS FOUND THAT THE STERILE PACKAGING WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. IT WAS REPORTED THAT NO FURTHER INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21920711 ↗
  • MalfunctionJan 30, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT THE STERILE PACKAGING WAS FOUND DAMAGED. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21272801 ↗
  • MalfunctionDec 12, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT DURING INSPECTION OF STOCK, THE STERILE POUCH WAS FOUND DAMAGED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20918872 ↗
  • MalfunctionNov 11, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20650922 ↗
  • MalfunctionJul 24, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    NO FURTHER INFORMATION AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19818355 ↗
  • MalfunctionJun 3, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING INSPECTION OF STOCK, THE STERILE PACKAGING WAS FOUND DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19447826 ↗
  • MalfunctionJan 31, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    NO FURTHER INFORMATION AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18614894 ↗
  • MalfunctionJan 10, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18483576 ↗
  • MalfunctionNov 21, 2023Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT WHEN INSPECTING ITEM IN STOCK, IT WAS FOUND THAT THE PACKAGING WAS DAMAGED. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18181746 ↗
  • MalfunctionMay 8, 2023Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT DURING AN INSPECTION OF CIRCULATED ITEMS, IT WAS FOUND THAT THE STERILE PACKAGE WAS DAMAGED. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16887118 ↗
  • MalfunctionJan 11, 2023Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE SURGEON HAD DIFFICULTY WITH THE STEM NOT SEATING TO THE EXACT BROACH SIZE. THE BROACHES WENT DOWN TO WHERE THEY WERE SUPPOSED TO; HOWEVER, THE IMPLANT SAT BETWEEN 1-4 MM PROUD. THE SURGEON WOULD THEN DROP DOWN 2-3 SIZES AND REBROACH. AFTER THAT, THE IMPLANT WOULD SIT IN PLACE. THERE WERE NO PATIENT COMPLICATIONS. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16136852 ↗
  • MalfunctionNov 29, 2022Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT DURING WAREHOUSE INSPECTION, THE STERILE PACKAGE WAS FOUND TO BE DAMAGED. IT IS UNKNOWN IF THERE WAS A BREECH IN STERILITY. THERE WAS NO PATIENT INVOLVEMENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15876086 ↗
  • MalfunctionOct 13, 2022Packaging Problem; Packaging Problem

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15600257 ↗
  • MalfunctionSep 26, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED DAMAGED STERILE PACKAGING WERE IDENTIFIED WHILE INSPECTING CIRCULATED ITEMS. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15484001 ↗
  • MalfunctionJul 21, 2022Separation Failure; Separation Failure

    IT WAS REPORTED STEM AND HEAD WERE IMPLANTED. AFTER REDUCING, SURGEON WANTED TO CHANGE HEAD SIZES. A TAMP WAS INTRODUCED AND ATTEMPTED TO REMOVE IMPLANTED HEAD. SUBSEQUENTLY, THE HEAD WOULD NOT FREE FROM STEM. HAD TO REMOVE STEM AND HEAD AS A MONOBLOCK CONSTRUCT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15075410 ↗

Other Prosthesis, hip, hemi-, femoral, metal devices